Shujun Lin1, Bradley Sutton2, Richard Magin1, Aaron Anderson2, and Dieter Klatt1
1Richard and Loan Hill Department of Bioengineering, University of Illinois at Chicago, Chicago, IL, United States, 2Beckman Institute, University of Illinois at Urbana-Champaign, Urbana-Champaign, IL, United States
Synopsis
Simultaneous acquisition of diffusion tensor imaging (DTI) and magnetic
resonance elastography (MRE) has been proven feasible in a preliminary study of in
vivo
human brain. However, the
experimental parameters have to be optimized in order to prevent mutual
interferences of DTI and MRE acquisitions. We identified in simulations three
experimental parameter sets for in vivo
human brain DTI-MRE that we classify as good, moderate and poor with regard to
optimization and present a pilot study using these parameter sets. The experimental
results verify the simulations as we found the best performance of DTI-MRE for
the good parameters set.
INTRODUCTION:
Magnetic Resonance
Elastography (MRE) and Diffusion Tensor Imaging (DTI) are two non-invasive MRI
techniques that use the phase and magnitude of the complex-valued MRI signal,
respectively. The simultaneous acquisition of MRE and DTI is beneficial by reducing
the scanning time by a factor of ~2 and providing immediately co-registered
images. We previously have shown the feasibility of simultaneous acquisition of
DTI and MRE, which we name DTI-MRE, on in vivo human brain.1 However,
a limitation of DTI-MRE is the tissue vibration might induce signal loss on the
magnitude images due to the intra-voxel phase dispersion (IVPD) for certain
experimental parameters. Therefore, a careful selection of experimental
parameters is needed for DTI and MRE acquisitions not to interfere with each
other. We identified optimized experimental parameter sets in simulations for in
vivo human brain DTI-MRE.1 In the presented preliminary study we
compare the performance of three experimental parameter sets we classify as
good, moderate, and poor. METHODS:
The conditions for finding
the three experimental parameter sets are a b-value within the range of 950-1000 s/mm2, an encoding efficiency of vibration
above 1.5x105 rad/m, vibration frequencies within range of 20-70
Hz, and a separation time between motion encoding gradients (MEG)
selected to be integer multiples of vibration periods.3 The three parameter sets were identified based on the
extent of vibration-induced signal loss min{R(x)}, which is listed in Table 1
and was calculated using an equation that was described previously.2
The experimental DTI-MRE
study on in vivo human brain was approved by the Institutional Review
Board (Protocol # 2018-1042) at the University of Illinois at Chicago. The
study was conducted on a 3T human scanner (Prisma, Siemens). Previously
modified single-shot, spin-echo echo planar imaging (SS-SE-EPI) sequence for
DTI-MRE was used in this study. The voxel size, number of slices, matrix size, and
TR/TE are 3x3x3 mm3, 48, 80x80 and 6000/80 ms, respectively.
Standard EPI MRE without diffusion encoding and EPI DTI acquisitions without
vibration were acquired for comparison with the DTI-MRE acquisition. Diffusion
property maps and the complex shear modulus G* were calculated as previously
described.2 The Spearman correlation
was determined voxel-wise between DTI-MRE and conventional
methods in a central slice excluding fluid-filled ventricles. RESULTS:
The simulated parameter
sets classified as good, moderate, and poor used in our experiments are displayed
in the Table 1. Figure 1 shows the reconstructed mean diffusivities (MD) of one
subject. The MD value of the good parameter set in DTI-MRE is comparable to
values from conventional DTI, and the difference in MD values decreases from good
to moderate and poor. Figure 2 displays the Spearman’s correlation coefficients
on absolute stiffness maps in a central brain slice of the same subject between
DTI-MRE and conventional MRE. A similar correlation trend is observed that
confirms the simulations. A direct comparison of reconstructed MRE and DTI
parameter maps with DTI-MRE is shown in figs. 3 and 4.DISCUSSION:
Our preliminary results
suggest that the simulations provide DTI-MRE experimental parameters optimized
for best performance of the new technique. We will conduct a study with a
larger sample size in the next months to confirm these observations. DTI-MRE
has the potential to increase the clinical acceptance of MRE and DTI by
providing fast acquisitions and tissue mechanical and diffusion property maps
that are immediately co-registered.Acknowledgements
This
work was supported by the NIH NIBIB under grant R21EB026238, Adding MRE to DTI for free. The work represents the views of the authors and not
of the NIH.References
1.
Lin S, Sutton B, Magin RL, Klatt D:
Development of in vivo human brain DTI-MRE. Proceedings of the 28th
Annual Meeting of the ISMRM , p. 3325,
virtual, 2020.
2.
Yin Z, Kearney SP, Magin RL, Klatt D.
Concurrent 3D Acquisition of Diffusion Tensor Imaging and Magnetic Resonance
Elastography Displacement Data (DTI-MRE): Theory and In Vivo Application. Magnetic
Resonance in Medicine 2017; 77(1): 273-284.
3.
Yin Z, Magin RL, Klatt D. Simultaneous MR
elastography and diffusion acquisitions: Diffusion-MRE (dMRE). Magnetic
Resonance in Medicine 2014; 71(5): 1682-1688.