Nikolaos Panagiotopoulos1, Thekla Helene Oechtering1, David Rutkowski2, Jean Brittain2, Diego Hernando1,3, and Scott B Reeder1,3,4
1Department of Radiology, University of Wisconsin-Madison, Madison, WI, United States, 2Calimetrix LLC, Madison, WI, United States, 3Department of Medical Physics, University of Wisconsin-Madison, Madison, WI, United States, 4Departments of Biomedical Engineering, Medicine, and Emergency Medicine, University of Wisconsin-Madison, Madison, WI, United States
Synopsis
A phantom placed alongside a patient, i.e.,
“pocket phantom,” can enable quality assurance of proton density fat fraction
(PDFF) measurements. To determine the optimal positioning, the spatial
dependence of PDFF measurements must be understood. To address this unmet need,
we filled the bore of a 3T MRI with pocket phantoms. Though PDFF bias was low, it
increased with distance from isocenter. Absolute bias was <2.5% within 18cm (in-plane)
and within 10cm (z-direction). We conclude that the optimal position for a PDFF
pocket phantom is on the table, centered below the patient at the level of the examined
organ.
Introduction
Quality assurance (QA) of MR acquisitions is an important
aspect of medical imaging. With the introduction of quantitative MRI into
clinical routine, QA must go beyond qualitative assessment. In the case of proton
density fat fraction (PDFF) measurements using advanced chemical shift encoded MRI
(CSE-MRI), the use of standardized “pocket” phantoms with predefined PDFF
values in a clinical scan would allow for QA on a patient-to-patient basis.
With the patient positioned at isocenter, the
pocket phantom position will necessarily be off-isocenter. Currently, the
impact of off-isocenter PDFF-quantification is an important gap in knowledge. Thus,
the purpose of this study is to establish the spatial dependence of PDFF
measurement bias using CSE-MRI.Methods
Phantom: Commercially
available sets of vials-only pocket PDFF phantoms (Calimetrix, Madison, WI)
were used. Each vial set contains 5 vials (each 25x25x91mm) with 0%,10%,20%,30%, and 40% PDFF concentration, respectively. 28 sets (140 vials) were stacked
to fill the bore (Figure 1).
Acquisition and Reconstruction: Acquisitions were performed on a clinical 3T MRI system (SIGNATM
Premier, GE Healthcare, Waukesha, WI) with a 30-channel anterior array receiver
coil using a commercial CSE-MRI method (IDEAL IQ, GE Healthcare) for PDFF
quantification (TR/TE1/ΔTE = 9.5/1.1/0.9ms, flip angle=4°, pixel
bandwidth=488Hz/pixel, FOV=48x38x8.8cm2, spatial resolution
3.0x3.0x3.0mm3). Acquisitions were repeated with different z-direction
offsets (0cm,10cm,15cm,20cm). Subsequently, the whole phantom setup was rotated
180° around the y-axis and acquisitions repeated to facilitate measurement of different
PDFF concentrations at the same locations. To avoid swap artifacts, confounder-corrected
PDFF was reconstructed using a magnitude-based reconstruction where the
iterative parameter estimation algorithm is initialized with the assumption
that PDFF=0%. The confounding effects of T2* decay, spectral complexity of fat and
temperature were also corrected1.
PDFF measurements:
Regions of interest (ROI, 1cm²) were placed in the center of all vials using
HorosTM (Horos Project, USA). ROIs were semi-automatically
co-registered on three consecutive slices and measurements averaged.
Statistics: Absolute
and relative bias between the true PDFF of the vials and the PDFF measurements
for each ROI were calculated and color-coded bias maps were created. Bias was
plotted against the distance from isocenter in-plane for all z-offsets. Bias per
PDFF value as a function of z-offset was depicted in boxplots.Results
Bias maps: PDFF
measurement bias increased with distance from isocenter in-plane and along the
z-axis (Figure 2). The absolute measurement bias was <2.5% for 81% of vials with
no z-offset and 84% for 10cm z-offset. This value dropped to 71% and 15% with increasing
z-offset of 15cm and 20cm, respectively. For all z-offsets, high bias was
recorded in the upper edges of the FOV.
In-plane distance from isocenter: Absolute PDFF measurement bias was <2.5% (<5%) for in-plane
distance from isocenter of <18cm (<21cm) with no offset in z-direction. For
distances >21cm, bias increased steeply (Figure 3). The same trend was
observed for higher z-offsets but occurred closer to isocenter in-plane (bias <5%
at <18cm, 17cm, 0cm in-plane distance to isocenter for z-offset of 10cm, 15cm, and 20cm, respectively).
Distance from isocenter along z-axis: Absolute PDFF measurement bias was <2.5% (<5%) for offsets in
z-direction of <10cm (<15cm) when measurements within a 17cm in-plane
radius from isocenter were made. With 20cm z-offset, the bias increased steeply
(Figure 4).Discussion
We examined the effects of off-isocenter
imaging on bias in PDFF measurements using CSE-MRI. Overall, bias was very low
but increased with increasing distance from isocenter. We found that a distance
from isocenter of up to ±18cm in-plane and up to ±10cm in z-direction translates
into clinically acceptable PDFF measurement bias of <2.5%. Practically, this
means that pocket PDFF phantoms should be placed under the patient, centered in
left-right direction. Further, care must be taken to align the pocket phantom
within ±10cm of the examined organ in z-direction.
Although small, bias related to off-isocenter
imaging may be related to several factors. Concomitant gradients are known to
impact PDFF estimation, depending on the characteristics of the echo train2. B0
inhomogeneities are naturally accounted for with CSE-MRI which naturally estimates
them as part of PDFF estimation. B1 inhomogeneities are generally avoided through
low flip angles. Spatially dependent eddy currents or gradient timing delays
could potentially introduce bias in PDFF estimation, although these should be
avoided using magnitude-based CSE-MRI fitting.
With between-scan variability of PDFF being as low as
1%3 and cut-offs for the prediction of metabolic syndrome
ranging between 3-5%4,5, a low bias for low PDFF values (0-10%) is essential.
For these PDFF values, we demonstrated that the absolute bias for vials within
the range of 12-15cm distance to isocenter in-plane for a z-offset of 0-10cm was
<1%. Hence, the spatial constraints of low PDFF measurements are narrower
than for higher PDFF. Finally, we note that these results are study-specific and
may depend on specific sequence parameters and hardware. Further work with a
wider variety of acquisition and reconstruction parameters is needed.Conclusion
Based on this study, the optimal location for
placing a pocket PDFF phantom lies within 18cm distance to isocenter in-plane
and 10cm distance to isocenter in z-direction allowing for maximum absolute bias
of 2.5%. We therefore propose to place the phantom below the patient centered on
the table and at the level of the examined organ.Acknowledgements
We wish to thank Calimetrix for use of the pocket
phantoms for this study. Further, we wish to acknowledge support from the NIH (NIH R01 DK088925, R44EB025729), UW Institute for Clinical and
Translational Research, and the Clinical and Translational Science Award of the
NCATS/NIH, as well as GE Healthcare who provides research support to the
University of Wisconsin. Finally, Dr. Reeder is a Romnes Faculty Fellow, and
has received an award provided by the University of Wisconsin-Madison Office of
the Vice Chancellor for Research and Graduate Education with funding from the
Wisconsin Alumni Research Foundation.
David Rutkowski, PhD, and Jean Brittain, PhD,
are employees of Calimetrix LLC. Scott Reeder, MD, PhD, Diego Hernando, PhD,
and Jean Brittain, PhD have ownership interests in Calimetrix LLC.
References
1. Reeder SB, Cruite I,
Hamilton G, Sirlin CB. Quantitative Assessment of Liver Fat with Magnetic
Resonance Imaging and Spectroscopy. J. Magn. Reson. Imaging 2011;34:729-49.
2.Colgan TJ, Hernando D,
Sharma SD, Reeder SB. The effects of concomitant gradients on chemical shift
encoded MRI. Magn. Reson. Med. 2017;78:730-8.
3. Colgan TJ, Van Pay AJ,
Sharma SD, Mao L, Reeder SB. Diurnal Variation of Proton Density Fat Fraction
in the Liver Using Quantitative Chemical Shift Encoded MRI. J. Magn. Reson. Imaging 2020;51:407-14.
4. Puchner SB, Lu MT,
Mayrhofer T, et al. High-risk coronary plaque at coronary CT angiography is
associated with nonalcoholic fatty liver disease, independent of coronary
plaque and stenosis burden: results from the ROMICAT II trial. Radiology
2015;274:693-701.
5. Rehm
JL, Wolfgram PM, Hernando D, Eickhoff JC, Allen DB, Reeder SB. Proton density
fat-fraction is an accurate biomarker of hepatic steatosis in adolescent girls
and young women. Eur. Radiol. 2015;25:2921-30.