Tsuyoshi Matsuda1, Ikuko Uwano1, Yuji Iwadate2, and Makoto Sasaki1
1Division of Ultrahigh Field MRI, Institute for Biomedical Sciences, Iwate Medical University, Iwate, Japan, 2Global MR Applications and Workflow, GE Healthcare Japan, Hino, Japan
Synopsis
Flip angles (FAs) might
be underestimated during the actual FA imaging (AFI) method due to the aliasing
phenomenon that occurs when the actual FAs exceeds 90°. To optimize nominal FA
values for the AFI scan performed at 7 T, at which in-plane FA values markedly
vary, we attempted to detect pixels showing erroneous FAs by comparing the
phase difference between two types of AFI source images. AFI-FA maps with
nominal FAs of ≥ 60° include substantial areas with FA underestimation, which
is unsuitable for accurate FA measurements.
Introduction
The actual flip angle
imaging (AFI) 1, which consists of images with two types of
repetition times (TRs), i.e., TR1 and TR2, can rapidly measure distribution of
flip angles (FAs) on MR images with high accuracy, particularly when using the
highest nominal FA values. However, when the value of actual FAs exceeds 90° on
the TR1 and/or TR2 images, the FAs might be underestimated as longitudinal
magnetization reduces inversely when the value of FAs exceeds 90° (an aliasing
phenomenon), indicating that optimization of the nominal FA is crucial for
accurate FA mapping while using the AFI method. The optimization appears
difficult particularly at 7 T because the FA distribution within images
markedly vary mainly due to marked B1+ heterogeneity 2. Hence, in this
study we attempt to optimize the nominal FA value for the AFI method at 7 T by
detecting the aliasing phenomenon using phase difference information of the TR1
and TR2 images.Methods
We
used a 7 T MRI scanner (Discovery MR950; GE Healthcare, Waukesha, WI, USA) with
a two-channel transmission and 32-channel receive head coil system
(NM008-32-7GE-MR950; Nova Medical, Wilmington, MA, USA). This study was
approved by the Institutional Ethics Committee (H25-53).
Phantom
experiment and volunteer scan
We
performed three dimensional (3D) AFI scans of a 16 cm sphere phantom which
contained agarose with a T1 value of 1591 ms with nominal FA values of 5 to 130° (Table 1). In addition, to determine the truth FA values,
we performed time-consuming 3D variable FA (VFA) scans with nominal values of FA
of 5 to 90° (Table 1). Further, one healthy
subject underwent a 3D AFI head scan with nominal values of FA of 10 to 90° (Table 1) after obtaining a written consent.
Image
Analyses
We generated FA maps of the
phantom from the data obtained using the AFI and VFA methods. In addition,
absolute differences in phase between the TR1 and TR2 images of the AFI method
were calculated. Then, we analyzed the relationships between FA and
phase-difference values of the AFI method with the FA values of the VFA method
and determined the threshold of the phase difference causing the aliasing
phenomenon. In addition, using the threshold value obtained, we generated error
maps that indicate pixels having aliased or unmeasurable FA values. Furthermore,
we generated FA maps of the volunteer head from the AFI data and obtained error
maps using the threshold of the phase difference.Results
In the
phantom study, the aliasing phenomenon was observed in FA maps by the AFI
method when the true FA value using the VFA method was more than 90° (Fig. 1).
In addition, the differences in phase between the TR1 and TR2 increased with the
true FA values. Furthermore, we found that the half value of the maximum phase
difference, i.e., 1.6 rad, can be used as a threshold for predicting the
aliasing phenomenon (Fig. 1).
On the FA
maps of the phantom, the FA distribution obtained by the AFI method was
remarkably different from that by the VFA method (Fig. 2). On the error maps generated by applying 1.6 rad as the threshold
of the phase difference, a small aliased FA area was found at the center
of the image with the nominal FA of 50°. In addition, substantial aliased or
unmeasurable FA areas were observed when the nominal FAs were 90 to 130° (Fig.
2).
On the AFI-FA maps of the
volunteer, the aliasing phenomenon was visually evident when the nominal FAs exceeded
70° (Fig. 3). In addition, the aliased FA areas were detected on the error maps
when the nominal FA values were 60° or greater (Fig. 3). Discussion
We successfully detected
pixels showing erroneous FAs due to the aliasing phenomenon in the AFI images
of the human head at 7 T by determining the threshold for this purpose using
the phase difference between the TR1 and TR2 of the phantom. These results revealed
that nominal FA values of ≥ 60° are unsuitable for the AFI scan, suggesting that
the highest FA values under 60° should be selected for accurate FA measurement during
the AFI method at 7 T. Optimization of the nominal FA values using this method
may contribute to the improvement of the B1+ field mapping in parallel
transmission systems as well as concentration correction of MR spectroscopy and
chemical exchange saturation transfer imaging. However, this method has a
limitation that full echo sampling is mandatory to calculate the phase
information, resulting in being sensitive to magnetic susceptibility.Conclusions
The phase difference
between the two types of AFI images can be used to detect pixels that cause the
aliasing phenomenon. This method is considered helpful to optimize nominal FAs
of the AFI method for accurate measurement of FA distribution at 7 T.Acknowledgements
This work was partly
supported by JSPS KAKENHI (18H00483).References
1. Yarnykh
VL. Actual flip-angle imaging in the pulsed steady state: A method for rapid
three-dimensional mapping of the transmitted radiofrequency field. Magn Reson
Med 2007; 57:192-200.
2. Hoult DI, Phil D.
Sensitivity and power deposition in a high-field imaging experiment. J Magn
Reson Imaging 2000; 12:46-67.