Faraz Amzajerdian1, Tahmina Achekzai1, Luis Loza1, Hooman Hamedani1, Yi Xin1, Harilla Profka1, Ian Duncan1, Stephen Kadlecek1, Kai Ruppert1, and Rahim Rizi1
1University of Pennsylvania, Philadelphia, PA, United States
Synopsis
Xenon-polarization
Transfer Contrast (XTC) imaging is a powerful technique for quantifying
exchange rates between gas- and dissolved-phase xenon compartments and involves
a series of radiofrequency (RF) saturation pulses applied to the targeted dissolved-phase
resonance. Although increasing the number of pulses and their spacing generates
greater contrast, it also increases acquisition time. In this work, in an
effort to reduce acquisition time, particularly for free-breathing imaging
protocols, we explored the use of continuous RF irradiation to depolarize
dissolved-phase xenon, with the goal of producing results similar to those
achieved with pulsed saturations, but in a significantly shorter time period.
Introduction
Hyperpolarized
xenon-129 (HXe) MRI provides a valuable tool for assessing disease-associated changes
in lung structure and function by measuring the distinct chemical shifts HXe
experiences when dissolved in the tissue/plasma (TP) or bound to red blood
cells (RBCs). Direct imaging of these dissolved-phase (DP) components is often
challenging, since only approximately 2% of HXe is dissolved at any given time
and certain lung diseases, such as emphysema, decrease gas exchange, further
reducing available signal. An alternative approach, Xenon-polarization Transfer
Contrast (XTC) imaging, indirectly quantifies DP xenon by taking advantage of
the high gas-phase (GP) signal1. By saturating the targeted DP
resonance, the GP is depolarized proportionally to the exchange between those
components. Traditionally, multiple saturation pulses interspersed with
specific delay times to allow for gas exchange were used to achieve sufficient
contrast. This approach is time inefficient, however, and prolongs the
acquisitions due to the need for an extended contrast preparation period. In
this work, we explored the feasibility of using long, compartment-selective RF
pulses2,3,4 for faster, more efficient saturation in healthy and
emphysematous rats. Methods
Imaging was performed in 4 male
Wistar rats (300 – 400 g) according to protocols approved by the Institutional
Animal Care and Use Committee. In one animal, emphysema was induced via a localized
elastase injection (100 µL/g) approximately one month prior to imaging. Animals
were ventilated with room air and then switched to an 80% HXe, 20% oxygen
mixture for 4 wash-in breaths, followed by imaging during a 10 second
breath-hold. All imaging adhered to SAR limitations and was performed on a 1.5
T scanner (Magnetom Avanto, Siemens) with a custom xenon-129 birdcage coil (Stark
Contrast, Germany). A prototype commercial system (XeBox-E10, Xemed LLC, NH)
was used to polarize 87% enriched xenon-129.
XTC data sets were obtained from
three consecutive 2D GP GRE projection images (S1, S2,
and S3). Imaging parameters included: flip angles (αi)
of 4°, 9°, and 13°, respectively, TR/TE = 8.1/3.9 ms, matrix size = 36x48, and
FOV of 60 mm. For continuous RF acquisitions, a series of rectangular
saturation pulses with varying lengths (25-100 ms) and flip angles (1,800°-54,000°)
was applied on-resonance (213 ppm for RBC or 199 ppm for TP) in-between the
second pair of images to generate contrast, and off-resonance (-213 ppm for RBC
or -199 ppm for TP) between the first pair as a control for apparent T1 decay
and any inadvertent GP magnetization loss. Due to hardware limitations, 500 ms
and 1 s saturation pulses had to be implemented by repeating 100 ms pulses with
minimal separation (< 0.3 ms). For pulsed saturations, 60 8 ms Gaussian 180°
RF pulses with 30 ms spacing were used. Depolarization ratio maps, RDepol,
were calculated from the ratio of these images, as shown in the equation below, where RCorr
and RExch are the T1-correction and exchange-weighted
depolarization maps, respectively, adjusted for excitation flip angle and
number of phase-encoding lines (PE).
$$ R_{Depol}=\frac{R_{Exch}}{R_{Corr}} =\frac{S_3 S_1}{S_2^2} (\frac{sin^2 \alpha_2}{sin \alpha_1sin \alpha_3})(\frac{cos \alpha_1}{cos \alpha_3})^{PE/2} $$Results and Discussion
Figure 1
depicts the depolarization ratios associated with the control saturation scheme,
RCorr, illustrating the minimal impact of the off-resonance
continuous saturations on the GP magnetization in a healthy rat. An average RCorr
of 0.82
for 500 ms saturation (five 100 ms pulses) with flip angles ranging from 1.8°/ms to 72°/ms—compared to 0.80 seen with a
0° flip angle—suggests that these continuous pulses do not directly depolarize
the GP. Figure 2 illustrates the effect of increasing flip angle on the
depolarization of the TP resonance in another healthy rat. There is a clear
trend despite the long saturation time, which allows xenon to exchange
repeatedly between the various compartments and introduces uncertainty with
regards to the flip angle that each xenon atom experiences. In order to
demonstrate consistency and RF amplifier stability, very large flip angles were
evaluated. As seen in Figure 3, saturations with the same flip angle per pulse
length at various flip angle and pulse length combinations produced similar
depolarization levels. In Figure 4, discrete and continuous saturation schemes
are compared in an emphysema rat model. Due to the lower than expected xenon
gas exchange in this model, 1 s of saturation (ten 100 ms pulses) was applied
at a rate of 18°/ms. The distributions were consistent between the two methods
with average depolarization ratios of 0.42 ± 0.18 and 0.49 ± 0.18 for TP and 0.70
± 0.16 and 0.75 ± 0.16 for RBC, respectively. However, the continuous
saturation scheme significantly reduced overall acquisition time by 45%,
displaying significant promise for use in free-breathing imaging protocols,
especially with further optimizations of imaging and saturation parameters. Conclusion
XTC MRI with continuous
RF irradiation produces depolarization distributions similar to those obtained
with pulsed saturation schemes, but with a significant time advantage that
decreases required breath-hold durations. Acknowledgements
No acknowledgement found.References
[1] Ruppert, Kai, et al (2000). Probing
lung physiology with xenon polarization transfer contrast (XTC). MRM 44(3),
349-357.
[2] Amzajerdian, Faraz, et al (2020). Measuring pulmonary
gas exchange using compartment‐selective xenon‐polarization transfer contrast
(XTC) MRI. MRM 00, 1-14.
[3] Sun, Philip, et al (2011). Simulation
and optimization of pulsed radio frequency irradiation scheme for chemical
exchange saturation transfer (CEST) MRI—demonstration of pH-weighted
pulsed-amide proton CEST MRI in an animal model of acute cerebral ischemia.
MRM 66(4), 1042-1048.
[4]
Kunth, Martin, et al (2015). Continuous-wave saturation considerations for
efficient xenon depolarization. NMR in Biomedicine 28(6) 601-606.