W. Patricia Bandettini1, Christine Mancini2, Sujata M. Shanbhag2, Jennifer Lynn Henry2, Margaret M. Lowery2, Marcus Y. Chen2, and Adrienne E. Campbell-Washburn2
1NIH/NHLBI, Bethesda, MD, United States, 2NATIONAL INSTITUTES OF HEALTH/NHLBI, BETHESDA, MD, United States
Synopsis
Low-field MRI may offer reduced artifacts in patients with
implanted metallic devices. In this early comparison, we sought to
evaluate the appearance of artifacts seen on cardiac MR at a 0.55T versus 1.5T
for common cardiovascular devices, including sternal wires, valves, surgical
clips, and cardiac implantable electronic devices (CIEDs). For smaller discrete
implants, the artifact profile was reduced at 0.55T, as expected. Whereas, CIED
generators contributed significant artifacts at both field strengths,
illustrating the complex relationship with material properties, sequence
details and field strength. Overall, lower field may offer some advantage
for cardiac imaging of patients with common implanted devices.
INTRODUCTION
Lower field strength MRI combined with modern hardware,
acquisition methods and reconstruction methods has recently become a topic of
increasing interest for cardiac magnetic resonance (CMR) (1, 2). One area of potential
benefit is the imaging of patients with implanted metallic devices, which
continues to be a challenge for conventional CMR imaging due to severe
susceptibility artifacts. Devices can
include sternal wires, valves, surgical clips and bypass graft markers, septal
occluders, as well as cardiac implantable electronic devices (CIEDs) such as
pacemakers, defibrillators and loop recorders. Although absolute susceptibility
reduces linearly with field strength, artifact size of metallic implants
demonstrates a more complex relationship with material properties, orientation
relative to B0, and sequence type/parameters (3). In this early
observational comparison, we sought to evaluate the appearance of implanted
metal seen on CMR at a 0.55T field strength compared to 1.5T.METHODS
Subjects (n = 11) with one or more known implants underwent
imaging on both a prototype 0.55T scanner and clinical 1.5T scanner (both
MAGNETOM Aera, Siemens Healthineers, Erlangen, Germany). Artifact size was compared qualitatively between
field strengths in similar views on a steady state free precession sequence. For
the patients with right ventricular leads that could be visualized in
cross-section, measurements of lead diameter were performed. bSSFP imaging generates artifacts that are a
combination of “blooming” susceptibility artifacts and banding. Sequence
parameters were independently optimized for each field strength, and therefore
were not matched (4). Gradient recalled echo
(GRE) cine imaging was used in one patient with CIED to minimize artifact from
the generator. RESULTS
We imaged patients with bioprosthetic valves (4), sternal
wires (4), defibrillators (2), loop recorders (2), pacemakers (2), Amplatz
septal occluder (1), aortic bypass graft marker (1) and vertebral rods
(1). Implanted metallic devices caused
susceptibility artifacts visible at both field strengths, but the artifact
profile for all devices was more prominent at 1.5T, as expected. Figures 1, 2 and
3 compare artifact size for a patient with post-surgical implants (sternal
wires, bioprosthetic valve, and aortic bypass graft marker). These images
demonstrate the relative reduction in blooming artifact and banding between
1.5T and 0.55T on bSSFP imaging. For example, the metallic valve ring artifact
(Figure 3), measured in the cross section (linear dimension across the valve ring
wall), reduced from 6.1mm at 1.5T to 2.6mm at 0.55T. Figure 4 illustrate
examples CIED generators imaged with bSSFP and GRE in two different patients.
In these patients, the generator contributed significant artifacts at both
field strengths, whereas the pacing leads produced smaller artifacts at 0.55T. For
the four right ventricular lead cross-sectional measurements, lead diameters
were consistently larger at 1.5T compared to the 0.55T diameters (1.5T vs.0.55T
diameters: 8.3 vs. 6.6 mm; 6.1 vs. 2.6
mm; 12.8 vs.6 mm, and 7.0 vs. 4.4 mm). Figure
5 demonstrates lead artifacts in two patients and illustrates reduced artifact
at 0.55T and limited dephasing of blood as it flows by the leads. DISCUSSION
Most metallic and implant artifacts decreased at the
lower field strength of 0.55T, as expected; however, the artifact reduction
varied based on device specifications and material properties. Quantitative assessment of artifact size from
2D imaging for comparison between 0.55T and 1.5T is challenging and was not
pursued here. Most medical metallic implants are made from paramagnetic materials,
but some stainless steels can be saturated at clinically-relevant field
strengths (5), meaning artifact size may
not change between 0.55T and 1.5T. Notably, the dense, heavy CIED generator
artifacts created significant susceptibility artifacts on both 0.55T and 1.5T
field strengths; whereas smaller discrete implants uniformly showed less artifact
at 0.55T.CONCLUSION
These early observational findings suggest that lower
field strength CMR may lend some advantage in decreasing the susceptibility
artifact that may be seen with implanted metallic devices and CIEDs. Additional work to characterize the full
array of implanted metal artifacts is ongoing. Acknowledgements
Acknowledgements: The
authors would like to acknowledge the assistance of Siemens Healthcare in the
modification of the MRI system for operation at 0.55T under an existing
cooperative research agreement (CRADA) between NHLBI and Siemens Healthcare. We also thank Delaney McGuirt and Kendall
O’Brien for their technical support and patient care.References
REFERENCES
1. Campbell-Washburn
AE, Ramasawmy R, Restivo MC, Bhattacharya I, Basar B, Herzka DA, et al.
Opportunities in Interventional and Diagnostic Imaging by Using
High-performance Low-Field-Strength MRI. Radiology. 2019:190452.
2. Simonetti OP,
Ahmad R. Low-Field Cardiac Magnetic Resonance Imaging: A Compelling Case for
Cardiac Magnetic Resonance's Future. Circ Cardiovasc Imaging. 2017;10(6).
3. Lee M-J, Kim S,
Lee S-A, Song H-T, Huh Y-M, Kim D-H, et al. Overcoming Artifacts from Metallic
Orthopedic Implants at High-Field-Strength MR Imaging and Multi-detector CT.
RadioGraphics. 2007;27(3):791-803.
4. Bandettini WP,
Shanbhag SM, Mancini C, McGuirt DR, Kellman P, Xue H, et al. A comparison of
cine CMR imaging at 0.55 T and 1.5 T. J Cardiovasc Magn Reson. 2020;22(1):37.
5. Peeters JM, van Faassen EE, Bakker CJ. Magnetic resonance
imaging of microstructure transition in stainless steel. Magnetic resonance
imaging. 2006;24(5):663-72.