Xu Lulu1, Shen Yong2, Dou Weiqiang3, and Qi Liang1
1Radiology, Jiangsu Province Hospital, Nanjing, China, 2GE Healthcare, MR Enhanced Application China, Beijing, China, 3GE Healthcare, MR Research China, Beijing, China
Synopsis
This
study aimed to determine the optimal direction of the frequency-encoding
gradient when acquiring T2WI-STIR images with titanium alloy. To investigate this,
we made in vitro MRI experiment. The water-pork phantom with ten
different-length titanium alloys was constructed, and the short- and long-axis
ratios calculated in different frequency-encoding directions were compared with
each other, respectively. We found that when the frequency-encoding axis was perpendicular
to the screw, the metallic artifact was smaller in whatever long axis or short
axis. This result may improve observing more anatomical structures around the embedded
metal.
Introduction
Magnetic
resonance imaging (MRI) for high soft-tissue contrast has been applied in the musculoskeletal
(MSK) system. However,
it remains challenging to acquire T2WI fat-suppressed images for MSK tissues
with metal implanted for clinical purposes, largely due to strong magnetic
field inhomogeneity.
So far,
few studies have focused on the metal-reduced artifact of T2WI fat-suppressed
imaging, which is often applied in MSK imaging, although this artifact on T1WI
was investigated previously.1 On the other
hand, the directional frequency-encoding gradient was strongly related to the
metal artifact, and the choice of the direction of the frequency-encoding
gradient depended on the area of clinical interest. We have previously shown
that the STIR technique effectively reduces metal artifact,2 so it is
interesting to find the relationship between T2WI-STIR imaging and the directional
frequency-encoding gradient.
Therefore,
the main goal of this study was to investigate if T2WI with STIR incorporated
can achieve minimal metal artifact in a phantom with metal embedded by
optimizing the frequency-encoding gradient direction. Materials and methods
A phantom was self-made using ten different-length
titanium alloy pedicle screws suspended in the water-pork phantom center.
All sagittal
MR images were obtained parallel to the long axis of the screw, using STIR
technique. In T2WI-STIR images, according to the American Society for Testing
and Materials (ASTM) F2119-01,3 the apparent long-
and short-axis lengths of the screw were measured when the frequency-encoding
gradient was parallel or perpendicular to the long axis. The long- and
short-axis ratios were defined as (measured length) / (actual length). Imaging
parameters were TR 3029 ms, TE(eff) 85 ms, ETL 15, matrix 256×192, FOV
256×256, BW
83.34 kHz, and slice thickness 3 mm. Total scan time was 2 minutes, 56 seconds.
The obtained
long- and short-axis ratios were compared when the frequency-encoding gradient
was parallel to the long-axis or perpendicular to the long axis. Additionally,
the long-axis ratios and short-axis ratios calculated in different
frequency-encoding directions were also compared with each other, respectively.
All statistical analyses
were performed using the Wilcoxon signed-rank test. The significance threshold
was set as P < 0.05. Results
In
phantom measurement, the short-axis length was approximately three times
greater than the actual screw length, while the long-axis length was similar to
the true measurements (table 1). When the direction of the frequency-encoding
gradient was parallel to the long axis of the screw, both short-axis and
long-axis ratios were smaller compared with those at the frequency-encoding
perpendicular to the screw (P < 0.05)
( table 1).Discussion
In this
phantom study, when the frequency-encoding gradient was applied parallel with
the long-axis screw, less metal artifact was displayed in whatever long-axis or
short-axis direction than setting the frequency-encoding gradient perpendicular
to the long-axis screw.
The
region around the metal implanted was usually a specific
area of interest in clinical diagnosing. However, the metal artifact
often limits the observation of anatomy. Since our previous study demonstrated that
the STIR technique showed the least metal artifact compared to FSFS and IDEAL,2 STIR imaging was
utilized in this experiment. In the
T2WI-STIR images, metal artifacts were composed of inhomogeneous signals,
including low and high signal intensities (Fig.1(a-b)). The lower signal
outside the screw was originated from fat. It can be explained that the field perturbations
caused faster transverse relaxation, thus led the fat signal to almost zero.
The high signal around the screw was probably the unsaturated fat from in-plane
or through-plane. This phenomenon can be explained that the titanium alloy was a
paramagnetic material,so that
it may shorten the T1 value of fat, resulting in a change of the null point in
STIR.
Additionally,
the short- and long-ratios revealed that the artifact length was longer in the
phase-encoded direction(Fig.1(c-d)). One
reason may be that the orientation of the screw in the phantom was
perpendicular to the static magnetic field leading to the particular field perturbations;4 another reason
may be that the screw shape has a larger interface with pork oil in phase-encoded
direction, resulting in changed local magnetic fields, further reducing the
effectiveness of fat suppression. Furthermore, longer artifact length in the short
axis was produced when the frequency-encoding gradient was applied perpendicular
to the long-axis screw. The reason was that susceptibility artifacts were more
extended in the frequency direction. However, it was not clear why the longer
artifact length in the long axis. This finding was slightly different from
Frazzini’s results,5 which
showed the shorter artifact in the long axis.Acknowledgements
We
thank Synthes GmbH company (Switzerland) for generously providing the screw used
in this study.References
1. Lewin JS, Duerk
JL, Jain VR, et al. Needle localization in MR-guided biopsy and aspiration:
effects of field strength, sequence design, and magnetic field orientation. AJR
Am J Roentgenol. 1996;166(6):1337-1345.
2. Lulu X, Liang Q,
Weiqiang D, et al. The evaluation of the metal artifact and SNR changes in
multiple fat-suppression techniques:a phantom
study. ISMRM, 2019.
3. ASTM. Standard
test method for evaluation of MR image artifacts from passive implants. 2001.
4. Schenck JF. The
role of magnetic susceptibility in magnetic resonance imaging: MRI magnetic
compatibility of the first and second kinds. Med Phys. 1996;23(6):815-850.
5. Petersilge CA,
Lewin JS, Duerk JL, et al. Optimizing imaging parameters for MR evaluation of
the spine with titanium pedicle screws. AJR Am J Roentgenol. 1996;166(5):1213-1218.