Caspar Florin1 and Jürgen Finsterbusch1
11Department of Systems Neuroscience, University Medical Center Hamburg–Eppendorf, Hamburg, Germany
Synopsis
Inner field-of-view EPI provides a good
image quality for diffusion-weighted imaging of the spinal cord in healthy
subjects but suffers from severe artifacts in the vicinity of metallic implants
that may be present in patients with traumatic injuries. Here,
inner-field-of-view imaging based on cross-sectional RF pulses is used to
obtain a single-shot fast spin echo technique that combined with multi
acquisition variable-resonance imaging (MAVRIC) and view angle tilt (VAT) offers a more
robust access to small target regions close to metallic implants and, thus, may
be feasible for diffusion-weighted imaging of patients with traumatic spinal
cord injury.
Introduction
Inner field-of-view EPI, e.g. (1,2), provides a good image quality for
diffusion-weighted imaging (DWI) of the spinal cord in healthy subjects. But it
suffers from severe artifacts in the vicinity of metallic implants making it
unfeasible for most patients with traumatic injuries. Fast-spin-echo (FSE) imaging,
in particular when combined with multi acquisition
variable-resonance imaging (MAVRIC)(3) and view angle tilt (VAT)(4) is much more
robust, however, may not be applicable in a single-shot as desired for DWI.
Here, inner-field-of-view imaging based on cross-sectional RF pulses is combined
with FSE imaging to enable single-shot imaging that in combination MAVRIC and
VAT can provide a much better image quality close to metallic objects than EPI
and, thus, may be feasible for diffusion-weighted imaging of patients with
traumatic spinal cord injury.Materials and Methods
The basic pulse sequence (Fig. 1) has been derived
from a standard fast-spin-echo sequence. An extra refocusing section is
inserted prior to the readout interval that could be used to apply diffusion
weighting and is tilted by an angle ɸ compared to the
image plane. The latter allows to reduce the FOV in the phase-encoding
direction without aliasing and can shorten the echo train accordingly to make
it feasible for single-shot acquisitions. To introduce the
frequency-selectivity required for MAVRIC, the initial RF excitation is applied
without slice-selection gradient pulse but different frequency offsets that
differ between subsequent acquisitions of the same slice; the same frequency offsets
are also considered for all subsequent slice-selective RF pulses and add to the
frequency required for the conventional slice selection. The readout is
performed with view angle tilt, i,e. a gradient pulse in the slice direction
during the data acquisition.
Measurements were
performed on a 3T whole-body MR system (Siemens PrismaFit) using a 64 channel head-neck
coil and a 32 channel spine coil on phantoms with metallic objects (four aluminum screws; see
Fig. 2) and a volunteer with a metallic implant from whom informed consent was
obtained prior to the examination. A resolution of 1.0x1.0x4.0mm³ covering a
FOV of 32x128 mm² was used yielding an echo spacing of 7.4ms, an echo time of
35ms for an ascending sampling for the FSE. With a frequency width of the RF
pulses of about 1000Hz,
the MAVRIC frequency range between +10kHz and -10kHz for phantom measurements
and 2000 Hz and -2000Hz in vivo was sampled in steps of 500Hz in an interleaved
order. All frequencies were stepped through for a slice with a TR of 900ms yielding
a total acquisition time per slice of 1.13min for phantom and of 8.1s for in
vivo measurements before continuing with the next slice. A maximum intensity
projection across frequency offsets was used to calculate a single image for
each slice on the fly. The EPI measurement yielded a total measurement duration of 5 s
with a TR of 4200 ms and a TE of 65 ms. Results and Discussion
Figure 2 summarizes results obtained in the phantom with metallic
objects (four aluminum screws). Signal losses close to and even at a distance
from the screws are obvious in the localizer image (Fig. 2a) and within the FSE acquisitions without MAVRIC
(Fig. 2b). With MAVRIC, significant signal can be regained and almost
the full phantom becomes visible, although with some residual interferences and
remaining signal losses. These images underline the significant influence of
even smaller metallic objects and the capability of the MAVRIC approach to
recover signal.
In figure 3 measurements of a volunteer with metallic implants are
presented. Thereby the cancellations
due to the implants are apparent in the localizer (a). Further in Fig 3
transversal in vivo measurements with inner FOV EPI (b) and inner FOV FSE (c)
are shown. In case of the EPI images the spinal cord is nearly not identifiable, the
distortions due to the metallic screws and plates are too strong. With
the combination of FSE and MAVRIC the spinal cord can be observed
and in nearly all slices parts of it are recognizable, even though the SNR is
rather low. The image quality difference in between FSE and EPI is due to the different
refocusing techniques used, RF refocused and gradient refocused, respectively,
as the least reacts more prone to inhomogeneities. The reason for the varying
measured frequency ranges during the phantom and in vivo acquisitions are
due to the different susceptibilities of the implant materials. The
susceptibility of aluminum, as used for the phantom, is by a factor of
10² larger compared to titanium as present in the volunteer and thus leads to stronger distortions.Conclusion
The combination of FSE with ZOOM and MAVRIC proves to be a tool with
which the image acquisition near metallic artifacts can be improved
significantly. Still work has to be done to accelerate the acquisition time. A
possibility to accelerate the image acquisition could lie in combination of
MAVRIC with simultaneous multi-slice imaging.Acknowledgements
This
work was supported by a grant of Wings for LifeReferences
[1] Saritas EU. DWI of the spinal cord with
reduced FOV single-shot EPI. MRM, 2008;60:
468-473
[2] Wheeler-Kingshott CA. ADC mapping of the human optic nerve:
Increased resolution, coverage and reliability with csf-suppressed ZOOM-EPI,
MRM, 2002 (47) 24-31
[3] Koch KM et al., Imaging
near metal with a MAVRIC-SEMAC hybrid. Magn Reson Med, 2011 (65) 71-82
[4] Cho ZH et al Total inhomogeneity correction
including chemical shifts and susceptibility by view angle tilting, A. A. Phys.
In Medicine 1988, (15), 7-11