Koji Sakai1, Yasuhiko Tachibana2, Toshiaki Nakagawa3, Hiroyasu Ikeno3, Takayuki Obata2, and Kei Yamada1
1Kyoto Prefectural University of Medicine, Kyoto, Japan, 2National Institute of Radiological Sciences, Chiba, Japan, 3Kyoto Prefectural University of Medicine Hospital, Kyoto, Japan
Synopsis
We compared DTI measures among three different anisotropic diffusion
phantoms: a commercially available astriction cotton; a glass capillary plate;
a hand-bundled polyethylene fibers. The purpose of this study was to examine whether the cotton is useful as
an anisotropic diffusion phantom. Differences
in ADC and FA were
evaluated by comparing coefficients of variation (CVs).
No significant
difference was seen for the largest eigen value: 2.06 - 2.14 x 10-3 mm2/sec.
Clear differences in FA were seen among the three DTI phantoms (p = 0.0079). CVs
from five acquisitions for ADC and FA from every phantoms were all less than
2%.
INTRODUCTION
In
general, a large data set gives larger statistical power for study results. While
collecting large data sets is appropriate for research, accomplishing this
using only a single magnetic resonance (MR) scanner is not possible. As a
result, multi-scanner studies are often needed to fulfill the statistical
requirements of MR studies. Nevertheless, intrinsic differences among scanners have
been recognized as inevitable. The Quantitative Imaging Biomarkers Alliance (QIBA)
was established to address these intrinsic differences and obtain statistically
meaningful results from multi-scanner studies 1. Isotropic diffusion
has already been discussed by the QIBA 2. However, anisotropic
diffusion has not been included among QIBA endeavors because of the absence of
appropriate anisotropic diffusion phantoms. A suitable phantom needs to satisfy
these points: stable measurement of values from diffusion
tensor imaging (DTI); cost-effectiveness; and availability. Anisotropic
diffusion phantoms have been used, including the solid capillary type 3-5 and bundled fiber type 5, 6. However, these phantoms have various drawbacks.
The solid
capillary type shows only
hindered diffusion and is expensive, and unsuitable for widespread use. On the other hands, most bundled fiber-type phantoms are handmade, and preparation of multiple
identical pieces for industrial application is difficult. This
study compared the DTI values of apparent diffusion coefficient (ADC), eigen
value (λ), and fractional anisotropy (FA) among three different anisotropic
diffusion phantoms: commercially available astriction cotton (AC)7; a
glass plate (CP)5; and hand-bundled polyethylene
fibers (Dy)5. The purpose of this study was
to examine whether AC is useful as an anisotropic diffusion phantom for
multi-scanner studies.MATERIALS & METHODS
Phantoms: AC (30 mm in length, 20 mm in diameter; Hakujuji
Co., Tokyo, Japan) in acryl tubes (diameter,
25 mm) was arranged in layers of parallel tubes arranged perpendicular
to the underlying layer (Figure 1a). The phantom was 12.5 cm wide x 17.0 cm long x 12.0 cm high (polyethylene container) 7.
The capillary
glass plate (Hamamatsu Photonics K.K., Shizuoka, Japan) was 9 mm in diameter
and 3 mm thick, with uniformly spaced micro-pores (diameter, 20 μm). Twenty
plates were stacked in a plastic tube with ultra-pure water (Figure 1b) 5. The polyethylene
fibers (Dynnema®; Toyobo Co., Osaka, Japan) were bundled together with a
thermal shrinkage tube 5. These bundled fibers were placed in a
plastic tube with ultra-pure water (Figure 1c). Table 1 summarizes
the features of the phantoms.
MR scanner: This study used a 3-T MRI system
(MAGNETOM Verio 3T; Siemens Healthcare K.K., Erlangen, Germany) equipped with a
12-channel head coil. The phantoms were placed in parallel with the Z-axis
direction of the MRI gantry.
Data acquisition: DTI was performed using b values of 0 and 1000 s/mm2
in 10 different diffusion-encoding directions. The
field of view was 240 x 240 mm. DTI was acquired using an EPI
technique with a matrix of 120 x 108. Of note,
diffusion time was automatically selected by the MRI system for this
additional scan, but was not displayed on the user interface as is usual for
clinical machines. Other parameters were: TR = 6300 ms; TE = 92 ms; average = 3; number
of b0 images = 1; number of readout
segments = 5; resolution = 2.0 × 2.0 × 2.0 mm; room temperature, approximately
22°C; and total scan time = 4 min. Five acquisitions were performed for each phantom.
DTI
analysis: DTI analysis was performed with Diffusion Toolkit/TrackVis
version 0.5.2.28, and hand-drawn regions of interest (ROIs) based
average diffusion metrics were calculated by 3D slicer 9.
Evaluation of stability: The
stability of diffusion metrics (ADC and FA) of the phantoms was evaluated by comparing
coefficient of variation (CV) as CV [%] = standard deviation / mean x 100. When CV was < 5%, the measurement was considered
homogeneous
10.
Statistics: Comparisons
between DTI measurements from different phantoms were performed
using the Wilcoxon rank-sum test (Matlab®; The MathWorks, Natick, MA). The
difference was evaluated as statistically significant for Values of p < 0.05 were evaluated as
statistically significant.RESULTS
Diffusivity: In Figure 2,
the average ADC of three different phantoms were 1.34 – 1.80 x 10-3 mm2/sec
(p < 0.05). In contrast, no
differences in largest eigen value were seen among the three DTI phantoms: 2.06–2.14
x 10-3 mm2/s (p: AC vs. CP = 0.15, CP vs. Dy = 0.095, Dy vs. AC = 0.15).
Anisotropic diffusion: FA
values of the three phantoms were 0.19–0.54 (Figure
3a). Clear differences in FA were seen among the three DTI phantoms (p = 0.008), attributed to significant differences
in radial diffusivity (p = 0.008; Figure 3b).
Inter-5-scan difference: CVs of ADC (Figure 4a) and FA (Figure 4b)
for the phantoms were all <2%. Scan-rescan diffusivity and anisotropy were thus stable
for all three phantoms.DISCUSSION & CONCLUSION
CP and
Dy showed high FA and are considered appropriate as anisotropic diffusion
phantoms. However, chemically etched glass plates are expensive and unsuitable for
widespread use. Since Dy is handmade, preparation of multiple
identical pieces for industrial application is difficult. In contrast, AC is a
mass-produced industrial material and inexpensive. AC is expected to solve the drawbacks of CP and Dy. AC shows relatively low FA, but the range of values observed in the living brain 11.Acknowledgements
This work was supported by JSPS KAKENHI Grant
Number JP17K10413.References
1. QIBA,
https://www.rsna.org/qiba/.
2. http://qibawiki.rsna.org/images/1/1b/QIBA_IMI_August_7_2013-M_Boss.pdf.
3.Yanasak N et al., Magnetic
Resonance Imaging. 2006; 24: 1349–1361.
4.Oida T et al., Magn
Reson Med Sci. 2011: 10(2); 121-128.
5. Tachibana
et al., Magn Reson Med Sci. 17(2018),
pp.251-258.
6.Lemberskiy G et al., NMR in Biomedicine. 2017;30:e3708
7. Sakai
et al., Proc. Intl. Soc. Mag. Reson. Med.
24 (2016), 2032.
8. Diffusion
toolkit, https://www.nitrc.org/projects/trackvis/.
9. 3D
slicer, https://www.slicer.org/
10. Kamagata
et al., Magn Reson Med Sci. 2015; 14:
227-233.
11. Alexander
AL et al., Neurotherapeutics, 2007;
4: 316-329.