Gaspar Delso1, Laura Farre2, Daniel Lorenzatti3, Santi Sotes3, Adelina Doltra3, Susanna Prat3, Rosario J Perea3, Teresa M Caralt3, José T Ortiz3, and Marta Sitges3
1GE Healthcare, Barcelona, Spain, 2Universitat de Barcelona, Barcelona, Spain, 3Hospital Clínic de Barcelona, Barcelona, Spain
Synopsis
Cardiac
T1-mapping methods often require motion correction. The contrast changes
intrinsic to the inversion recovery series often used for this purpose can
occasionally cause registration errors, resulting in inaccurate T1 values. It
has been shown, using a large database of clinical cases, that accounting for those
contrast changes markedly increases the robustness of the correction.
A new metric
of cardiac anatomy alignment had to be defined, in order to automate the
quantitative analysis of the database. This metric was shown to correlate with
the visual scoring of misalignment in MOLLI series.
Introduction
Myocardial
tissue T1 constitutes a reliable indicator of heart diseases related to extracellular
changes (oedema, fibrosis, etc.) as well as fat, iron and amyloid content1,2.
Quantitative T1 mapping delivers repeatable, objective diagnostic criteria
based on non-invasive tissue characterization.
T1-mapping
is typically achieved by exponential fitting of a series of inversion or
saturation recovery measurements. Good anatomical alignment between these
measurements is essential for accurate T1 estimation. In practice, however,
this alignment is limited by such factors as patient motion and gating
accuracy.
Post-reconstruction
image registration is usually applied to improve alignment. However, in the
case of inversion recovery sequences3,
registration is compromised by the intrinsic contrast variation between frames.
We present the
evaluation of a dedicated motion correction method for MOLLI series, on a large
database of cardiac clinical cases. To automate the quantitative analysis of registration results,
a custom metric was defined and validated.Methods
A cohort of
186 patients (115 M / 71 F; weight 75±15 Kg; age 55±16), referred for
a clinically indicated cardiac scan, was included in this study. Scans were performed on a 3.0T GE Signa
Architect at Hospital Clínic (Barcelona). The
acquisition protocol included one or more MOLLI
sequences, with target parameters: 2D bSSFP, 160x148, pFOV 0.8-1.0, 1.4x1.4mm²,
ST 8mm, TE 1.4ms, TR 3.0ms, FA 35deg, NEX 1, BW 100kHz, 2x ASSET, 5(3s)3, Body
48.
Three
reconstruction methods were applied retrospectively, prior to T1 mapping:
Standard Cartesian reconstruction; Cartesian reconstruction with
motion correction, as implemented in the scanner; Cartesian reconstruction
combined with a new correction algorithm, based on a similarity criterion that
accounts for the intensity changes caused by T1 relaxation. The algorithm
consisted on an iterative approach, alternating polarity estimation, T1
fitting, relaxation series simulation and frame-by-frame registration.
Due to the
large number of cases being analyzed, a framework for automated quantitative
evaluation had to be implemented. For this purpose, a metric of anatomical
alignment was defined as the percentage of cardiac pixels T1 fitting
coefficient of determination greater than 95%. This metric was validated by
comparison with the qualitative scoring of 200 MOLLI series, using the
three-class data alignment scale defined in table 1.Results
All data
were successfully reconstructed, resulting in a total of
1133 MOLLI series for each method. A
representative example is shown in figure 1. Series were
independently analysed and T1 maps automatically generated.
On average, the
percentage of cardiac pixels showing good T1 fit (R²>0.95) went up from
85%±9% with standard motion correction to 90%±7% with the dedicated method. In comparison
to uncorrected series, the percentage improved by +3%±8% with standard motion
correction and by +9%±8% with the dedicated correction.
The new similarity criterion was found to limit the incidence of
registration failures, defined as those cases where motion correction decreased
alignment by 5% or more. Failures went down to 0.3% of cases with the new
method, compared to 12% with standard motion correction. The relative performance of the
different methods can be appreciated in figure 2.
The results of
the qualitative analysis are shown in figure 3. The mean value for each of the
three alignment classes was, respectively: 90%±7%, 85%±7% and 75%±11%. These
were found to be statistically significant with P<10-2. Discussion
Cardiac T1
mapping is a valuable and widely accepted diagnostic tool, but remains
sensitive to patient motion and triggering inaccuracies, being particularly
vulnerable to arrhythmia episodes. Multiple instances of misalignment leading
to partially inaccurate T1 maps were identified in the database collected for this
study.
In clinical
practice, the orientation and parameters of MOLLI acquisitions for T1 mapping
vary greatly, depending on factors such as patient condition, use of contrast
agents, etc. Similarly, image quality is strongly dependent on patient
condition and compliance. For this reason, this study aimed at evaluating our
motion correction on a large, representative set of clinical cases.
The results
of the quantitative evaluation show that accounting in the similarity criterion
for the contrast changes caused by inversion recovery markedly decreases
registration failures. The validity of the alignment
metric used for this purpose was confirmed by the results of the qualitative
analysis.Conclusion
A dedicated
motion correction method for cardiac MOLLI series has been evaluated on a large
clinical database, demonstrating increased robustness. Quantitative
measurements have been automatically extracted using a metric designed to
reflect the impact of anatomical alignment on cardiac T1 maps. This metric has
been validated by comparison with the qualitative ranking of a subset of the
available datasets.Acknowledgements
No acknowledgement found.References
1. Jellis, C. L. &
Kwon, D. H. Myocardial T1 mapping: modalities and clinical applications. Cardiovasc
Diagn Ther 4, 126–137 (2014).
2. Radenkovic, D., Weingärtner, S.,
Ricketts, L., Moon, J. C. & Captur, G. T1 mapping in cardiac MRI. Heart
Fail Rev 22, 415–430 (2017).
3. Messroghli, D. R. et al.
Modified Look-Locker inversion recovery (MOLLI) for high-resolution T1 mapping
of the heart. Magn Reson Med 52, 141–146 (2004).