Stephen E Russek1, Kathryn E Keenan1, Karl F Stupic1, Teryn S Wilkes2, Ramesh Karki3, and Todor Karaulanov4
1NIST, Boulder, CO, United States, 2Intermountain Neuroimaging Consortium, University of Colorado, Boulder, CO, United States, 3University of Colorado Anschutz, Radiological Sciences, Aurora, CO, United States, 4CaliberMRI, Inc., Boulder, CO, United States
Synopsis
Using
the MRI system phantom we demonstrate the utility of the embedded MR-readable
thermometer to measure and correct for temperature variations, demonstrate the
use of the fiducial array to characterize scanner geometric distortions and the
efficacy of software distortion corrections, and finally look at the effect of
the electromagnetic properties of the phantom fill on quantitative
measurements.
INTRODUCTION
NIST
and the ISMRM ad-hoc Committee on Standards for Quantitative Magnetic Resonance,
designed and commercialized a standard MRI system phantom1,2 to assess scanner performance, stability,
inter-comparability and accuracy of quantitative relaxation time imaging3-5. Several important measurement protocols need to be
improved and made more rigorous. The items addressed here are: 1) using the
embedded liquid crystal (LC) MR-visible thermometer6 to automatically record phantom temperature; 2) ascertaining
efficacy of geometric distortion corrections; and 3) determining the effects of
the fill conductivity on the MR measurements.METHODS
Two commercial MRI system phantoms were
purchased7, along with reference
solutions. The solution T1, T2 relaxation times were traceably
calibrated at 3T over a temperature range from 16C to 26C. Geometric accuracy
was calibrated using traceable CT. The system phantoms are included in the
NIST/ NIBIB phantom lending library https://www.nist.gov/programs-projects/nistnibib-medical-imaging-phantom-lending-library and
are available for check-out. The
phantoms were characterized on a 3T scanner using a 32-channel head coil with
the plates in the sagittal plane (Figure 1A), a 90° rotation of the phantom
from its default orientation. All data were analyzed using an open-source
package PhantomViewer (https://github.com/MRIStandards/PhantomViewer).
The LC thermometer consists of 10 cylindrical cells with transition temperatures
varying in 1C increments from 15C to 24C. Circular 4mm diameter regions of interest
were used to collect signal data for thermometry. Geometric distortion analysis
was done by fitting the fiducial array, consisting of 56 precisely-machined
10.0mm spheres on a 40.0mm grid (1). The phantoms were imaged with both a deionized
water fill and a phosphate-buffered saline fill with 20C conductivities of
< 10-4S/m and 0.43S/m, respectively. The saline conductivity was
chosen to be comparable to the conductivity of human tissue. RESULTS
Figure
1 B,C,D show the system phantom configuration, the top plate with serial number,
and the LC thermometer. The LC
thermometer shows bright/dark cells where the LC critical temperatures are below/above
the phantom temperature, respectively. The
on-cells have a small ~ 1mm chemical shift artifact along the frequency-encode
direction. Figure 2A,B show the LC thermometer at two points during the
scanning session indicating the phantom is warming. Figure 2C shows the MR signal for each cell
for three isotropic 3D scans taken at different times during the imaging
session. The temperature can be determined by using a power law fit to the
data, or by manually taking the midpoint between the last on-cell and the first
off-cell critical temperature. These phantom temperatures are shown in Fig. 2D
along with Pt-resistance thermometer measurements.
Figure
3A,B show geometric distortion, defined as the difference of the apparent
position of each fiducial sphere from the real position, obtained from a 3D GRE
scan with and without software-based distortion corrections. The scans were
done in the plane of the phantom plates, with the y-direction being
perpendicular to the plates and scan plane. The software correction improves
the in-plane distortion to < ±0.4mm and < ±0.2mm perpendicular and
parallel to the scanner bore, respectively. The distortion correction applied
here did not correct for the out of plane distortions.
The
change in fill causes a small shift in the image signal distribution across the
phantom (Figure 4), with signal loss in the center of the phantom for the
saline fill. A change in MR signal can also
be observed for the proton density array (Figure 5) for the saline and water
fills. The primary measurement, MR proton density, where the signal is
normalized to the local background signal is not strongly affected. The biggest
effect of the fill conductivity was observed in the variable flip angle (VFA) T1
protocol (Figure 5C). VFA is sensitive
to nonuniformities in the RF transmit field which will be a function of the
electromagnetic properties of the phantom fill. Other effects, including
prescan shimming and RF power transmit/receive gains may have important effects
which may lead to much greater variability in VFA. DISCUSSION
The
MR-readable thermometer allows the temperature to be embedded in the MR data at
scan time and provides a more reliable method of recording phantom
temperature. The thermometer accuracy is
±0.5C, which is sufficient to push the thermal uncertainty for T1, T2
measurements on the NiCl2 array below 2%. The geometric distortion
measurements are quick and useful ways to understand the intrinsic scanner
accuracy due to nonuniform gradients as well as to verify that the software
corrections are adequate. Figure 3B shows
that current scanners can have geometric distortion less than 1 part in 103,
hence calibration standards must exceed this accuracy. The effect of fill has a
small effect on primary measurements but may have a considerable effect on less
robust protocols such as T1 VFA. The effect and optimization of the
electromagnetic properties of the phantom fill need to be further explored.CONCLUSION
Here we describe
improvements and applications of the MRI system phantom including the
incorporation of an MR-readable thermometer, simple evaluation of scanner
geometric distortion and efficacy of distortion corrections, and the effect of
varying the electromagnetic properties of the fill solution. While the MRI system phantom has limitations
and considerable complexity, it provides necessary information and validation
for many types of quantitative measurements. Acknowledgements
We thank all of the members of the ISMRM SQMR committee for input and guidance. We thank Nikki Rentz for long hours doing traceable NMR calibrations. References
1. Russek SE, Boss M, Jackson EF, Jennings DL, Evelhoch JL, Gunter JL, Sorensen AG. Characterization of NIST/ISMRM MRI System Phantom. Proc Intl Soc Mag Reson Med 20 2012:2456.
2. Keenan KE, Ainslie M, Barker AJ, Boss MA, Cecil KM, Charles C, Chenevert TL, Clarke L, Evelhoch JL, Finn P, Gembris D, Gunter JL, Hill DLG, Jack Jr. CR, Jackson EF, Liu G, Russek SE, Sharma SD, Steckner M, Stupic KF, Trzasko JD, Yuan C, Zheng J. Quantitative magnetic resonance imaging phantoms: A review and the need for a system phantom. Magnetic Resonance in Medicine 2018;79(1):48-61.
3. Bane O, Hectors SJ, Wagner M, Arlinghaus LL, Aryal MP, Cao Y, Chenevert TL, Fennessy F, Huang W, Hylton NM, Kalpathy-Cramer J, Keenan KE, Malyarenko DI, Mulkern RV, Newitt DC, Russek SE, Stupic KF, Tudorica A, Wilmes LJ, Yankeelov TE, Yen Y-F, Boss MA, Taouli B. Accuracy, repeatability, and interplatform reproducibility of T1 quantification methods used for DCE-MRI: Results from a multicenter phantom study. Magnetic Resonance in Medicine 2018;79(5):2564-2575.
4. Keenan KE, Gimbutas Z, Dienstfrey A, Stupic KF. Assessing effects of scanner upgrades for clinical studies. J Magn Reson Imaging 2019;50(6):1948-1954.
5. Jiang Y, Ma D, Keenan KE, Stupic KF, Gulani V, Griswold MA. Repeatability of magnetic resonance fingerprinting T1 and T2 estimates assessed using the ISMRM/NIST MRI system phantom. Magn Reson Med 2017;78(4):1452-1457.
6. Keenan KE, Stupic KF, Russek SE, Mirowski E. MRI-visible liquid crystal thermometer. Magn Reson Med 2020;84(3):1552-1563.
7. CaliberMRI. Web site. https://qmri.com/. 2020.