Kosuke Ito1 and Atsushi Kuratani1
1Healthcare Business Unit, Hitachi, Ltd., Kashiwa, Japan
Synopsis
T1
weighted image is acquired using Spin Echo sequence in almost clinical
routine practices. FID signal induced by 180 degrees pulse causes zipper
artifact at the center of the image. To move the artifact to the edge of the
image, the phase of RF pulse is controlled as 0 and 180 degrees alternately.
However, higher parallel imaging factor cannot be applied due to the FID
artifact still exist at the edge of FOV. In this study, FID artifact was dispersed
uniformly by modulating phase of 180 degrees pulse with quadratic function, and
higher parallel imaging factor was applied in vivo.
Purpose
The purpose of this study is dispersing FID
artifact uniformly in T1 weighted image (T1WI) acquired with Spin Echo (SE)
sequence and apply higher parallel imaging factor.Introduction
T1WI is acquired in clinical routine practices
by using SE sequence with short TR and short TE. In T1WI, FID signal induced by
the 180 degrees pulse causes zipper artifact in the center of the image [1]. To
move the artifact to the edge of the FOV, phase of RF pulse is controlled as 0
and 180 degrees alternately. However, the artifact still exists at the edge of
the FOV, the use of parallel imaging is limited to small acceleration factor. Recently,
we proposed algorithm of decomposing FID artifact from T1WI [2]. The method
uses information of phase distribution measured from center of k-space data. So
the accuracy of decomposition depends on the accuracy of phase estimation. In
this study, the FID artifact was dispersed uniformly by modulating RF phase of
180 degrees pulse, and signal intensity of FID artifact was suppressed as $$$1/\sqrt{N}$$$ . And higher parallel imaging
acceleration factor was applied in vivo.Materials and Methods
Modulation of RF phase of 180 degrees
pulse
We propose modulation of RF phase of 180
degrees pulse as equation 1.
$$θ_k=\frac{\pi}{N}k^2 \hspace{10pt} (1)$$
Where
k is phase encode number, and N is total number of phase encoding: N=#ph/R. Then
FID signal in k-space is modeled as equation 2. The amplitude of FID signal is
constant along phase encode direction, the amplitude is assumed as 1 in equation
2.
$$S(k)=\exp(iθ_k ) \hspace{10pt} (2)$$
FID
artifact in image domain is FFT of FID signal, signal intensity of FID artifact
is described as equation 3, and magnitude of signal intensity of FID artifact
is described as equation 4.
$$\begin{eqnarray}I(x)&=&\frac{1}{\sqrt{N}}\sum_{k=0}^{N-1}\exp\left(iθ_k\right)\exp\left(\frac{2\pi i kx}{N}\right) \\&=& \frac{1}{\sqrt{N}}\sum_{k=0}^{N-1}\exp\left(\frac{i\pi\left(k^2+2kx\right)}{N}\right) \\&=&\frac{1}{\sqrt{N}}\sum_{k=0}^{N-1}\exp\left(\frac{i\pi\left(\left(k+x\right)^2-x^2\right)}{N}\right) \hspace{10pt}(3)\end{eqnarray}$$
$$\begin{eqnarray}\mid I(x)\mid^2 &=&\frac{1}{N}\left[\sum_{k=0}^{N-1}\exp\left(\frac{i\pi\left(\left(k+x\right)^2-x^2\right)}{N}\right)\right]\left[\sum_{k=0}^{N-1}\exp\left(\frac{-i\pi\left(\left(k+x\right)^2-x^2\right)}{N}\right)\right]\\&=&\frac{1}{N}\left[\sum_{k=0}^{N-1}\exp\left(\frac{i\pi\left(\left(k+x\right)^2\right)}{N}\right)\right]\left[\sum_{k=0}^{N-1}\exp\left(\frac{-i\pi\left(\left(k+x\right)^2\right)}{N}\right)\right]\\&=&\frac{1}{N}\left[\sum_{k=1}^{N-1}\left\{\exp\left(\frac{i\pi\left(\left(k+x\right)^2\right)}{N}\right)+\exp\left(\frac{i\pi x^2}{N}\right)\right\}\right]\left[\sum_{k=1}^{N-1}\left\{\exp\left(\frac{-i\pi\left(\left(k+x\right)^2\right)}{N}\right)+\exp\left(\frac{-i\pi x^2}{N}\right)\right\}\right] \hspace{10pt} (4)\end{eqnarray}$$
Magnitude
of signal intensity at the next pixel is described as equation 5.
$$\begin{eqnarray}\mid I(x+1)\mid^2 &=&\frac{1}{N}\left[\sum_{k=0}^{N-1}\exp\left(\frac{i\pi\left(\left(k+x+1\right)^2-\left(x+1\right)^2\right)}{N}\right)\right]\left[\sum_{k=0}^{N-1}\exp\left(\frac{-i\pi\left(\left(k+x+1\right)^2-\left(x+1\right)^2\right)}{N}\right)\right]\\&=&\frac{1}{N}\left[\sum_{k=0}^{N-1}\exp\left(\frac{i\pi\left(\left(k+x+1\right)^2\right)}{N}\right)\right]\left[\sum_{k=0}^{N-1}\exp\left(\frac{-i\pi\left(\left(k+x+1\right)^2\right)}{N}\right)\right]\\&=&\frac{1}{N}\left[\sum_{k=1}^{N}\exp\left(\frac{i\pi\left(k+x\right)^2}{N}\right)\right]\left[\sum_{k=1}^{N}\exp\left(\frac{-i\pi\left(k+x\right)^2}{N}\right)\right]\\&=&\frac{1}{N}\left[\sum_{k=1}^{N-1}\left\{\exp\left(\frac{i\pi\left(k+x\right)^2}{N}\right)+\exp\left(\frac{i\pi\left(N+x\right)^2}{N}\right)\right\}\right]\left[\sum_{k=1}^{N-1}\left\{\exp\left(\frac{-i\pi\left(k+x\right)^2}{N}\right)+\exp\left(\frac{-i\pi\left(N+x\right)^2}{N}\right)\right\}\right] \hspace{10pt} (5)\end{eqnarray}$$
When N is even, equation 6 holds.
$$\begin{eqnarray}\exp\left(\frac{i\pi\left(N+x\right)^2}{N}\right)&=&\exp\left(\frac{i\pi\left(N^2+2Nx+x^2\right)}{N}\right)\\&=&\exp\left(\frac{i\pi x^2}{N}\right) \hspace{10pt} (6)\end{eqnarray}$$
So $$$\mid I(x)\mid^2=\mid I(x+1)\mid^2$$$ holds, and signal intensity is constant along phase
encoding direction. FID artifact was dispersed uniformly by modulating RF phase
of 180 degrees pulse as equation 1.
Numerical calculation
Numerical calculation was conducted to
confirm signal distribution and signal intensity of FID artifact. Number of
phase encoding was set 128, 192, 256, and 512.
Experiments
Experiment was conducted on a 3 Tesla whole
body MRI system (Hitachi, Ltd.). A phantom and a healthy volunteer were evaluated.
This study was approved by the ethics committee of Hitachi group headquarters. 32
channels receive coil was used. Because T2 of phantom is shorter than
volunteer, to confirm FID artifact clearly, crusher pulse was not applied on
phantom experiment. Scan parameters of T1WI were as follows; TR / TE = 500 / 11
ms, thickness = 5 mm, FOV = 230 mm, Freq# x Phase# = 256 x 256, and scan time
was 2 min 10 sec. In this study, parallel imaging factor of 2 was applied.
Evaluation
SNR was evaluated on phantom study. And line
profile along the FID artifact was evaluated on phantom and volunteer study.Results and discussions
Figure
1 shows proposed phase modulation of 180 degrees pulse described as equation 1
and corresponding signal intensity of FID artifact. The phase is expressed
modulo 360 degrees. Signal intensity of FID artifact was constant. Magnitude of
FID artifact calculated by equation 3 is showed in Table 1. As number of phase
encode increases, magnitude of FID artifact is suppressed to $$$1/\sqrt{N}$$$. Figure
2 shows phantom image. By modulating phase as equation 1, FID artifact
dispersed along phase encoding direction as confirmed on red arrow. In Figure 2
(c), the zig-zag behavior of signal intensity due to the FID artifact on
conventional image was improved by modulating phase with quadratic function. In
table 2, SNR is summarized. SNR was almost same between conventional image and
proposed image. Thought proposed method disperses FID artifact in entire image,
magnitude of FID artifact is suppressed to $$$1/\sqrt{N}$$$, degradation
of SNR was little. Figure 3 shows a volunteer image. By modulating phase of 180
degrees pulse, FID artifact was not confirmed on Figure 3 (b). And in Figure 3
(c), FID artifact was confirmed on the line of white arrow where FID artifact
was confirmed on Figure 3 (a). Figure 3 (d) is line profile at the line of FID
artifact. Same as phantom case, the zig-zag behavior of signal intensity was
improved.Conclusions
In
this study, simple method of modulation of 180 degrees pulse to disperse FID
artifact uniformly was proposed. The distribution of signal intensity of FID
artifact was calculated theoretically, and the magnitude was numerically
calculated as a function of phase encode number. The signal intensity of FID
artifact was suppressed to $$$1/\sqrt{N}$$$
compared to conventional phase control. FID
artifact was not confirmed on volunteer image, where the artifact was confirmed
on conventional phase controlled image. The proposed method of phase modulation
enables applying higher acceleration factor for acquisition of T1WI with SE
sequence in clinical routine practices.Acknowledgements
References
[1] Hoff, Michael & Andre, Jalal &
Stewart, Brent. (2016). Artifacts in Magnetic Resonance Imaging.
10.1201/b19609-10.
[2] K.
Ito et al. JSMRM 2020 O-053