Hitoshi Kubo1, Yuya Abe2, Tomoya Yokokawa2, Seira Yokoyama2, and Koji Hoshi2
1Fukushima Medical University, Fukushima, Japan, 2Hoshi General Hospital, Koriyama, Japan
Synopsis
We aim to assess fundamental noise
reduction performance using deep learning reconstruction with a phantom at a
1.5 T MR scanner. In this study, the relationship among parameters for noise
reduction, signal-to-noise ratio, image quality, and spatial resolution of
images was evaluated. SNRs were increased higher significantly by DLR in all
SNR ranges. Increasing ratio of SNR was varied by means of parameter settings.
Combination of the DLR parameters affected varies to SNR, SSIM, and spatial
resolution of the images. We should exercise caution to select DLR parameters
when this technique applies to clinical images.
[Introduction]
Image quality in magnetic
resonance (MR) is based on the MR physics, sometimes happens some conflicts of
parameters to increase image quality such as signal-to-noise ratio (SNR),
spatial resolution, scan time, and so on. (1) Strength of magnetic field is one
of the parameters for defining SNR, it is very difficult to change it to
increase SNR. (2) Deep learning is a branch of artificial intelligence and has
been developing a lot of algorithms to solve complex problems related to
medical images. Deep learning reconstruction (DLR) (Advanced Intelligent
Clear-IQ Engine {AiCE}, Canon Medical Systems, Tokyo, Japan) was the first
commercialized DLR tool to reduce noise of MR images. (3) It has some
parameters to manage the strength of noise reduction, and we have little
evidence to refer to determine parameters. The purpose of this study was to clarify
fundamental characteristics of this tool using a phantom to provide properties
for clinical use.[Methods]
1.5T MR equipment (Vantage Orian,
Canon Medical Systems) with body and spine Coils was used. Type 90-401 phantoms
containing polyvinyl alcohol (PVA) gel (Nikko Fines Industries Co. Ltd.) was
used to evaluate image quality. 2-dimensional spin-echo (SE) sequence with TR =
500 ms., TE = 15 ms., number of average = 1, pixel bandwith = 217 Hz, FOV = 25.6
x 25.6 cm2, matrix = 368 x 368, spatial resolution = 0.7 x 0.7 mm2,
scan time = 3:05. Images with 8 mm, 4 mm, and 2 mm of slice thickness were
obtained to change SNR. 6 slices of axial images with the gap more than 2 times
wider slice thickness were obtained for each phantom section. After obtained
these images, DLR was applied to reduce noise with various parameters.
To evaluate image quality, ROIs
were set on the images and measure mean and SD on them. Although some SNR
calculation methods have been proposed in the past literature, SNR was defined
as a mean value divide by SD on the same ROI in this study. (4-6) SNRs were
calculated on each ROI on each slice with various DLR parameters. Structural
similarity (SSIM) were calculated using MATLAB R2020b. In addition to these
analyses, visual evaluation with a scoring method of pin section was performed
to evaluate changes of spatial resolution by four radiological technologists
who have various experience of clinical MR exam. All statistical analyses were
performed with SPSS statistics 27. Repeated-measures-ANOVA were performed to detect
significant differences. Then Friedman test was performed as a posthoc test.[Results]
Figure 1 shows SNR changes in
terms of DLR on images with 2 mm, 4 mm, and 8 mm slice thickness, respectively.
Original indicated without DLR. DLR made SNR of images higher in comparison to
that of original images. Figure 2 shows SSIM changes in terms of DLR.
Decreasing of SSIM was shown in more thin slices. Edge enhancement function
recovered SSIM higher, especially in the lower SSIM parameters. The results of
visual estimation showed that mean scores were recovered by using edge
enhancement, especially in the phase encoding direction compared to that in the
readout direction.[Discussion]
The fundamental noise reduction
performance of DLR was evaluated using a phantom in this study. SNRs were
increased higher significantly by DLR in all SNR ranges. Increasing ratio of
SNR was varied by means of parameter settings. Decreasing of SSIM higher in
lower SNR images is compared to that in higher SNR images. Edge enhancement
recovered SSIM higher, especially in lower SNR images. The results of visual
estimation indicated that little changes of a mean score by means of parameters
were shown and edge enhancement has little effects on improvement of spatial
resolution. Different trends of mean score of visual evaluation by means of the
DLR parameters between phase encode direction and readout direction of the
images were found.[Conclusion]
DLR had a powerful performance to
increase SNR in all SNR ranges. Combination of the DLR parameters affected
varies to SNR, SSIM, and spatial resolution of the images. We should exercise
caution to select DLR parameters when this technique applies to clinical
images.Acknowledgements
No acknowledgement found.References
1. Saloner D, et al., Magn Reson
Imaging Clin N Am. 2015;23(1):1–6.
2. Rutt BK, et al., JMRI.
1996;1:57-62.
3. Kidoh M, et al., Magn Reson Med
Sci. 2020;19(3):195-206.
4. NEMA Standards Publication MS
1-2008 (R2014)
5. Dietrich O, et al., J Magn
Reson Imaging. 2007;26:375-385.
6. Goerner FL, et al., Med Phys.
2011;38(9):5049-5057