Rohit Chacko Philip1, Ali Bilgin1,2,3, and Maria I Altbach1,2
1Medical Imaging, University of Arizona, Tucson, AZ, United States, 2Biomedical Engineering, University of Arizona, Tucson, AZ, United States, 3Electrical and Computer Engineering, University of Arizona, Tucson, AZ, United States
Synopsis
Streaking artifacts in
radial MR imaging due to gradient nonlinearities are suppressed using a
beamforming algorithm where region growing image segmentation is used to automatically
generate the interference covariance matrix. The performance of the automatic
streaking artifact suppression algorithm (RGB-STAR) is compared to algorithms
based on coil removal and coil weighting and a beamforming algorithm with manual
segmentation.
Introduction
Streaking artifacts that spread
over the field-of-view (FOV) are a drawback of radial MRI techniques.1
Streaking artifacts are most commonly associated with azimuthal undersampling
and motion but also arise as a consequence of gradient nonlinearities. This
type of streaking artifact tends to originate from the anatomy at the periphery
of the FOV, due to gradient imperfections close to the edge of the magnet.
Since the level of artifact varies across the phased array RF coil elements,1 several coil selection and combination methods as well as phased-array
processing algorithms have been proposed.2-5 Among these, the recently
proposed beamforming for streaking artifact reduction (B-STAR) algorithm6
has the advantage that coils do not need to be removed or weighted, thus
preserving the signal gains of phased array RF technology.6
B-STAR estimates an
interference covariance matrix using a region-of-interest (ROI) containing the
artifact and then generates a phased array coil combination map using the
signal correlation matrix at the spatial location and the fixed interference
covariance matrix. B-STAR relies on the manual placement of the ROI around areas
from which streaks emanate which is not practical for routine use, in particular
if several ROIs need to be selected. The automatic generation of the
interference covariance matrix by identifying the sources of streaking artifact
can save time and produce more robust artifact reduction. Here, we present the
RGB-STAR method which incorporates region growing image segmentation
to automatically generate the interference covariance matrix which is then used
to suppress streaking artifacts without reducing the signal in the anatomy of
interest. Methods
For the automatic
selection of the ROIs, we used a region growing (RG) method7 with
four seed points automatically placed at the four corners of the FOV of the
intensity normalized image. RG selects the background area of the FOV (i.e.,
areas not including anatomical features). The complement of the segmented
background region in abdominal MRI typically produces three foreground (or
anatomical) regions based on their spatial location within the image: (i) the
abdomen (region closest to the center of the image), (ii) the left arm (region closest
to the bottom right of the image), and (iii) the right arm (region closest to
the bottom left of the image) (Fig. 1). These foreground regions are used to
generate the interference covariance matrix in two ways: a) the pixels in the complement
of the central abdomen region (Auto Center) form the interference covariance
matrix and b) the pixels in the union of
the two arm regions (Auto Arms) form the interference covariance matrix.
RGB-STAR was tested
on a set of abdominal images acquired with a T2-weighted radial turbo spin echo
pulse sequence. We selected 26 slices (from 13 subjects) with strong streaking
artifacts originating from unsuppressed fat signal in the arms. The performance of RGB-STAR using the Auto
Center and Auto Arm ROI selection was compared qualitatively and quantitatively
against reference images (reconstructed with adaptive coil combination3
and without streak suppression) as well as images processed with a coil removal
with outlier pruning (CR+OP) method4 and a sensitivity weighted (SW)
coil combination method.5Results
Figure 2 compares the
level of streak removal achieved by RGB-STAR to the reference image and images
processed with SW, CR+OP, and B-STAR.
Note that the image reconstructed with SW has a similar level of
artifact as the reference image. CR+OP suppressed the streaks effectively but
at the expense of a significant signal reduction across the anatomy. B-STAR and
RGB-STAR (Auto Center and Auto Arm) are comparable and offer the best streaking
artifact suppression capabilities while preserving signal intensity through the
abdomen. The advantage of RGB-STAR
is that it is fully automated.
Figure 3 shows the mean/standard
deviation (SD) (evaluated on streak corrupted ROIs within the abdomen) and the
Interference Strength (defined as the mean signal intensity evaluated on
background ROIs)6 for all methods. Note that B-STAR and RGB-STAR yield
the lowest Interference Strength while maintaining high a mean/SD in foreground
regions.
The correlation between
the mean/SD values evaluated on 26 test ROIs across two central abdominal MRI
slices for 13 subjects using B-STAR and RGB-STAR are shown in Fig. 4. The Pearson’s
correlation coefficient,8 r, between B-STAR and the RGB-STAR (Auto Center)
method is 0.9837 (p-value: $$$2.2035e^{-19}$$$), and between B-STAR and
the RGB-STAR (Auto Arms) method is 0.8694 (p-value: $$$8.2366e^{-9}$$$). Figure
5 shows Bland-Altman plots9 indicating the magnitude of disagreement
(both error and bias), outliers, and trends between B-STAR and each of the
automated interference covariance matrix generation methods. Thus, both RGB-STAR approaches work
well and are comparable to B-STAR as indicated by strong positive correlation
in mean/SD values (Fig. 4) and further corroborated by strong agreement in
Bland-Altman analysis (Fig. 5).Conclusion
We propose a seeded
region growing image segmentation scheme combined with a beamforming algorithm to
automatically generate the interference covariance matrix to suppress streaking
artifacts in radial MRI images. Automatic seed point placement for the region
growing image segmentation scheme ensures that the entire streaking artifact
suppression method is fully automated, thereby reducing tediousness and
operator dependence. RGB-STAR yields images with artifact reduction superior to existing
methods and is fully automated.Acknowledgements
We would like to acknowledge grant
support from NIH (CA245920), the Arizona Biomedical Research Commission
(ADHS14-082996), and the Technology and Research Initiative Fund (TRIF). References
1. Xue
Y, Yu J, Kang HS, Englander S, Rosen MA, Song HK. “Automatic coil selection for
streak artifact reduction in radial MRI.” Magn Reson Med. 2012; 67:470-476.
2. Holme
HC, Frahm J. “Sinogram-based coil selection for streak artifact reduction in
undersampled radial real-time magnetic resonance imaging.” Quant Imaging Med
Surg. 2016; 6:552-556.
3. Walsh
DO, Gmitro AF, Marcellin MW. “Adaptive reconstruction of phased array MR
imaging.” Magn Reson Med., 2000; 43:682-690.
4. Grimm
R, Forman C, Hutter J, Kiefer B, Hornegger J, Block T. “Fast automatic coil
selection for radial stack-of-stars GRE imaging.” Proc. ISMRM, 2013, p.
3786.
5. Kholmovski
EG, Parker DL, Di Bella EV. “Streak artifact suppression in multi-coil MRI with
radial sampling.” Proc. ISMRM, 2007. p. 1902.
6. Mandava
S, Keerthivasan MB, Martin DR, Altbach MI, Bilgin A. “Radial streak artifact
reduction using phased array beamforming” Magn Reson Med. 2019;
81:3915-3923.
7. Adams
R, Bischof L. “Seeded Region Growing”, IEEE T-PAMI, 1994; 16(6):641-647.
8. Kutner
MH, Nachtsheim C, Neter J, Li W. “Applied Linear Regression Models.” 4th
ed., McGraw-Hill/Irwin, 2004.
9. Altman
D, Bland J. “Measurement in medicine: the analysis of method comparison
studies,” J R Stat Soc Ser D Stat. 1983; 32:307-317.