Belinda Ding1, Iulius Dragonu2, Patrick Liebig3, and Christopher T Rodgers1
1Wolfson Brain Imaging Centre, University of Cambridge, Cambridge, United Kingdom, 2Siemens Healthcare Limited, Firmley, United Kingdom, 3Siemens Healthineers, Erlangen, Germany
Synopsis
In this study, we assessed the performance of jointly optimised pTx excitation and refocusing pulses in 2D spin-echo based sequences on a 7T Terra scanner. In conventional pTx acquisitions, the excitation and refocusing pulses are designed independently based a set of fieldmaps. Here, we compared two approaches for jointly optimising the excitation and refocussing against an approach of separately optimised pTx pulses and the traditional circularly polarised pulses. We observed that pTx pulses significantly improve image quality in both phantom and in vivo acquisitions at 7T. The image quality is further improved with joint optimisation of excitation and refocusing pulses.
Introduction
With the recent release of the first clinically approved 7T
MRI scanner, we have seen increased interest in utilising ultra-high field
scanners. However, one of the main
disadvantages of scanning at ultra-high fields is the increase in RF transmit field
(B1+) inhomogeneity, resulting in spatially non-uniform
signal magnitudes across the brain, including signal dropouts in lower brain
regions1. These effects are more
severe for higher flip angle pulses such as the refocusing pulse needed in
spin-echo imaging. RF parallel transmission (pTx) has been used as one of the
techniques to mitigate these effects, but most research have been
focused on 3D non-selective kT points pulse designs2.
Here,
we aim to make an initial comparison in performance between jointly optimised
and separately optimised slice-selective pTx spokes pulses in spin-echo based
sequences.Pulse Design
Four different groups of pulses were used in both phantom
and in vivo acquisitions(Figure
1):
- Circularly polarized(CP) pulses: CP pulses were
identical to pulses played out on a traditional single transmit system, with
all 8 transmit elements playing pulses with identical shapes and amplitudes,
varied by a fixed phase shift.
- Separately optimised pTx: Both the excitation pulse and
the refocusing pulse were optimised separately, using a linear least square
optimisation method3 based on a set of
subject-specific B0 and per-channel B1+ fieldmaps.
- Matched excitation pTx: The refocusing pulse was optimised
first using the method described for (2). Bloch simulation using Cayley-Klein
parameters4, and neglecting intra-pulse
relaxation, was then performed to produce a flip-angle map. Based on the FA
map, a target excitation pattern was calculated (no longer a homogenous
excitation), from which the excitation pulse was then designed with the aim of
improving homogeneity in the final refocused signal.
- Jointly optimised pTx: Using pulses designed in (2) as
starting points, amplitude and phase weightings for each channel and spoke
across both pulses were passed to a simulated annealing global optimisation
algorithm. Parameters for simulated annealing include reanneal interval=50;
maximum iterations=400; and initial temperature=300. New points for
iterations were generated with steps of length square root of temperate, with
direction uniformly at random and temperature schedule was
InitialTemperature/ln(k), where k is the iteration number until reannealing.
Across all methods, pTx pulses had the following parameters:
3 spokes per pulse, sinc pulse shape with a TBW=2.7,
pulse duration of 2.56ms/spoke for excitation and 3.2ms/spoke for refocusing.
The spokes positions were determined using an inverse Fourier transform method.
All pulse optimization algorithms were performed in MATLAB. The CP pulses had
matching pulse parameters, except the pulse duration of the refocusing pulse
was increased to 5.12ms.
Data acquisition
All experiments were performed on a Magnetom 7 Tesla Terra
system (Siemens Healthcare, Erlangen, Germany). Signal acquisition was performed
using an 8Tx/32Rx head coil (Nova Medical).
Phantom acquisition: A uniform spherical agar phantom was
scanned with a spin-echo sequence adapted to include pTx pulses. SE images were
obtained with two different sets of FAs (excitation/refocusing FA=60/120 and
90/180) for each group of pulses. For each set of FA and pulse group, 6 TEs
ranging from 50ms to 200ms were used. A TR of 3000ms guaranteed more than
90% recovery of longitudinal magnetisation. Phantom experiments had an
isotropic resolution of 3mm. T2 maps were computed offline by
fitting to a mono-exponential decay.
In vivo acquisition: A healthy volunteer (male, age = 27 years) was scanned using a 2D SE echo-planner imaging(EPI) sequence adapted to
include pTx pulses. Single slice SE-EPI images were obtained with two different
sets of FAs (excitation/refocussing FA=70/140° and 90/180°) for each group of
pulses. Other imaging parameters included: FOV=240×240 mm2; voxel
size=0.5×0.5×3(slice)mm3; GRAPPA acceleration factor R=3 and
6/8 partial Fourier; TE=60 ms; and TR=6000 ms.Results and discussion
SE images acquired in phantom and the resultant T2
maps are shown in Figure 2.
The distributions of signal intensity and fitted T2 across the
phantom are shown in Figure
3.
Compared to CP pulses, all three pTx optimisation methods gave narrower (better)
distributions of signal intensity and fitted T2 values.
In vivo SE-EPI images are shown in Figure 4
and 5.
For the case of CP pulses, clear signal dropouts can be observed in the
temporal lobes. All three proposed pTx pulse optimisation methods are able to
recover the signal in this area. Similar image quality with excellent
homogeneity is obtained for all three pTx pulse optimization methods. Subtle
differences can be detected on closer inspection as displayed by arrows
in Figure 5. Joint optimisation of both pulses (method 4)
outperformed the other two methods by producing clearer definition in both the
pons (yellow boxes in Figure 5) and tip of the temporal pole (red
arrows in Figure 5).Conclusion
In this abstract, we have shown the importance of using
parallel transmit pulses when acquiring SE based data at 7T. Joint optimisation
of excitation and refocusing pulses improves the image quality. We believe that
these results will be applicable to a wide range of SE-based sequences which
requires slice selective excitation such as diffusion imaging and inner-volume
selected SE-MRSI. Further optimisation of the joint pulse design pipeline including
volume-by-volume optimization for whole brain imaging will allow for very
homogeneous resulting magnetisation and increased performance compared to separately
optimised pTx pulses.Acknowledgements
BD is supported by Gates Cambridge Trust. CTR is funded by a Sir Henry
Dale Fellowship from the Wellcome Trust and the Royal Society [098436/Z/12/B]. This study was funded by the NIHR Cambridge
Biomedical Research Centre and MRC Clinical Research Infrastructure Award for
7T and has also received funding from the European Union’s Horizon 2020
research and innovation programme under grant agreement No 801075.References
-
Wu
X, Auerbach EJ, Vu AT, et al. High-resolution whole-brain diffusion MRI at 7T
using radiofrequency parallel transmission. Magn Reson Med.
2018;80(5):1857-1870. doi:10.1002/mrm.27189
- Gras V, Mauconduit F, Vignaud A, et al.
Design of universal parallel-transmit refocusing k T -point pulses
and application to 3D T 2 -weighted imaging at 7T. Magn Reson Med.
2018;80(1):53-65. doi:10.1002/mrm.27001
- Setsompop K, Wald LL, Alagappan V, Gagoski
BA, Adalsteinsson E. Magnitude least squares optimization for parallel radio
frequency excitation design demonstrated at 7 Tesla with eight channels. Magn
Reson Med. 2008;59(4):908-915. doi:10.1002/mrm.21513
- Pauly J, Nishimura D, Macovski A, Roux P
Le. Parameter Relations for the Shinnar-Le Roux Selective Excitation Pulse
Design Algorithm. IEEE Trans Med Imaging. 1991;10(1):53-65.
doi:10.1109/42.75611