Jan Sedlacik1,2, Raphael Tomi-Tricot1,3, Pip Bridgen1,2, Tom Wilkinson1,2, Sharon Giles1,2, Karin Shmueli4, Jo V Hajnal1,2, and Shaihan J Malik1,2
1Centre for the Developing Brain, School of Biomedical Engineering & Imaging Sciences, King’s College London, London, United Kingdom, 2Biomedical Engineering Department, School of Biomedical Engineering & Imaging Sciences, King’s College London, London, United Kingdom, 3MR Research Collaborations, Siemens Healthcare Limited, Frimley, United Kingdom, 4MRI Group, Department of Medical Physics and Biomedical Engineering, University College London, London, United Kingdom
Synopsis
Quantitative
Susceptibility Mapping (QSM)
used for microstructural
assessment of white matter (WM)
is very attractive at
ultra high
magnetic field strengths,
due to the
increased
signal-to-noise ratio (SNR) and
phase sensitivity. This allows
shortening
echo and repetition times and, therefore,
acquisition time. Further
advantages of QSM are
the low flip
angles used for scanning,
which results
in low specific absorption rates,
and the B1
insensitivity of
the signal phase.
However,
suboptimal choice
of imaging parameters will result in
suboptimal SNR.
The purpose of
this work is to find and test
optimal scanning parameters for QSM of the
WM at 7T.
INTRODUCTION
Quantitative
Susceptibility Mapping (QSM)
used for microstructural
assessment of white matter (WM)1
is very attractive at
ultra high
magnetic field (UHF) strengths,
due to the
increased
signal-to-noise ratio (SNR) and
phase sensitivity. This allows
shortening
echo and repetition times and, therefore,
acquisition time. Further
advantages of QSM are
the low flip
angles used for scanning,
which results
in low specific absorption rates (SAR),
and the B1
insensitivity of
the signal phase.2,3
However,
suboptimal choice
of imaging parameters will result in
suboptimal SNR.
The purpose of
this work is to find and test
optimal scanning parameters for QSM of the
WM at 7T.METHODS
The gradient echo (GRE) signal in the WM was calculated using the
equations for RF spoiled4 and unspoiled5 GRE:
$$S_{RFspoilON}=\frac{\sin\!FA\,(1-e^{-TR\diagup T_1})}{(1-\cos\!FA\,e^{-TR\diagup\,T_1})}e^{-TE\diagup\,T_2^*}\qquad\qquad\qquad\,\,(1)\\S_{RFspoilOFF}=\tan\!\left(\frac{FA}{2}\right)\left(1-\frac{(E_1-\cos\!FA)(1-E_2^2)}{\sqrt{p^2-q^2}}\right)\qquad(2)\\\text{with}\,E_1=e^{-TR\diagup
T_1},\,E_1=e^{-TR\diagup\,T_2}\qquad\qquad\qquad\\p=1-E_1\cos\!FA-E_2^2(E_1-\cos\!FA)\\q=E_2(1-E_1)(1+\cos\!FA)\qquad\quad\,\,\\\text{and
the WM relaxation times at 7T:}\qquad\qquad\qquad\\T_1=1220\text{ms},^6\,T_2=45.9\text{ms},^7\,\text{and}\,T_2^*=26.8\text{ms}.^8$$
The echo time (TE) was set to be 3ms shorter than
the repetition time (TR) to accommodate rephase and spoiler gradients
after the readout. The resulting maximum WM
signals
from Equation
1 and 2
for a given TR and its corresponding FA are
shown in Fig.1. The phase SNR was calculated by multiplying the phase
sensitivity, which increases linearly
with TE, with the maximum
WM signal for a
given TR. The WM scanning
efficiency of the GRE phase
images was then
calculated by dividing
the phase SNR by
the square root of the TR.
A
healthy volunteer (male, 29y)
was scanned on
a MAGNETOM Terra
(Siemens Healthcare, Erlangen, Germany) 7T MRI scanner in clinical
mode using the
1/32 transmit/receive head
coil (Nova
Medical, Wilmington MA, USA) with informed consent and
Institutional
Research Ethics Committee approval. Multi
echo GRE images
for QSM were
acquired with the following parameters:
TEs=2.68/7.37/12.06/16.75/21.44/26.13ms,
TR=29ms, RF spoil
ON with FAs=12.5 and 15.5°
and RF spoil OFF with FAs=15.5 and 19°
(see results and
below for further information),
isotropic 0.7mm voxel size and
2x2 parallel acquisition acceleration using CAIPIRINHA9
and scan time for
each acquisition 5:35min.
To better assess
the effect of the RF spoiling on the signal, the scan with RF spoiling ON was
repeated with FA=15.5° to match the optimal
FA of the scan with RF spoiling OFF. The impact on the signal of such a
24% FA increase was also tested for the scan with RF spoil OFF, which was repeated with 19° FA.
The data sets
were realigned to the mean of all scans using FSL-FLIRT10
the realignment was repeated twice with an updated reference. The
brain was extracted using FSL BET.11
All processing used MATLAB (The Mathworks Inc.,
Natick MA, USA). R2*
maps were computed using the Moore-Penrose pseudoinverse. For QSM,
the phase images of each receive channel were combined using the
phase difference between two consecutive echoes, which eliminates the
individual coil sensitivities. Complex
fitting12
was used to fit the phase over all TEs and the local phase was
extracted by Laplacian background field removal.13
Susceptibility calculation used Iterative Tikhonov regularisation
(α=0.05).14RESULTS
Figure
1A shows that the maximum
WM signal declines
with increasing TR
with RF spoiling
OFF, while with RF
spoiling ON the signal is also low for shorter TRs. However,
the phase SNR for
short TRs does
not greatly differ between RF spoiling ON or OFF,
because it is already low due to the even
shorter TEs (Fig.1B). Consequently, the
efficiency for scanning the WM phase shows
no substantial
difference between RF spoiling ON or OFF (Fig.1C). In both cases
the most efficient TR is 29ms. The corresponding FAs shown in Fig.1D are rounded
to the closest 0.5° FA increment of the scanner with FA=12.5° for
RF spoiling ON and FA=15.5° for RF spoiling OFF.
Signal difference maps between the
scans with different FAs, but RF spoil ON, (Fig.2,top) show up to
30% signal increase for cerebellar and temporal regions. These
regions typically suffer from low B1 with similar (single transmit, multi receive) head coils at 7T. The centre of
the brain, which has higher B1 in this setting,
shows signal loss due to increased saturation caused by the
increased FA. The comparison between the two scans with RF
spoiling ON/OFF, but equal FAs, shows an overall signal gain which is very strong for CSF (Fig.2,middle).
CSF benefits especially from switching RF spoiling OFF, due to its
long T2.15 Increasing the FA for RF
spoiling OFF shows overall more signal losses than gains (Fig.2,bottom).
R2* maps show some visual
differences most likely caused by subject motion during data acquisition
(Fig.3). The QSM maps appear even more similar between the different
scanning options (Fig.4).DISCUSSION & CONCLUSION
QSM
at 7T for assessing WM microstructure has maximum scan efficiency using RF spoiling OFF, TR=29ms and
FA=15.5°.
Increasing the flip angle to compensate for low B1 regions did not improve the overall signal
magnitude of these optimal scan parameters.
The
different scan options do not obviously impact the R2*
and QSM maps.
However, a further quantitative analysis of these
maps and more scanned subjects may show subtle differences. Based on the demonstrated signal increase it is advantageous to acquire QSM using the found optimal scanning parameters.Acknowledgements
This work was supported by Wellcome Trust Collaboration in Science
Award 201526/Z/16/Z, the Wellcome EPSRC Centre for Medical
Engineering at Kings College London (WT 203148/Z/16/Z) and by the
National Institute for Health Research (NIHR) Biomedical Research
Centre based at Guy’s and St Thomas’ NHS Foundation Trust and
King’s College London. The views expressed are those of the authors
and not necessarily those of the NHS, the NIHR or the Department of
Health.References
1. Marques JP, Khabipova D, Gruetter
R. Studying cyto and myeloarchitecture of the human cortex at
ultra-high field with quantitative imaging: R1, R2* and magnetic
susceptibility. Neuroimage. 2017 Feb 15;147:152-163.
2. Abduljalil AM, Schmalbrock P, Novak
V, Chakeres DW. Enhanced gray and white matter contrast of phase
susceptibility-weighted images in ultra-high-field magnetic
resonance imaging. J Magn Reson Imaging. 2003 Sep;18(3):284-90.
3. Deistung A, Rauscher A, Sedlacik J,
Stadler J, Witoszynskyj S, Reichenbach JR. Susceptibility weighted
imaging at ultra high magnetic field strengths: theoretical
considerations and experimental results. Magn Reson Med. 2008
Nov;60(5):1155-68.
4. Ernst, R. R., and W. A. Anderson.
1966. “Application of Fourier Transform Spectroscopy to Magnetic
Resonance.” Review of Scientific Instruments 37 (1):
93–102.
5. Hänicke, Wolfgang, and Horst U.
Vogel. 2003. “An Analytical Solution for the SSFP Signal in MRI.”
Magnetic Resonance in Medicine 49 (4): 771–75.
6. Rooney, William D., Glyn Johnson,
Xin Li, Eric R. Cohen, Seong-Gi Kim, Kamil Ugurbil, and Charles S.
Springer. 2007. “Magnetic Field and Tissue Dependencies of Human
Brain Longitudinal 1H2O Relaxation in Vivo.” Magnetic Resonance in
Medicine 57 (2): 308–18.
7. Yacoub, Essa, Timothy Q. Duong,
Pierre-Francois Van De Moortele, Martin Lindquist, Gregor Adriany,
Seong-Gi Kim, Kâmil Uğurbil, and Xiaoping Hu. 2003. “Spin-Echo
FMRI in Humans Using High Spatial Resolutions and High Magnetic
Fields.” Magnetic Resonance in Medicine 49 (4): 655–64.
8. Peters, Andrew M., Matthew J.
Brookes, Frank G. Hoogenraad, Penny A. Gowland, Susan T. Francis,
Peter G. Morris, and Richard Bowtell. 2007. “T2* Measurements in
Human Brain at 1.5, 3 and 7 T.” Magnetic Resonance Imaging 25 (6):
748–53.
9. Breuer FA, Blaimer M, Mueller MF,
et al. 2006. “Controlled aliasing in volumetric parallel imaging
(2D-CAIPIRINHA).” Magnetic Resonance in Medicine 55: 549–556.
10. M. Jenkinson and S.M. Smith. A
global optimisation method for robust affine registration of brain
images. Medical Image Analysis, 5(2):143-156, 2001.
11. S.M. Smith. Fast robust automated
brain extraction. Human Brain Mapping, 17(3):143-155, November 2002.
12. Liu T, Wisnieff C, Lou M, Chen W,
Spincemaille P, Wang Y. Nonlinear formulation of the magnetic field
to source relationship for robust quantitative susceptibility
mapping. Magn Reson Med. 2013 Feb;69(2):467-76.
13. Schweser F, Deistung A, Lehr BW,
Reichenbach JR. Quantitative imaging of intrinsic magnetic tissue
properties using MRI signal phase: an approach to in vivo brain iron
metabolism? Neuroimage. 2011 Feb 14;54(4):2789-807.
14. Karsa, A, Punwani, S, Shmueli, K, et al. An optimized and highly
repeatable MRI acquisition and processing pipeline for quantitative
susceptibility mapping in the head‐and‐neck region. Magn
Reson Med. 2020; 84: 3206– 3222.
15. Daoust, A., S. Dodd, G. Nair, N.
Bouraoud, S. Jacobson, S. Walbridge, D. S. Reich, and A. Koretsky.
2017. “Transverse Relaxation of Cerebrospinal Fluid Depends on
Glucose Concentration.” Magnetic Resonance Imaging 44:
72–81.