Susanne Rauh1, Oliver Maier2, Oliver Gurney-Champion3, Melissa Hooijmans3, Rudolf Stollberger2,4, Aart Nederveen3, and Gustav Strijkers1
1Department of Biomedical Engineering and Physics, Amsterdam UMC, location AMC, Amsterdam, Netherlands, 2Institute of Medical Engineering, Graz University of Technology, Graz, Austria, 3Department of Radiology and Nuclear Medicine, Amsterdam UMC, location AMC, Amsterdam, Netherlands, 4BioTechMed-Graz, Graz, Austria
Synopsis
Model-based reconstruction can overcome many short-comings of
IVIM and DTI by accelerating the acquisition, suppressing systematic errors
from Rician noise and improving SNR. In this study we assess the feasibility of
model-based reconstruction to obtain IVIM and combined IVIM-DTI parameter maps
in the liver and kidneys in healthy volunteers. Results were compared to a
conventional IVIM and IVIM-DTI least-squares fitting. Model-based
reconstructions produced maps with less artifacts in liver and more details in
the kidneys as compared to those from conventional fitting. Mean parameter values
were similar for both methods.
Introduction
Diffusion tensor imaging (DTI) and intravoxel incoherent
motion (IVIM) yield detailed information about tissue microstructure and can
improve our understanding of the pathophysiology of several diseases1,2. In well-perfused organs with ordered microstructure, such as the kidneys, a
combination of both DTI and IVIM is desirable. However, conventional IVIM, DTI
and the combination of both face some challenges: (1.) Protocols suffer from
long acquisition times due to the need for many b-values and/or diffusion
directions; (2.) Fits from noisy (diffusion-weighted) images result in noisy
parameter maps; (3.) Rician noise introduces a noise-dependent bias in
parameter values.
Model-based reconstruction has the potential to overcome these
shortcomings. It facilitates parameter estimation directly from k-space by
combining the model and the MRI signal generation in the reconstruction
process. Higher undersampling factors can be obtained while simultaneously
preserving image quality3–5. Furthermore, as noise is
Gaussian in k-space, no systematic noise-bias is introduced.
The aim of this work was to assess the feasibility of
model-based reconstructions for pure IVIM and combined IVIM-DTI fitting in
well-perfused abdominal organs.Methods
All scans were performed with a 3T MRI (Philips Ingenia,
Best, The Netherlands) on healthy participants. Raw and image data were
exported. Liver data was acquired in three volunteers, kidney data in two. Diffusion-weighted EPI scan parameters were as followed: Liver scan
in free-breathing, TE/TR=63ms/7000ms, 27 slices, 3x3x6mm³ (1 volunteer: 3x3x5mm3),
1mm slice gap, SENSE factor 2, 18 b-values (range 0-700s/mm²), 3 diffusion
directions each, 9 b=0s/mm² scans. Kidneys: navigator-triggered scan, 15-18
slices to cover whole kidneys, 2.5x2.5x5mm³ resolution, 0.3mm slices gap, SENSE
factor 2, TE/TR=62ms/982ms, 10 b-values (range 0-600s/mm²), 32 diffusion directions
for high b-values (200-400s/mm²), 6 directions for lower b-values, 16 b=0s/mm²
scans.
Model-based reconstruction was performed on the raw data
using the PyQMRI Toolbox6 written in Python after
applying basic corrections. An IVIM and IVIM-DTI model were implemented. PyQMRI
uses an iteratively regularized Gauss-Newton approach with total generalized
variation regularization7. In this work 7
Gauss-Newton iterations were used. An illustration of the conventional fitting
and model-based reconstruction process is shown in Figure 1.
For comparison the image data obtained from the scanner was
fitted voxel-wise to an IVIM or IVIM-DTI model in Python using the DiPy8 toolbox.
The IVIM-DTI fit was performed in two steps: first, the IVIM fit was performed,
yielding the perfusion fraction f and pseudo-diffusion coefficient D*; then the
perfusion component was subtracted from the signal and the diffusion tensor was
fitted to the remaining signal. The liver data was fitted to a pure IVIM model:
$$S(b)=S_0\cdot(f\cdot e^{-bD^{*}}+(1-f)\cdot e^{-bD})$$
S(b) denotes the signal at given b-value b, S0 is
the signal at b=0s/mm², D is the diffusion coefficient, D* pseudo-diffusion
coefficient and f perfusion fraction. Kidney data was fitted to a combined
IVIM-DTI model with diffusion tensor $$$\underline{\underline{D}}$$$ and diffusion gradient orientations $$$\underline{g}$$$:
$$S(b)=S_0\cdot(f\cdot e^{-bD^{*}}+(1-f)\cdot e^{-\underline{g}^{T}\underline{\underline{D}}\underline{g}})$$
From
the diffusion tensor the mean diffusivity (MD) and fractional anisotropy (FA)
were calculated. Regions of interest (ROIs) were drawn on the DiPy output
S0 maps in all slices covering the liver or kidney. ROIs of the kidney
were drawn in the cortex only. Outcome measures are shown as mean values and
standard deviation over the full segmented volume.Results
Figure 2 shows the
value of the cost function after each Gauss-Newton iteration for the IVIM and
IVIM-DTI model. The largest decrease of the residuals was observed in the first
three iterations for each dataset and model. After five iterations the cost
function reached a plateau and more iterations did not further decrease the
residuals substantially.
Representative parameter maps are shown in Figure 3 for the
liver and Figure 4 for the
kidneys. The liver maps obtained from the conventional fit had a smooth
appearance, but displayed abnormal regional heterogeneity, such as a high
perfusion anterior and high D in the left liver lobe – a known artefact from
cardiac motion. The model-based reconstructions seems to be unaffected by these
unexpected heterogeneities. For the kidneys, the model-based reconstructions yielded
consistent parameter maps, which revealed more anatomical detail than those
obtained with conventional fitting. This was particularly observed in the perfusion
map (Figures 4e and 4f).
A quantitative comparison of the mean ROI parameter values
obtained with model-based and conventional fitting is shown in Figure 5. The mean
values and standard deviations were similar. With model-based reconstruction, D was slightly lower in the liver, which could have resulted from
cardiac motion artificially increasing D in the conventional fit. In the
kidney, MD was slightly higher for model-based fitting, which is consistent
with proper handling of Gaussian noise. The model-based reconstruction yielded lower
f values in both organs and all datasets. An interesting observation was that
the mean and standard-deviation of D* were much lower with model-based reconstruction. Discussion
Model-based reconstruction for IVIM and IVIM-DTI fitting resulted
in good-quality parameter maps. These initial results could help to overcome
the challenges in conventional IVIM and IVIM-DTI. We plan to scan more subjects
to further optimize regularization parameters and investigate the robustness of
the reconstructions. Additionally, we plan to include image acquisition acceleration
by undersampling k- and diffusion-space. Conclusion
We showed the feasibility of IVIM and combined IVIM-DTI
model-based reconstructions in liver and kidney. Acknowledgements
No acknowledgement found.References
1.
Notohamiprodjo M, Glaser C,
Herrmann KA, et al. Diffusion tensor imaging of the kidney with parallel
imaging: Initial clinical experience. Invest Radiol. 2008;43(10):677-685. doi:10.1097/RLI.0b013e31817d14e6
2. Wei Y, Huang Z, Tang H, et al. IVIM improves preoperative assessment of microvascular invasion in HCC. Eur
Radiol. 2019;29(10):5403-5414. doi:10.1007/s00330-019-06088-w
3. Doneva M, Börnert P, Eggers
H, Stehning C, Sénégas J, Mertins A. Compressed sensing reconstruction for
magnetic resonance parameter mapping. Magn Reson Med.
2010;64(4):1114-1120. doi:10.1002/mrm.22483
4. Roeloffs V, Wang X, Sumpf
TJ, Untenberger M, Voit D, Frahm J. Model-based reconstruction for T1 mapping
using single-shot inversion-recovery radial FLASH. Int J Imaging Syst Technol. 2016;26(4):254-263.
doi:10.1002/ima.22196
5. Maier O, Schoormans J, Schloegl M, et al. Rapid T 1 quantification from high resolution 3D data with
model‐based reconstruction. Magn Reson Med. 2019;81(3):2072-2089.
doi:10.1002/mrm.27502
6. Maier O, Spann SM, Bödenler
M, Stollberger R. PyQMRI : An accelerated Python based Quantitative MRI
toolbox. J Open Source Softw. 2020;5(56):2727. doi:10.21105/joss.02727
7. Bredies K, Kunisch K, Pock
T. Total generalized variation. SIAM J Imaging Sci. 2010;3(3):492-526.
doi:10.1137/090769521
8. Garyfallidis E, Brett M,
Amirbekian B, et al. Dipy, a library for the analysis of diffusion MRI data. Front
Neuroinform. 2014;8(FEB):1-17. doi:10.3389/fninf.2014.00008