Minjie Zhu1, Stephen DeVience2, and Dirk Mayer1
1University of Maryland Baltimore, Baltimore, MD, United States, 2Scalar Magnetics, LLC, Cuyahoga Falls, OH, United States
Synopsis
The
goal of this study is to incorporate the spectral dimension into the Low Rank
Plus Sparse Reconstruction algorithm with application of undersampled dynamic Hyperpolarized 13C
imaging. The
proposed method can be used to increase temporal and/or spatial resolution
without significantly compromising image quality.
Introduction
Iterative low rank plus sparse (L+S) reconstruction has
been applied to dynamic hyperpolarized (HP) 13C metabolic
imaging in previous studies to accelerate image acquisition with incoherent
dynamic sampling patterns1,2. For spectroscopic imaging, the
existing L+S reconstruction framework has robust performance by reconstructing
the dataset for each frequency point individually1. However, this requires
performing the iterative reconstruction multiple times without consideration of
the low-rank property in the frequency dimension, leading to excessive
reconstruction time and failure when the dimension of the low-rank space is too
small. In this study, we extend the L+S reconstruction framework to spiral
chemical shift imaging (spCSI) by enforcing the consistency between image
space and k-space of the undersampled 2D-spatial-1D-spectral dataset using 3D
NUFFT operator. The proposed method was evaluated with digital simulations and
retrospective study on preclinical in vivo scan.Methods
Algorithm: The L+S reconstruction fill the under-sampled k-space
through iterative soft thresholding the singular values of the low-rank image
space L and the entries of the sparsity image space S. The L+S algorithm for joint
spectral-spatial acquisition applies the partial Fourier transform of the under-sampled
kx-ky-t data d0 into 3D image space M with dimension [Nx, Ny, Nf] for each dynamic time point. This is achieved
through a modified 3D NUFFT/iNUFFT operation given the undersampled k-t
trajectory. The L+S reconstruction pipeline follows the same as previous studies1-3, shown
in Fig 1.
Simulations: The modified L+S algorithm was first tested on digitally simulated 2D
spatial-1D spectral dynamic imaging data comprising of 4 discs representing the
vasculature, kidneys, and liver/body, and 3 spectral peaks (pyruvate, alanine,
and lactate, off-resonance [0Hz, 179Hz, 391Hz] at 3T), each intensity
characterized by the typical dynamic measurement in a rat following
hyperpolarized pyruvate injection. A previously acquired in vivo rat brain imaging dataset performed
on a clinical 3T GE 750w MR scanner (GE Healthcare, Waukesha, WI, USA) was used
for retrospective undersampling simulations.
Scan parameters: For both the digital simulation
and in vivo scan, the fully sampled 2D single slice spCSI sequence has the
following parameters: FOV 4x4 cm, matrix size 16x16, 24 echoes, 8 interleaves,
280 Hz spectral width. The simulated data consisted of 24 dynamic time points
over a total scan time of 72 seconds, while the in vivo scan had 16 dynamic
time points over 48 seconds. The under-sampled data was extracted from the
fully sampled data by choosing 4 out of 8 (50% undersampling) of the
interleaves. The order of the sampled and unsampled interleaves was random for
each dynamic time point to create incoherence between the k-space and image space
for the L+S algorithm to converge.
Assessment: The
simulated and retrospectively undersampled datasets were reconstructed via the
proposed modified L+S algorithm. Metabolic images of pyruvate, lactate, and
alanine (for simulated data) were calculated by phasing the spectrum in each
voxel and integrating the resulting peak from image space M. The reconstructed
images were assessed compared to the images reconstructed from fully sampled
k-space data using normalized root mean squared error (nRMSE) and the
structural similarity index (SSIM)4. A series of combinations of
the threshold parameters λL and λS were tested to determine the optimal values
for general application.
Spectral aliasing: With the given 280Hz spectral width, the alanine and
lactate resonances are aliased, which requires an additional linear phase
correction step during reconstruction5. Two methods were implemented
and compared: 1) apply alias correction on the undersampled k-t data for each
metabolite of interest and then run the L+S algorithm to get image space M separately. 2) run the
L+S algorithm on the undersampled k-t data for only once and get estimated
fully sampled k-t data, apply alias correction for each metabolite of interest and
get image space M, respectively. Results and Discussion
Due to the limited dimension
of the low-rank space ([Nx Ny T] = [16,16,24]), reconstruction of the simulated and retrospectively
undersampled data with the previously published method (individual L+S iteration on
each frequency point) failed as the iterative algorithm did not converge. The
metabolic images reconstructed with the proposed method for the simulated
dataset at selected time points are shown in Fig 2c. An
optimal value of λL = 0.02 and λS = 0.01 were selected based on the nRMSE
and SSIM measurements. Assessment results for the two
methods to resolve aliased spectral peaks are shown in Fig 3. They lead to similar results in terms of deviation
from ground truth, while the latter only require one round of L+S iteration,
preferred for the reduction in total reconstruction time.
Results for the retrospective study on the in vivo rat
brain imaging is presented in Fig 4. Similar to the digital simulations, the L+S
algorithm can successfully reconstruct the undersampled in vivo data without
significant artifacts.Conclusion
The presented data demonstrate that the L+S image
reconstruction can be applied to undersampled 2D-spatial-1D-spectral
spCSI imaging data in a joint 3D scheme to accelerate image acquisition. By
incorporating the spectral dimension in a single L+S process, the proposed
method has less restriction on the spatial/temporal dimension of the image
matrix to make the L+S algorithm converge and significant reduction in
reconstruction time.Acknowledgements
This
work was supported by NIH grants R21 NS096575, R01 DK106395, R21 CA213020, and
R21 CA202694.References
[1] DeVience, Stephen J., Mayer, Dirk. "Speeding up dynamic spiral
chemical shift imaging with incoherent sampling and low‐rank matrix
completion." Magnetic resonance in medicine 77.3 (2017): 951-960.
[2] Milshteyn, Eugene, et al. "Using a local low rank plus sparse
reconstruction to accelerate dynamic hyperpolarized 13C imaging using the bSSFP
sequence." Journal of Magnetic Resonance 290 (2018): 46-59.
[3] Otazo, Ricardo, et al. "Low‐rank plus sparse matrix
decomposition for accelerated dynamic MRI with separation of background and
dynamic components." Magnetic resonance in medicine 73.3 (2015):
1125-1136.
[4] Wang, Zhou, et al. "Image quality assessment: from error
visibility to structural similarity." IEEE transactions on
image processing 13.4 (2004): 600-612.
[5] Mayer, Dirk, et al. “Fast Metabolic Imaging of Systems with Sparse
Spectra: Application for Hyperpolarized 13C Imaging.” Magnetic resonance in medicine
56.4 (2006): 932–937.