Haonan Zhang1, Qingwei Song1, Jiazheng Zhang2, Yishi Wang2, Renwang Pu1, Nan Zhang1, and Ailian Liu1
1Department of Radiology, the First Affiliated Hospital of Dalian Medical University, Dalian, China, 2PHILIPS——Philips Healthcare, beijing, China
Synopsis
Compared with
echo planar imaging diffusion weighted iamging (EPI-DWI), turbo spin echo
diffusion weighted imaging (TSE-DWI) can significantly reduce magnetic
sensitivity artifacts in skull base imaging. However, the longer scan time
limits its clinical promotion. The purpose of this study is to investigate the
effect of the compression sensing acceleration factor on the image quality of
TSE-DWI in the skull base area.
Introduction
Diffusion weighted imaging (DWI), as one of the most commonly used MR
functional sequences for the diagnosis of brain diseases, can non-invasively
quantify the diffusion characteristics of water molecules using ADC1.
The traditional DWI sequence uses echo planar imaging (EPI) technology, which
can complete whole brain imaging in a short time. However, due to the variation
of the magnetic field in the skull base area, it is prone to susceptibility artifacts
and misdiagnosis2. DWI imaging based on turbo spin echo sequence
(TSE) can eliminate magnetic sensitivity artifacts, but the scan time is
relatively long, which limits its clinical promotion3. Compressed
sensing (CS) can significantly shorten the scan time through sparse sampling,
while ensuring the image quality to meet the diagnostic requirements4.
The purpose of this study is to explore the effect of the compressed sensing
acceleration factor on the image quality of the fast spin echo diffusion
weighted imaging in the skull base area.Materials and methods
The
study population consisted of 24 healthy subjects(8 females, 53.4±18 years), 22
subjects with dentures(8 females, 57.8±19.7 years) and 13 subjects with skull
base lesions(8 females, 45.1±19.7 years), defined as group A, B, and C,
Respectively. All subjects underwent the EPI-DWI and TSE-DWI scan of Skull Base
on a 3.0 T MR scanner (Ingenia CX, Philips Healthcare, Best, the Netherlands). TSE-DWI
adopted a routine clinical setup of SENSE acceleration factor 2.6, and CS
factors of 2.6, 3.6, 5.6 and 7.6. Other scan parameters were shown in Table 1.
Regions of interest were placed manually at both sides of the cerebellar
hemisphere or unilateral lesion by two experienced radiologists for the measurement
of ADC. The two observers used a five-point scoring method to evaluate the
quality of the images (the scoring system was listed in Table 2). The Kappa
test was adopted to evaluate the consistency of the scores from the two radiologists. If the consistency
was in good agreement, the corresponding images would be adopted for further analysis
by senior physicians. In the following analysis, the Kruskal-Wallis test was
used to assess the difference of ADC and score between groups, and the
Mann-Whitney U test was employed to make a pairwise comparison. This study has
been approved by the local IRB.Results
The scores by
the two observers were in good agreement (p=0.509).
In Table 3, it was shown that there were statistically significant differences
in ADC-L and score of group A, ADC-L, ADC-R and score of group B, scores of group
C. In Table 4, the ADC-L,
score of group
A and B were significantly different from the conventional sequences
when CS factor was
equal to 7.6, and scores of group B and C were significantly different from the corresponding conventional
sequences when CS factor was equal to 2.6, 3.6. However,
when CS=2.6, 3.6, the subjective score is higher than the conventional sequence.Discussion and Conclusions
Scan
time for the skull base TSE-DWI could
be reduced significantly by CS technique. However, if a large CS acceleration factor (≥7.6) is used, the image quality
degrades evidently. CS acceleration factor of 5.6 is recommended for clinical skull base TSE-DWI to achieve an optimal
balance between imaging time and image quality.Acknowledgements
No acknowledgement found.References
1. Kolff-Gart, AS; Pouwels, PJ; Noij, DP, et al.
Diffusion-weighted imaging of the head and neck in healthy subjects:
reproducibility of ADC values in different MRI systems and repeat sessions.AJNR
Am J Neuroradiol,2015,36(2):384-390.
2. Wan Q, Lei Q, Wang P, et al. Intravoxel Incoherent Motion
Diffusion-Weighted Imaging of Lung Cancer: Comparison Between Turbo Spin-Echo
and Echo-Planar Imaging. J Comput Assist Tomogr,2020,44(3):334-340.
3. Elefante, A; Cavaliere, M; Russo, C, et al. Diffusion
weighted MR imaging of primary and recurrent middle ear cholesteatoma: an assessment by readers with different
expertise. Biomed Res Int,2015,2015:597896.
4. Bratke, G; Rau, R; Weiss, K, et al. Accelerated
MRI of the Lumbar Spine Using Compressed Sensing: Quality and Efficiency. J
Magn Reson Imaging,2019,49(7):e164-e175.