George Hutchinson1 and Penny Gowland1
1Sir Peter Mansfield Imaging Centre, University of Nottingham, Nottingham, United Kingdom
Synopsis
Analysing diffusion data from the placenta using with biexponential Intravoxel incoherent motion model and an Inverse Laplace approach gave similar final measures for the IVIM fraction and Kurtosis but not for D.
Introduction
A
biexponential diffusion model is commonly used to fit diffusion data in the
placenta and other organs. One limitation of this approach is the need to predetermining
the number of exponents to be fit for: commonly in the placenta a two-component
bi-exponential decay is assumed, but this is not necessarily correct in such a
complex organ. The goal of this abstract is to compare the fitting diffusion
weighted placental images using the IVIM and an ILT approaches.Method
This study was approved by the local ethics committee and 8
women gave informed consent to participate. Data was acquired on a Philips 3T
Achieva system using a respiratory gated pulse gradient spin echo (PGSE) EPI
sequence at 19 b-values (0-500 s/mm2), with the women in the left or right decubitus position to
avoid vena cava compression. The placenta was masked by hand into four ROIs (chorionic
plate, placenta, basal plate, and uterine wall) on the b=0 image with regions
of very low signal thresholded out as they could not be fit properly. Data
was fit using an L2
regularised inverse Laplace model (Figure 1A), where the coefficient of the L2
term was selected by inspecting data using an L-curve [1] method
and picking a suitable value to use over all subjects. 80 diffusion
coefficients were evenly spaced through log-space from 10-1 to 103 (x10-3mm2/s) to be used for the final weightings from the ILT. The ILT
could not fit negative weightings and had no boundary conditions. The weights
over each ILT were normalised to sum to 1.
Data was then fit voxelwise using non-linear least squares to a biexponential
IVIM model with kurtosis included, $$$S=S_0[(1-f_{ivim})e^{-bD+(1/6)b^2D^2K}+f_{ivim}e^{-bD^*}$$$ limiting fit parameters to plausible
biological values (Figure 1B).
We selected various measures from the ILT
to compare to the IVIM model:
IVIM comparison:
The ILTs were then analysed using a similar method to Slator et al.[2], splitting
the ILT into 3 sections (Figure 1A), with the section boundaries altered to better
correspond to our data. The normalised weights were summed over each section as
a measure of how much weighting each region had in a particular voxel. The normalised
weights averaged over each section of the ILT was plotted against the
corresponding mean IVIM value for each ROI (placenta etc) and each participant.
K comparison:
The value of the ILT at D = 10-1 mm2/s was averaged over all voxels for each
ROI for each subject and this was compared to the corresponding value of K,
since kurtosis would be expected to correspond to very low diffusion
coefficients in the ILT.
D comparison:
A peak finding algorithm was used to find the location of every peak in the ILT
for each voxel, and histograms of these peak positions were found over each ROI
(Figure 2). The mode peak D for each section was then found for each ROI and each
participant, and was plotted against the mean D from the IVIM fit.Results
Figure 3 shows the summed weightings across
sections 2 and 3 of the ILT both increase with increasing $$$f_{ivim}$$$.
Figure 4 shows that the mean of the first
ILT weighting (at D = 10-1 (x10-3 mm2/s)) shows positive correlation with K.
Histograms of the positions of peaks
within the ILT (figure 4) all contain a large peak within section 1 with a
well-defined mode. This mode is plotted against the average diffusion
coefficient from the biexponential fit in figure 5. The values are comparable but
are typically larger from the ILT.Discussion
fIVIM from the IVIM model is strongly correlated
with the weightings of regions 2 and 3 from the ILT of the data for all
placental ROIs. This is expected as an increase in weighting in regions 2 and 3
correspond to regions that have a higher proportion of spins travelling faster
than the diffusion coefficient of a voxel.
Kurtosis data from the IVIM model also correlated
well with the first value of the ILT. As the ILT has no way to compensate for
Kurtosis.
However the ILT tends to estimate a
higher diffusion coefficient than the IVIM fit and the results are not well
correlated. It is possible that selecting the mode of the peaks from the ILT
histograms skews the results.
These results suggest that similar
information is available using both ILT and IVIM analysis of the placental.
However the data from the chorionic plate always correlated differently
compared to the other ROIs, possibly related to the faster flows in that area
and the estimates of D did not agree well. This suggests that there may be
slightly different information available from the ILT. Future work will aim at
finding additional summary measures from the ILT that can provide additional
information over traditional IVIM analysis.Acknowledgements
This work was funded by the National Institute of Health and the EPSRC/MRC Oxford Nottingham Biomedical Imaging CDT. References
[1] - Hansen P. The l-curve and its use in the numerical treatment
of inverse problems. Computational Inverse Problems in Electrocardiology,
4:119–142, 2001.
[2] - Slator, PJ, Hutter, J, Palombo, M. Combined diffusionârelaxometry MRI to identify dysfunction in the human placenta. Magn Reson Med. 2019; 82: 95– 106. https://doi.org/10.1002/mrm.27733