Mo Kadbi1, Dawn Berkeley1, Brian Tymkiw1, Hung Do1, and Erin Kelly1
1Canon Medical System USA, Tustin, CA, United States
Synopsis
In MR physics, there is a fundamental tradeoff between image
spatial resolution, signal to noise ratio, and scan time. To acquire images
with high resolution and SNR, signal averaging is the most common solution, but
results in longer scan time. In this study, a Deep Learning Reconstruction
method was employed to remove the noise from clinical images and improve SNR.
This SNR improvement was devoted to increase the spatial resolution without the
need of signal averaging and increased scan time. Hence, higher resolution
images with high image quality can be obtained in shorter time in clinical
practice.
Introduction
In MR physics, there
is a tradeoff between spatial resolution, Signal-to-Noise Ratio (SNR), and
acquisition time. Increasing spatial resolution is desirable as it helps the
clinicians visualize and identify small anatomical structures and pathologies. Higher
spatial resolution negatively impacts SNR as there is less signal in a smaller voxel.
Typically, to achieve high resolution without SNR drop, the acquisition is
repeated multiple times and averaged. This repetition increases the scan time leading
to patient discomfort and potential image artifacts due to patient motion.
With Artificial
Intelligence (AI) evolution, Deep Learning Reconstruction (DLR) methods are starting
to play an essential role in MR imaging. Recently, commercial DLRs have
been used to achieve higher resolution and SNR images1,2. By selectively
removing the noise, the DLR improves the SNR without repeating the
acquisition. Increasing SNR can be utilized in different ways: it can be used
to improve the image resolution for better diagnosis; or to reduce the number
of signal acquisition (NAQ) to achieve shorter scan time; or a combination of
both. In this study, the DLR was used in
multiple anatomical regions to achieve high spatial resolution in shorter scan
time while preserving SNR and image contrast. Methods
Routine clinical brain and knee scans were performed on volunteers on a Canon
Vantage Orian 1.5T scanner. The scans were acquired using two protocols: a
routine clinical protocol with a conventional reconstruction (CR) method using
typical clinical post-processing filters to provide images with optimal smoothness
and sharpness at the same time, and a modified protocol to achieve higher
spatial resolution in a shorter scan time using the DLR method. Most sequence
parameters were kept identical between the CR and DLR protocols. Wherever the
DLR scan had an abundant SNR, spatial resolution was increased, and the NAQ was
reduced in order to shorten the scan time. The altering parameters between two the
protocols for each anatomical region and clinical sequence are shown in tables 1
and 2.
To evaluate the impact of the DLR on image quality in brain,
the SNR was measured in White Matter (WM) and Gray Matter (GM) and the contrast
change between WM and GM was calculated in axial T1, T2, and FLAIR scans.
Similarly, in knee, the SNR was measured in bone and muscle and the contrast
change between bone and muscle was calculated for each sequence. Results
Figure 1
demonstrates a brain case acquired with the CR and DLR methods. As shown in
table 1, the spatial resolutions in the DLR sequences were increased up to 37%.
The scan time for each sequence was shorter than the CR sequences by reducing
the NAQ in the DLR sequences, even though resolution was greater. The overall
exam time was reduced 31.5% using DLR scans. The SNR was measured in GM and WM
in several ROIs and averaged for each tissue type. As shown in table 3, the SNR
using the DLR scan with higher resolution and shorter scan time was increased
for all image contrasts. The contrast change between GM and WM was less than 3%
for all sequences.
Similar to the brain
case, figure 2 shows a knee case where the spatial resolution was improved in
DLR scan while the scan time for each sequence was reduced using only one NAQ.
The overall scan time using the DLR was reduced 36% compared to the CR method. The
measured SNR averaged in selected ROIs in femoral bone and muscle remained very
similar between the CR and DLR scans in axial PD and coronal T2 scans while it slightly
improved in bone in sagittal PD using the DLR scan. The contrast change between
bone and muscle was less than 8% between the CR and DLR scans. Discussions
As shown in figures
1 and 2, the DLR based images achieve a similar image quality to the CR images.
In the brain scan, the spatial resolution of sub-millimeter was achieved with the
DLR scan which resulted in sharper images. The higher resolution images were
obtained while the SNR was improved and the scan time was reduced, contrary to MR
Physics and conventional MR imaging techniques. The total routine brain scan
with high resolution images was completed in less than 10 minutes. This can be very
beneficial for clinical diagnosis as the shorter total scan time can provide
more patient comfort and reduce patient motion and artifact. In addition,
similar image quality with even higher resolution could improve the detection
of small lesions in the brain.
Similarly, in the
knee scan, the spatial resolution was improved with DLR in all the sequences
while the image quality, SNR and image contrast are similar to the CR scans. The
time saving for the entire exam was 36% which can be advantageous for the
musculoskeletal imaging wherein patients, usually experiencing severe pain,
could benefit from the shorter scan time. Conclusion
DLR facilitates higher
resolution imaging in a shorter scan time while the clinical image quality is
not compromised. The higher resolution with DLR scans results in sharper images
which can be helpful to identify smaller structural details in the images and
potentially lead to more reliable diagnosis and improved patient care.Acknowledgements
No acknowledgement found.References
1 Kidoh M, et al. Deep Learning Based Noise Reduction for Brain
MR Imaging: Tests on Phantoms and Healthy Volunteers. Magn Reson Med Sci. 2019;
doi:10.2463/mrms.mp.20190018.
2 Argentieri E, et al. Performance
of a Deep Learning-Based MR Reconstruction Algorithm for the Evaluation of
Peripheral Nerves. In Radiological Society of North America. 2019 Scientific
Assembly and Annual Meeting; 2019.