Quentin Lebret1,2, Pierre Bour1,2, Valéry Ozenne1,2, Nestór Pallares-Lupon1,2, Richard Walton1,2, and Bruno Quesson1,2
1IHU Liryc, Electrophysiology and Heart Modeling Institute, fondation Bordeaux Université, Pessac, France, 2Univ. Bordeaux, INSERM, Centre de recherche CardioThoracique de Bordeaux, U1045, Bordeaux, France
Synopsis
3D
MRI of the myocardium after an ischemic attack could provide an accurate scar
segmentation, much needed by the clinicians. However, this type of acquisition,
usually obtained by late gadolinium enhancement (LGE), is very time consuming. We
combined a wave acquisition with an inversion pulse and a variable density
Poisson undersampling strategy to accelerate 3D cardiac imaging.
Retrospectively subsampled images of a sheep heart were successfully
reconstructed with an acceleration factor of 4, opening the path to a fast high-resolution 3D LGE acquisition.
Introduction
3D electrical mapping of the heart
revealed that myocardial scar is the substrate of anarchic propagation of electrical
influx. Re-entrant activity developing in the scar tissue can lead to ventricular
fibrillation (sudden cardiac death). Incorporation of CT images, Late
Gadolinium Enhancement (LGE) MRI and segmentation of vessels was shown to help
optimizing planning of ablation therapy1. However, 3D CMR is time
consuming, and a trade-off between acquisition duration and spatial resolution
is often necessary in patients suffering for arrhythmia, for who acquisition
duration may be increased due to difficulty in gating the sequence on irregular
ECG. Parallel imaging techniques take advantage of the use of array receivers2 to reduce acquisition time by k-space subsampling. Among the techniques,
Wave-CAIPI–type sequences3 offer further aliasing control and higher
acceleration factors. In this study, we present a wave
gradient-echo sequence, containing an inversion pulse, implemented for in vivo
3D rapid scar imaging on the heart.Methods
- Pulse Sequence:
We implemented at 1.5T (MAGNETOM Aera, Siemens Healthineers) a wave-encoded
3D-GRE sequence, with a non-selective inversion pulse for T1-weighting and
respiratory navigator. 7 sinusoidal gradients (6mT/m intensity) are played
during the readout in both phase encoding directions.
In order to
calibrate the PSF, experimental trajectories are measured using 1D calibrations
scans prior to wave 3D-GRE acquisition4 (see chronogram in Fig.1).
- Animal model: A partial coronary occlusion (implantation of a
millimetric obstructive coil) of the first diagonal branch of left
anterior descending artery was induced in an anesthetised sheep (protocol approved
by ethical committee) under radiographic and ECG control. The animal was imaged
in the chronic phase 6 weeks post-myocardial infarction.
- In vivo experiment: Fully sampled wave 3D-GRE acquisitions were
performed in the thorax of an anesthetized sheep. The acquisitions were ECG
gated in cardiac diastole and a crossed pair navigator echo was used to track
and compensate respiratory motion (4 mm acceptance window). Acquisition
parameters: FoV =313*313*120mm2, 208*208*48 matrix (1.5*1.5*2.5mm3 spatial
resolution), TE/TR/FA = 2.86ms/27ms/10°, 350ms inversion time; acquisition of 40
segments per heartbeat.
- Inline Image Reconstruction: Raw data were streamed to the Gadgetron
reconstruction framework5, where calibrations and image
reconstruction were performed in a in-house written MATLAB gadget.
- Offline Image Reconstruction: The fully sampled in-vivo acquisition was
reconstructed to create a reference 3D volume. The effect of the wave gradient
was de-convoluted from the fully sampled data using the PSF formalism developed
by Bilgic and al3. The
corresponding images served as a gold standard.
The raw data were then
retrospectively subsampled and reconstructed to simulate accelerated
acquisitions (acceleration factor of 4) using a MATLAB script and the BART
toolbox6. For this, a 2x2 CAIPI mask and a x4 Poisson undersampling
mask were applied in ky-kz dimensions. Coil sensitivity maps
were estimated using ESPIRIT2 on low-resolution
wave images de-convoluted by low resolution PSFs7.
Results
The fully sampled dataset was
acquired in 12 minutes. The fully sampled image was reconstructed and sent
back to the MRI computer. This inline reconstruction pipeline offers flexibility to tune
the parameters during the experiment, especially the number of wave gradients
and their amplitude, depending on resulting image quality.
Figure 2 shows the difference between images
obtained with the subsampled k-spaces with acceleration factor of 4. Residual
ghosting is still visible on the image, as
we move away from the scanner isocenter, but they do not alter the signal in
the heart. The image subsampled with the CAIPI mask shows a RMSE of 4.1%,
higher than the one accelerated by the VDP disk. However, the infarct is
clearly visible in both images. Figure 3 displays orthogonal views of the 3D data sets for qualitative
comparison of image quality.
Discussion
Acceleration of acquisitions was performed offline to compare image quality with different subsampling strategies. Both accelerated images show acceptable RMSE
compared to the fully sampled data, yet the Poisson undersampling yields better
SNR due to the reduced coherence of the aliasing artefacts. The scar tissue observed were consistent between both the fully sampled and
undersampled images.Incorporating
Compressed Sensing in the SENSE reconstruction of the BART toolbox may further improve
image quality.Conclusion
The presented results indicate the feasibility
of rapid 3D LGE scar imaging using a navigated T1-weighted Wave-CAIPI
type sequence. The image quality after retrospective k-space subsampling by a
factor 4 was similar to the fully sampled image (albeit with lower SNR) using
either CAIPI or Poisson undersampling masks, showing potential for clinical usage.Acknowledgements
This work received financial support from the French National Investments for the Future Programs: ANR-10-IAHU-04 (IHU Liryc)References
1. Cochet
H, Dubois R, Sacher F, Derval N, Sermesant M, Hocini M, Montaudon M,
Haïssaguerre M, Laurent F, Jaïs P. Cardiac arrythmias: multimodal assessment
integrating body surface ECG mapping into cardiac imaging. Radiology.
2014;271:239–247
2. Uecker M, Lai P,
Murphy MJ, et al. ESPIRiT-an eigenvalue approach to autocalibrating parallel MRI: Where
SENSE meets GRAPPA. Magnetic Resonance in Medicine. 2014;71(3):990-1001.
3. Bilgic B, Gagoski
BA, Cauley SF, et al. Wave-CAIPI for highly accelerated 3D imaging. Magnetic Resonance in
Medicine. 2015;73(6):2152-2162.
4. Zhang Y, Hetherington HP, Stokely EM, Mason GF, Twieg DB. A novelk-space
trajectory measurement technique. Magnetic Resonance in Medicine.
1998;39(6):999-1004.
5. Hansen MS, Sørensen TS. Gadgetron: An open source framework for medical
image reconstruction: Gadgetron. Magn Reson Med. 2013;69(6):1768-1776.
6. Uecker M,
Ong F, Tamir J, Bahri D, Virtue P, Cheng J, Zhang T, and Lustig M. Berkeley
Advanced Reconstruction Toolbox. Annual Meeting ISMRM, Toronto 2015, In Proc.
Intl. Soc. Mag. Reson. Med. 23:2486
7. Chen F, Zhang T, Cheng JY, Shi X, Pauly JM, Vasanawala SS.
Autocalibrating motion-corrected wave-encoding for highly accelerated
free-breathing abdominal MRI: Wave-Encoding for Free-Breathing Abdominal MRI. Magn
Reson Med. 2017;78(5):1757-1766.