Dominik Ludwig1,2, Frederik B. Laun3, Karel D. Klika4, Mark E. Ladd1,2,5, Peter Bachert1,2, and Tristan A. Kuder1
1Medical Physics in Radiology, German Cancer Research Center (DKFZ), Heidelberg, Germany, 2Faculty of Physics and Astronomy, Heidelberg University, Heidelberg, Germany, 3Institute of Radiology, University Hospital Erlangen, Friedrich-Alexander-Universität Erlangen-Nürnberg (FAU), Erlangen, Germany, 4Molecular Structure Analysis, German Cancer Research Center (DKFZ), Heidelberg, Germany, 5Faculty of Medicine, Heidelberg University, Heidelberg, Germany
Synopsis
Diffusion
pore imaging (DPI) can be used to retrieve the pore space function of arbitrary
closed pores. In this study, we show that DPI of glass capillaries is possible
even under difficult experimental conditions. By separating the long gradient
into a CPMG-like gradient echo train, matching the magnetic susceptibility and
adding an additional filter diffusion weighting, it was possible to acquire
diffraction patterns of glass capillaries that were placed orthogonally to the
main magnetic field. Furthermore, the feasibility of doing DPI in the presence
extraporal water using our filtered approach was demonstrated.
Introduction
Diffusion
pore imaging enables the direct measurement of the shape of arbitrary closed
pores that are filled with an NMR-detectable medium [1-3]. The first
implementation of diffusion pore imaging on an NMR spectrometer [1] made use of
the long–narrow approach [4] in order to retrieve the phase information
necessary for reconstruction of the pore space function. To achieve correct
measurements for 20 µm capillaries, a CPMG train of 180° pulses was used in
[1], where the long gradient with low amplitude was split into many short
pulses. Also, the short high-amplitude gradient at the end of the pulse train
was spilt into two parts separated by a 180° pulse. More recently, it has been
shown that it is also possible to use a spin echo approach with a readout
gradient in order to counteract gradient imperfections [3]. In last year’s
ISMRM abstract [5] we showed, that it is also possible to acquire pore functions
in the presence of extraporal fluid. The aim of this study was to further
improve this method by adding an additional diffusion weighting acting as a
filter, similarly as in [6] for the faster-diffusing extraporal compartment,
which in turn significantly improves the outcome.Methods
Glass
capillaries of 3.5 mm
length and inner diameter of 10 µm were
stacked horizontally inside a
5 mm NMR tube, i.e. with orthogonal orientation
to the main magnetic field. The capillaries were filled with a saturated
aqueous NaCl solution – necessary to weaken the susceptibility effects
introduced by the glass capillaries – resulting in a free diffusion coefficient $$$D=1.48$$$ µm2/ms
for the water inside the capillaries at $$$T=298$$$ K. A schematic representation
of the pulse sequence used for the experiments is presented in Figure 1. The
long gradient of the long–narrow approach was split into a CPMG-like gradient
pulse train separated by inserting
180° pulses.
The long gradient was split into
11 segments of $$$\delta_\text{l}=10$$$ ms
each, yielding a total time of
$$$\delta_\text{L}=110$$$ ms. The
narrow gradient had a duration of $$$\delta_\text{S}=2.3$$$ ms with $$$G_{\text{S,max}}=6.3$$$ T/m, resulting
in $$$q_{\text{max}}=\gamma \delta_\text{S} G_{\text{S,max}}=3680$$$ mm-1.
q-Values
and the sampling of the gradient ramps had to be selected in order to match the
available 16-bit raster of the gradient amplifier for all experiments. This
resulted in small individual adjustments to
for the different q-values in order to
counterbalance these limitations. The necessary phase cycling for the
different 180° pulses is shown in Table 1. The filter consisted of 8 gradient
pulses with a duration of 10 ms each, generating
$$$b_{\text{filter}}=2750$$$ s/mm². The signal
decay in the intrapore space was simulated with a matrix-based approach as in [7].Results
Figure 2 shows
the measured signal for both an acquisition with and without the proposed
filter as the function of the q-value as well as simulation data. Signal for
the measurement with filter was normalized to $$$S(q=0)$$$, whereas the signal of the
measurement without the filter was normalized to $$$S(q=1102$$$ mm-1$$$)$$$(minimum of the
signal curve) of the measurement with the filter, in order to make them
comparable. It is seen that the experimental results, when using the method
with the additional filter, agree well with the matrix-based simulation. The
corresponding pore space function, calculated via Fourier transform of the
signal in Figure 2, is shown in Figure 3, which also yields the expected
diameter of 10 µm. For the measurement without the additional filter, it is
obvious that the signal is mainly governed by extraporal signal for the two lowest
q-values, while its contribution can be nearly completely suppressed with our
method. For higher q-values, both measurements are in good agreement as the
high q-values of the long-narrow sequence suppress the extraporal signal
contributions themselves.Discussion and Conclusion
The need for strong gradients is the main
limiting factor of DPI. Nevertheless, we were able to demonstrate the
feasibility of DPI with an ultra-high gradient system using 10 µm capillaries
without exploiting the full gradient strength available. Furthermore, we were
able to show that diffusion pore imaging is possible in the presence of water
outside the pores without the need of extrapolation or fitting of the signal
curve. Further, a very high extraporal signal fraction was used here; in
biological samples, this problem will be significantly reduced. Compared to the
initial approach of signal extrapolation [5], results are highly improved by
implementing a bipolar gradient pair, which is split up into segments, in the
first part of the CPMG train. By using this gradient pair as a filter for the
water inside the pores that is more restricted, it is possible to suppress the
surrounding water signal. This procedure enables the direct measurement of
smaller q-values instead of extrapolating them. Therefore, this work enables
the application to µm-sized objects, which have dimensions as living cells in a
realistic environment, i.e. in their natural surroundings.Acknowledgements
Financial support by the DFG (Grant No. KU 3362/1-1) is gratefully
acknowledged.References
[1]: Hertel, Stefan, Mark Hunter, and
Petrik Galvosas. "Magnetic resonance pore imaging, a tool for porous media
research." PRE 87.3 (2013): 030802.
[2]: Hertel, Stefan Andreas, et al.
"Magnetic-resonance pore imaging of nonsymmetric microscopic pore
shapes." PRE 92.1 (2015): 012808.
[3]: Bertleff, Marco, et al. "1D
and 2D diffusion pore imaging on a preclinical MR system using adaptive
rephasing: Feasibility and pulse sequence comparison." JMR 278
(2017): 39-50.
[4]: Laun, Frederik Bernd, et al.
"Determination of the defining boundary in nuclear magnetic resonance
diffusion experiments." PRL 107.4 (2011): 048102.
[5]: Ludwig,Dominik,
et al. "Water diffusion pore imaging on a 14.1 T spectrometer using glass
capillary phantoms in the presence of extraporal fluid." Proc. Intl. Soc. Mag.
Reson. Med. 28: 4496, (2020)
[6]:
Dhital, Bibek, et al. "PFG Filter for Oscillating
Gradient Diffusion Measurements." Proc. Intl. Soc. Mag. Reson. Med. 22 (2014)
[7]: Grebenkov, D.S. "Laplacian
eigenfunctions in NMR. I. A numerical tool." Concepts Magn. Reson., 32A: 277-301 (2008)