Andrew Tyler1,2, Justin Y C Lau1, Jane Ellis1, Jack J Miller1,2,3, Paul A. Bottomley4, Christopher T Rodgers1,5, Damian J Tyler1,2, and Ladislav Valkovic1,6
1Oxford Centre for Clinical Cardiac Magnetic Resonance Research, University of Oxford, Oxford, United Kingdom, 2Department of Physiology, Anatomy & Genetics, University of Oxford, Oxford, United Kingdom, 3Department of Physics, University of Oxford, Oxford, United Kingdom, 4The Division of MR Research, Johns Hopkins Medicine, Baltimore, MD, United States, 5Wolfson Brain Imaging Centre, University of Cambidge, Cambridge, United Kingdom, 6Department of Imaging Methods, Institute of Measurement Science, Slovak Academy of Sciences, Bratislava, Slovakia
Synopsis
The intrinsically low SNR and long acquisition times of 31P spectroscopy make translation to clinical practice very challenging, even at ultra-high fields. In this study, 12
healthy volunteers were scanned twice at 7T with a short 31P
CSI and reduced k-space acquisitions for reconstruction with the SLAM
and SLIM algorithms. PCr/ATP
ratio and PCr SNR were computed for each scan and coefficients of
reproducibility and variability were calculated. Compared
to a
Fourier based reconstruction of the short 31P
acquisition, SNR was significantly improved and PCr/ATP
was maintained when SLAM and SLIM reconstructions were used.
Introduction
MRS of high energy phosphorus metabolites provides a unique insight into cardiac metabolism. Of particular interest is the marked difference in PCr/ATP metabolite ratio between the healthy and diseased heart [1] which has been shown to be a significant predictor of cardiac mortality [2]. However 31P MRSI is fundamentally limited by poor SNR, leading to long scan times and poor spectral or low spatial resolution.
To reduce the number of phase encodes required for 31P-MRS several compartmentalized
spectroscopy techniques,
including SLIM[3],
SLAM[4],
SLOOP[5]
etc, have been proposed. These techniques seek to improve upon
conventional CSI by incorporating prior knowledge of the anatomical
structure of the subject into the reconstruction of spectra and/or
the optimization
of phase encode gradients.
The
fundamental assumption underlying these techniques is that the sample
can be divided up into n homogeneous compartments[3], each of
which has uniform MR properties, although the exact implementation of
SLIM/SLOOP and SLAM are markedly different. Using this knowledge a
series of simultaneous equations are constructed:
$$P=GC$$
Where $$$P$$$ is a vector of m phase encoded signals, $$$G$$$ a
phase encode operator with dimensions m x n and $$$C$$$ a
vector of n separated signals. The equation is then solved by
multiplying $$$P$$$ by the inverse of $$$G$$$ to give signals for each
compartment.
By
matching the spatial response function (SRF) to the compartment of
interest, a lower number of phase encodes can be used to achieve the
localization allowing for an increase in SNR or reduction in scan time compared to
regular CSI.
Ultra-high fields increase the achievable SNR in 31P-MRS almost 3-fold[6], however, a further increase in SNR could improve precision or reproducibility. At
7T, the B0 and B1 fields are highly
in-homogeneous and it is not yet clear if this will prevent the SLAM
and SLIM algorithms from effectively separating the spectra for each
compartment due to the underlying homogeneity assumption of the
techniques. This means that previous work at 1.5 and 3T may not be
applicable at 7T.
In
this work, we asses the feasibility of using compartmentalized
spectroscopy
techniques for cardiac 31P MRS at 7T.Methods
12
healthy volunteers (6M/6F,
BMI=23±3
kg/m2,
age=29±4
years) were scanned using a Siemens 7T Magnetom system (Erlangen,
Germany), equipped with a dual-tuned custom built RF-coil (quadrature
Tx/Rx for 31P
and 10cm loop for 1H[7]).
A short axis localizer,
matching the planned CSI acquisition was collected, followed by four
3D 31P CSI acquisitions (detailed in figure 1) to allow each of the
proposed reconstruction methods to be analysed.
While the non-optimized
acquisitions were completed (acquisition 1, 2) the matched short axis localisers were manually segmented (heart, chest wall, other) and optimized
fSLAM and SLOOP phase encode coordinates were calculated.
All CSI scans were matched to take a 6.5 minutes.
After
completion of the scan, the spectra were reconstructed using each of
the proposed methods and fitted using the OXSA Matlab toolbox AMARES
implementation[8] (figure 2).
For each fitted spectrum PCr/ATP ratios, PCr
SNR and coefficients of reproducibility (for PCr/ATP ratio) were
computed.
Wilcoxon
signed-rank paired
tests (α=0.05/comparisons, Bonferroni correction), were used
to compare the methods with the analysis of a single
midseptal
voxel, which is most commonly used for cardiac 31P CSI. A small number
of spectra were un-analysable and excluded from further analysis (1 participant with excessive movement + 3% of remaining data-points with very high liver contamination).Results and Discussion
An
example compartmentalization of one short axis slice through the
chest and the reconstructed spectra for that volunteer is shown in
figure 2. As expected the heart has a lower PCr to ATP ratio than the
chest wall and the 2,3-DPG resonance is clearly visible in the heart
spectra due to blood in the cardiac chambers.
Box-plots of PCr/ATP ratio and PCr SNR are shown in figures 3 and 4
respectively. A
table summarizing the computed coefficients of reproducibility and variation is shown in figure 5.
All compartmentalised spectroscopy techniques had a significantly higher SNR (Wilcoxon signed-rank, paired, α=0.05/6) than simple analysis of the
midseptal voxel, in line with previously reported results [4]. The
computed PCr/ATP values were within a plausible range of values[6][9],
however, the SLOOP value was significantly higher (Wilcoxon signed-rank, paired, α=0.05/6) than the midseptal value.
Unlike
the other compartmentalized
techniques, AW
SLAM and SLIM are a re-analysis of the midseptal voxel CSI acquisition, making implementation simple, however they still achieved lower (better)
coefficients
of variability and reproducibility
than the midseptal value, while maintaining
a similar PCr/ATP ratio.
Differences in the reproducibility reported here and by Ellis et al.[9] can be attributed to different RF hardware, however the relative improvement seen here suggests that an improvement would be seen with that hardware too.Conclusion
In
this study we show that compartmentalized
spectroscopy techniques can achieve a significant
increase in SNR and improved inter-scan reproducibility compared to our existing Fourier based midseptal voxel method while maintaining the PCr/ATP ratio. This suggests that the
B1
and B0
in-homogeneity present at 7T is not a significant
barrier to the use of compartmentalised MRS reconstruction techniques, with the right acquisition.Acknowledgements
This
work was supported by funding from the Engineering and Physical
Sciences Research Council (EPSRC) and Medical Research Council (MRC)
[grant number EP/L016052/1].
DJT is funded by a BHF Senior fellowship [FS/19/18/34252].
JJM
would like to acknowledge the support of a Novo Nordisk Postdoctoral
Fellowship, and the financal support of St Hugh's and Wadham College.
CTR
and LV are funded by a Sir Henry Dale Fellowship from the Wellcome
Trust [098436/Z/12/B]. Support
of the Slovak Grant Agencies VEGA [2/0003/20] and APVV [#19–0032]
is also gratefully acknowledged.References
[1]
O. J. Rider et al. Int. Journ. Card. Imag. 29:1043-1050, 2013
[2]
S. Neubauer et al. Circ. 96:2190-2196, 1997
[3]
X. Hu et al, Magn.
Reson. Med. 8, 314–322, 1988
[4]
Y. Zhang et al. J. Magn. Reson. 218, 66-76, 2012
[5]
M. Kienlin et al. J. Magn. Reson. 94, 268-287,1991
[6] C. T. Rodgers et al. Magn. Reson. Med. 72, 304-315, 2014
[7]
Schaller et al. Proc. Intl. Soc. Magn.
Reson. Med. Singapore, 2016
[8]
Purvis et al. PLoS ONE. 12(9):e0185356, 2017
[9]
J. Ellis et al. NMR in Biomed. 32:e4095, 2019