Yuheng Huang1,2, Xinheng Zhang1,2, Serry Fradad1, Lu Meng1, Ghazal Yoosefian1, Linda Azab1, Xinqi Li3, Alan Kwan4, Rohan Dharmakumar1, Hui Han1, and Hsin-Jung Yang1
1Biomedical Imaging Research Institute, Cedars Sinai Medical Center, Los Angeles, CA, United States, 2Bioengineering, UCLA, Los Angeles, CA, United States, 3Columbia University, New York, NY, United States, 4Cardiology, Cedars Sinai Medical Center, Los Angeles, CA, United States
Synopsis
A strong off-resonance field
around metallic implants induces unreliable phase maps and limits the
application of phase-based MRI techniques. State-of-the-art MR phase mapping
and unwrapping techniques perform poorly under the influence of metallic
implants because of their limited dynamic range and susceptibility to noise. In
this study, we developed a fully automatic phase mapping technique based on a multi-echo
GRE sequence with unevenly spaced echoes and a high-dynamic-range phase
reconstruction algorithm. We tested the proposed method with numerical
simulations and human subjects with metallic implants.
Introduction
Recent technical advancements in phase-based MRI
applications, such as susceptibility imaging, flow-encoded imaging, and B0
field mapping, have demonstrated major clinical values in diagnosis,
pathophysiological research, and image quality improvement.1, 2 The
accurate derivation of the phase terms in MR images serve as the foundation of
these modalities. However, accompanied by the uprising popularity of high field
MRI scanner and MRI safe metallic implants, strong B0 field inhomogeneity
around metallic implants became a common confounder in phase map derivation.
The strong B0 off-resonance induced by the metallic objects drastically impairs
the accuracy, reliability, and applicability of the phase-based MRI techniques,
particularly in the regions with strong phase accumulation and multi-fold phase
wrapping. Although algorithms had been proposed to unwrapping MR phase images,3,4,5 the state-of-the-art
algorithms were not suitable for the extreme conditions of the presents of
metallic implants and are confounded by limited phase dynamic range and high
susceptibility to noise. These challenges result in significant phase wrapping
artifacts, shifted central frequency and unreliable against sudden phase
changes at fat/water interfaces. In this study, we proposed a new phase
processing algorithm, named “Accurate mUlti-echo phase image wiTh uneven echO
spacing and Ultra-High Dynamic Range
(AUTO-HDR)”, that is robust against noise and is able to resolve
ultra-high dynamic range to the level of off-resonance near metallic
implants. We tested the proposed methods
in numerical phantoms and compared them with four other state-of-the-art phase
unwrapping techniques. We also validated the results in human subjects with an
implantable cardioverter-defibrillator on their chest.Method
The
image processing pipeline of AUTO-HDR is depicted in Fig.1A. A multi-echo GRE(mGRE)
sequence with unevenly spaced echoes was adopted(Fig.1B) to create an extended
dynamic range for the phase maps.6,7 To achieve a higher dynamic
range, a graph-cut based algorithm was used and guided by the second temporal
derivative phase images to utilize the high SNR in long TE images and preserve
the accurate and reliable unwrapping property from the images with short TEs. To
validate the proposed method, a numerical simulation, and in-vivo experiments
were performed and described as the following:
Numerical
Phantom:
Based on the regularized inversion method,8 a numerical phantom was
generated with a shepp-logan model9 and a common metallic object (implantable
cardioverter-defibrillator; ICD Medtronic VISIA AF), size. The position of imaging
FOV, ferromagnetic object, and the induced field map is illustrated in Fig 2A.
Images were simulated under different SNR levels to emulate the
imaging condition.10 An mGRE sequence
with the following experimental setups was simulated based using the Bloch
equation to generate the corresponding complex images. TE:(1.07, 3.07, 5.27,7.67)ms; TR:1124.67ms ; FOV:360*125*96mm3; Voxel Size:1.607*1.607*3.214mm3. Repeated experiments (N=20) were conducted to test
the robustness and reproducibility of the algorithms. All images were processed
with AUTO-HDR and four other state-of-the-art phase unwrapping techniques
(Quality Guide, SPURS, UMPIRE, and LAPLACIAN). The quality of the phase maps was
measured quantitatively in an ROI close to the off-resonance source(Fig. 2B). Normalized Root Means Square Error (NRMSE) and
SD of error maps was measured in the ROI and compared using bar plots. In-vivo experiment: To further validate the simulated results, an
in-vivo experiment was performed at a clinical 3T scanner (mMR, Siemens). A
human volunteer with a metallic off-resonance source(implantable
cardioverter-defibrillator; ICD) on the left anterior chest wall was scanned with
the same imaging parameter. 3D mGRE Images were acquired with a 25-sec breath-hold
and covering the whole chest.
Results
Numerical Phantom: Fig.3 showed the results from unwrapped phase
maps of the numerical phantom under different noise levels. Phase maps derived using AUTO-HDR and four
other state-of-the-art unwrapping algorithms were compared with the theoretical
inputs, and error maps were derived. In the area close to the off-resonance
source, SPURS, Laplacian, and Quality Guided with long TE(DTE=6.4ms), showed strong error in determining the central
frequency of the phase maps. Although quality Guided images using short TE(DTE=2ms) and UMPIRE can achieve correct central
frequencies, the strong amplification of noise corrupted the images significantly,
particularly in the region close to the off-resonance source. The proposed AUTO-HDR, however, was able to unwrap the images in the whole FOV and performed
robustly against different noise levels. Similar trends were observed in the
quantitative analysis(Fig.4). The proposed method showed unparallel robustness
compare to the other algorithms and demonstrated significantly lower RMSE and higher
SNR at the ROI under various noise levels. In-vivo experiment: Similar results to the simulation are presented
in the in-vivo experiment(Fig 5). Improved phase wrapping artifacts, off-centered
central frequency and lower noise level are shown with the proposed AUTO-HDR
compare to the peer algorithms. Conclusion
AUTO-HDR can accurately estimate MRI phase maps under
the presence of metallic implants. It demonstrates superior accuracy and
robustness compare to the state-of-the-art algorithms under insufficient SNR,
and ultra-high off-resonance range. A more comprehensive study on more human
subjects with quantitative measures is needed to exam its in-vivo reliability. Acknowledgements
This work is supported by the NIH (R43 NS120795)References
1.Haacke EM, Mittal S, Wu Z, Neelavalli J, Cheng YC. Susceptibility-weighted imaging: technical aspects and clinical applications, part 1. AJNR Am J Neuroradiol. 2009 Jan;30(1):19-30. doi: 10.3174/ajnr.A1400. Epub 2008 Nov 27. PMID: 19039041; PMCID: PMC3805391.
2. Juchem C, Umesh Rudrapatna S, Nixon TW, de Graaf RA. Dynamic multi-coil technique (DYNAMITE) shimming for echo-planar imaging of the human brain at 7 Tesla. Neuroimage. 2015 Jan 15;105:462-72. doi: 10.1016/j.neuroimage.2014.11.011. Epub 2014 Nov 8. PMID: 25462795; PMCID: PMC4262558.
3. K Bartusek et al 2006 Meas. Sci. Technol. 17 3293
4. Robinson SD, Bredies K, Khabipova D, Dymerska B, Marques JP, Schweser F. An illustrated comparison of processing methods for MR phase imaging and QSM: combining array coil signals and phase unwrapping. NMR Biomed. 2017 Apr;30(4):e3601. doi: 10.1002/nbm.3601. Epub 2016 Sep 13. PMID: 27619999; PMCID: PMC5348291.
5. Dong J, Liu T, Chen F, Zhou D, Dimov A, Raj A, Cheng Q, Spincemaille P, Wang Y. Simultaneous phase unwrapping and removal of chemical shift (SPURS) using graph cuts: application in quantitative susceptibility mapping. IEEE Trans Med Imaging. 2015 Feb;34(2):531-40. doi: 10.1109/TMI.2014.2361764. Epub 2014 Oct 8. PMID: 25312917.
6. Robinson S, Schödl H, Trattnig S. A method for unwrapping highly wrapped multi-echo phase images at very high field: UMPIRE. Magn Reson Med. 2014 Jul;72(1):80-92. doi: 10.1002/mrm.24897. Epub 2013 Jul 30. PMID: 23901001; PMCID: PMC4062430.
7. Dagher J, Reese T, Bilgin A. High-resolution, large dynamic range field map estimation. Magn Reson Med. 2014 Jan;71(1):105-17. doi: 10.1002/mrm.24636. Epub 2013 Feb 11. PMID: 23401245; PMCID: PMC3919137.
8. Shi X, Yoon D, Koch KM, Hargreaves BA. Metallic implant geometry and susceptibility estimation using multispectral B0 field maps. Magn Reson Med. 2017 Jun;77(6):2402-2413. doi: 10.1002/mrm.26313. Epub 2016 Jul 6. PMID: 27385493; PMCID: PMC5529184.
9. H. M. Gach, C. Tanase and F. Boada, "2D & 3D Shepp-Logan Phantom Standards for MRI," 2008 19th International Conference on Systems Engineering, Las Vegas, NV, 2008, pp. 521-526, doi: 10.1109/ICSEng.2008.15.
10. Rashid S, Rapacchi S, Vaseghi M, Tung R, Shivkumar K, Finn JP, Hu P. Improved late gadolinium enhancement MR imaging for patients with implanted cardiac devices. Radiology. 2014 Jan;270(1):269-74. doi: 10.1148/radiol.13130942. Epub 2013 Oct 28. PMID: 24086074; PMCID: PMC4228714.