Xiaoxi Liu1, Shuyu Tang1,2, Xucheng Zhu1,3, and Peder E.Z. Larson1
1Radiology and Biomedical Imaging, University of California, San Francisco, San Francisco, CA, United States, 2HeartVista, Inc., Los Altos, CA, United States, 3GE Healthcare, Sunnyvale, CA, United States
Synopsis
Hyperpolarized 13C magnetic
resonance imaging is a non-invasive imaging tool to assess metabolic process in-vivo that has been most commonly applied to imaging cancer and heart disease.
Spiral readout, as a rapid acquisition technique, is commonly used to acquire
hyperpolarized metabolic data. However, the spiral readout is sensitive to the
off-resonance effect and has blurring artifacts as a result. The blurring also
affects the measurement of metabolism. In this project, we investigated two off-resonance correction methods with self-estimated field maps and applied them on different anatomies.
Introduction
Hyperpolarized 13C magnetic
resonance imaging (MRI) is a non-invasive imaging tool for assessing metabolism,
including tumor lactate production1,2 as well as metabolism profiles in the
brain3,4 and heart5,6. Considering the finite decay and non-renewable magnetization generated
by hyperpolarization, the single-shot spiral-out readout combined with spectral-spatial
RF excitations is a common efficient technique to encode the k-space data7. However,
the spiral readout is sensitive to B0 inhomogeneity, which results in image blurring8. Off-resonance correction methods have
been applied in prior hyperpolarized 13C studies using spiral readouts,
including 1)a multi-frequency
automatic field map estimation scheme8,9; 2)a model-based
reconstruction to estimate the image and field map alternatively10. In this
work, we improved the current off-resonance correction methods to self-estimate
the field map and compared the performance in HP 13C studies in the
different anatomies.
Methods
Auto-Focusing
Correction
We modified
the auto-focusing method with spiral readout for proton MRI8, and adapted
this method to Hyperpolarized 13C data. The flowchart is shown in
Figure 1(a). To begin with, the coil sensitivity maps were measured by
smoothing the reconstructed pyruvate images. And the k-space data were
demodulated at several frequencies with a suitable range to cover the estimated
B0 inhomogeneity map. The field map can be estimated by maximizing the demodulated
images ($$$I(x,y,f)$$$) with a total variation (TV) as the
criteria, as shown below:$$
f(x,y)=\underset{f}{\mathrm{argmax}}\int\int_{A(x,y)}^{}{TV(I(x',y',f))dx'dy'}
$$Here, $$$A(x,y)$$$ is a
summation window function at the location $$$(x,y)$$$. With the estimated field map, the
corrected image was pieced together pixel by pixel from demodulated images ($$$I(x,y,f)$$$).
Time-Segmented
NUFFT Correction with Self-Estimated Field Map
Considering
the spiral readout, the image ($$$m_{i}(r,t)$$$) at $$$t$$$ time frame
for $$$i$$$-th channel can be represented as $$
m_i(r,t)=m_0(r,t)C_i(r)e^{(-i2\pi f(r,t)TE)}
$$Where
$$$C_i(r)$$$ is the
coil sensitivity map of $$$i$$$-th channel, $$$f(r,t)$$$ is the
off-resonance frequency. Estimation of field map is illustrated in Figure 1(b). Each channel image ($$$m_i(r,t)$$$) of all metabolites at all time points were reconstructed
with NUFFT11, then were averaged over time points. We estimated
the phase produced by the coil by finding the phase ($$$\sigma_i$$$) with highest value from the histogram of averaged
images. The magnitude ($$$\mid C_i(r) \mid$$$) of coil sensitivity map was extracted and then
merged into the composite coil sensitivity map ($$$C_i^*(r)$$$) with the relationship$$
C_i^* (r)=|C_i (r)| e^{iσ_i}
$$The
composite images ($$$\hat{m_0}(r,t)$$$) were then derived from the composite coil
sensitivity map to estimate the field map by$$
f(r,t)=angle(\hat{m_0}(r,t))/(-2{\pi}TE)
$$Finally, corrected images were reconstructed with time-segmented NUFFT12.
Phantom
Experiments
Ethylene
glycol (EG) phantom data were acquired on a 3T GE scanner using multi-slice 2D
single-shot spiral readout to test the two self-estimated off-resonance
correction methods. Data were acquired with a clamshell transmit coil and an
8-channel paddle receive array coil. Scan parameters were FOV=58X58cm2, resolution=10.58X10.58mm2, slice thickness=100mm, TE/TR=10/500ms,
readout time=22ms, NEX=2400.
In-Vivo
Experiments
We
compared the two methods on different in-vivo anatomies: brain, heart, kidney. All
in-vivo experiments were acquired on the 3T GE scanner with the same setting as
the phantom experiment. In the in-vivo test, slice thickness=21mm, NEX=30. Other scan
parameters were
1) brain
experiments FOV = 39X39cm2, resolution = 15X15mm2, TE/TR = 10/125ms.
2)
cardiac experiments FOV = 30X30cm2, resolution = 12X12mm2, TE/TR = 10/53ms.
3) renal
experiments FOV = 75X75cm2, resolution = 15X15mm2, TE/TR = 10/100ms.
Results
Figure 2
shows reconstructed images of EG phantom. Figure 3-5 show reconstruction
results 13C pyruvate brain images, bicarbonate cardiac images and lactate renal
images, respectively. Without off-resonance correction, the images have
blurring at the edge between different materials, shown in Figure 2(a). In
Figure 2, the auto-focusing correction method shows a better apparent
correction performance with a sharper edge, indicated by the yellow arrows. However, the autofocus estimated field map is
very different from the acquired field map while the time-segmented appears
more accurate. In Figure 5, the auto-focusing correction method also shows a better
correction performance with an intensity increase than the image without
off-resonance correction, indicated by the white circles and grey arrows. In
Figure 3&4, the self-estimated field map and
time-segmented NUFFT correction method shows a better correction performance. In
Figure 3, we chose a line profile to compare the intensity of three methods.
The results of using the iterative NUFFT correction
method has an improvement of intensity. In Figure 4, only using the method(c)
provides visualize of the entire left ventricle myocardium, shown by the grey
arrows. Discussion
After
comparing the results of two correction methods applying to different
anatomies, the two methods are suitable for different regions. Considering the
time-segmented NUFFT method may amplify the noise during iterations, this
method is suitable for high SNR images. At the same time, the self-estimated
field map is based on the assumption that the coil applies a constant phase on
the image, so the estimated field map may have an offset frequency. The
auto-focusing method uses a TV criteria to find the sharpest image, so this
method is more suitable to clear structure images. And because the corrected
image is pieced together, the SNR of corrected image should be same as the
image reconstructed by the conjugate phase. In conclusion, the off-resonance
effect in Hyperpolarized 13C data with spiral readout can be reduced by the
self-estimated B0 inhomogeneity.Acknowledgements
This
work is supported by the National Institute of Biomedical Imaging and
Bioengineering (P41EB013598, U01EB026412), Myokardia Myoseeds award, UCSF
Resource Allocation Program, and American Cancer Society Research Scholar Grant
131715-RSG-18-005-01-CCE.References
[1]
Hollingsworth JM, Miller DC, Daignault S, Hollenbeck BK. Rising incidence of
small renal masses: a need to reassess treatment effect. J Natl Cancer Inst.
2006; 98(18): 1331-1334.
[2]
Welch HG, Skinner JS, Schroeck FR, Zhou W, Black WC. Regional Variation of
Computed Tomographic Imaging in the United States and the Risk of Nephrectomy.
JAMA Intern Med. 2018;178(2):221-227.
[3] Mayer,
Dirk, et al. “Dynamic and high-resolution metabolic imaging of hyperpolarized
[1-13C]-pyruvate in the rat brain using a high-performance gradient insert.”
Magnetic resonance in medicine 65.5 (2011): 1228-1233.
[4] Josan,
Sonal, et al. “Effects of isoflurane anesthesia on hyperpolarized 13C metabolic
measurements in rat brain.” Magnetic resonance in medicine 70.4 (2013):
1117-1124.
[5] Cunningham, Charles H., et al. "Hyperpolarized 13C
metabolic MRI of the human heart: initial experience." Circulation research 119.11 (2016): 1177-1182.
[6] Rider, Oliver J., et al. "Noninvasive in vivo assessment
of cardiac metabolism in the healthy and diabetic human heart using
hyperpolarized 13C MRI." Circulation Research 126.6 (2020): 725-736.
[7]Tyler,
Andrew, et al. “A 3D hybrid-shot spiral sequence for hyperpolarized imaging. “ Magnetic
Resonance in Medicine 85.2 (2020): 790-801.
[8] Man,
Lai‐Chee, John
M. Pauly, and Albert Macovski. “Improved automatic off‐resonance correction without a field map in spiral imaging.” Magnetic resonance in medicine
37.6 (1997): 906-913.
[9] Lau,
Angus. Frequency-selective Methods for Hyperpolarized 13C Cardiac Magnetic
Resonance Imaging. Diss. 2012.
[10] Julia
Traechtler, et al. “Joint Image and Field Map Estimation for Multi-Echo Hyperpolarized
13C Metabolic Imaging.” ISMRM 2020, abstract ID: 3016.
[11] Fessler, Jeffrey A., and Bradley P. Sutton. “Nonuniform fast Fourier
transforms using min-max interpolation.” IEEE transactions on signal processing
51.2 (2003): 560-574.
[12]
Sutton, Bradley P., Douglas C. Noll, and Jeffrey A. Fessler. “Fast, iterative
image reconstruction for MRI in the presence of field inhomogeneities.” IEEE
transactions on medical imaging 22.2 (2003): 178-188.