Geli Hu1, Jiazheng Wang1, Qingping Gu1, Yang Zhang1, and Jianxia Cao1
1Philips Healthcare, Beijing, China
Synopsis
The balanced FFE (b-FFE) sequence has the
advantages of high resolution, SNR, CNR and superior differentiation between fluid and
tissue. Therefore, it can be used for lumbosacral plexus
imaging [1]. However, black band-like artifacts are frequently observed due to intra-voxel signal dephasing at the presence of magnetic field
inhomogeneity. Here we proposed using a RF phase cycling technique (b-FFE-XD)
to suppress the banding artifacts. The RF phase was set as [0,180+1*X, 2*X, 180+3*X,4*X,…]
in consequence to limit the intravoxel dephasing, where X=n*360/NSA, and n is
the nth average. Results demonstrated successful removal of banding
artifacts with the proposed technique.
Introduction
The lumbosacral plexus is a complex
anatomic area, disfunction of which can lead to severe disability and morbidity
[2]. Despite the use of balanced fast field echo (b-FFE) technique, this
sequence is prone to magnetic field inhomogeneity and is frequently disturbed
by banding artifacts due to intra-voxel signal dephasing with increased severity
at 3.0T. This study proposed a RF phase-cycling technique as an addition to the
traditional b-FFE sequence (b-FFE-XD) to suppress the potential banding
artifacts in lumbosacral plexus imaging.Method
Ten healthy volunteers (4 women, and 6 men,
mean age, 53.30±8.73
years; age range, 31-64 years) underwent B-FFE-XD and B-FFE scans (Coronal
orientation) at 3.0T (Ingenia Elition, Philips healthcare, Best, the
Netherlands) on the lumbosacral plexus with a body coil for RF transmitting and
an anterior coil for signal receiving. The detailed acquisition parameters for
both sequences were kept the same, listed in Table1 as below, except that a RF
phase cycling scheme ([0,180+1*X, 2*X, 180+3*X,4*X,…], where X=n*360/NSA, and n
is the nth average) was applied to the excitation pulse for b-FFE-XD.
Images were evaluated independently by two radiologists (>3 years radiology
experience) in terms of overall image quality, banding artifact suppression,
and diagnostic certainty using a 5-point scale criteria (5 for the best result,
the scoring system was listed in Table2).
Consistency of the evaluation results
between the two observers were assessed using intra-class correlation
coefficients (ICC) in SPSS (IBM). If the consistency was good, the scores were
averaged between the two observers and then used for the subsequent analysis. The
scores of the two sequences were compared using a Wilcoxon signed-rank test
with a significance threshold of P < 0.05.Results
Effective banding artifact suppression was
achieved by B-FFE-XD, representative images shown in Figure 1. Good agreement
was achieved between the scoring of the two observers, and the ICC of banding
reduction was 0.765 (P<0.001).
The scores were significantly
higher in b-FFE-XD than in b-FFE in all items (overall image quality,
diagnostic confidence, black band artifact) under scoring (4.35
0.67
v.s. 3.95
0.60,
4.70
0.57,
v.s. 3.85
0.49,
4.80
0.70
v.s. 3.40
0.60,
respectively; all P<0.05), as shown in Table 3. Discussion and Conclusion
Our results demonstrated the potential
clinical application of the proposed RF phase cycling scheme for lumbosacral
plexus imaging, which yielded improved image quality and reduced banding
artifacts when compared to the traditional b-FFE technique.Acknowledgements
N/AReferences
[1] F. J. Hans, M. H. Reinges, T. Krings.
Lumbar nerve root avulsion following trauma: balanced fast field-echo MRI.
Diagnostic neuroradiology. 2004; 46:144-147
[2]
Ethan A. Neufeld, Peter Yi Shen, etc. MR imaging of the lumbosacral plexus: a
Review of techniques and pathologies. J Neuroimaging. 2015; 25:691-703