Andreas Georg Berg1,2, Stefan Hengsbach3, and Klaus Bade3
1Center for Medical Physics and Biomedical Engineering High field MR-Center, Medical University of Vienna, Vienna, Austria, 2High-Field MR-Center (MRCE), Medical Universits of Vienna, Vienna, Austria, 3Karlsruhe Nano Micro Facility (KNMF), Karlsruhe Institute of Technology (KIT), Leopoldshafen-Eggenstein, Germany
Synopsis
Quality-control for
systematic improvements in 3D-isotropic spatial resolution of the MR-imaging
system up to the microscopic range becomes increasingly relevant not only for
preclinical imaging but also for High-Field-MR human scanners. The design of a
prototype phantom, comprising a set of 3D-cubes and an exemplary MR-microscopic
evaluation is presented. The extraordinary challenges on accurate bar-width-to-cavity
ratios for the cube range from periodicity a=128µm down to a=4µm are dealt with
using 2-photon-lithography as manufacturing technology. The spatial resolution
for 3D isotropic imaging can be checked qualitatively and quantitatively by
interpolation of the Modulation-Transfer-Function demonstrated on a 3D-GRE
sequence as an example.
Purpose/Introduction
The most important
criterions for image quality in MR-imaging are related to the
contrast-to-noise-ratio for the specific object and the spatial resolution of
the MR imaging system. High spatial resolution based on voxel- and pixel-sizes (ps) in
the microscopic range (ps < 100 x 100 µm2), is often claimed not only
on preclinical animal and investigational vertical high-field MR-microscopy
devices but also recently even on human MR-scanners1. However the
spatial resolution is only ultimately limited by the voxel size. The actual achieved
spatial resolution might be significantly worse and consequently the
qualitative and quantitative check is recommended in several norms on quality
control of human MR-scanners based on one dimensional periodic sets of plates
or tubes2-4. The spatial resolution is dependent on several
different technical and sample specific features, e.g. signal-to-noise ratio
(SNR), bandwidth/pixel and gradient strength, line width (T2), susceptibility and
chemical shift differences, B0-homogeneity and gradient switching behavior. Thus the improvement
of hard- and software components of the MR-imaging system including hard- and
software is to be proved at best for an adapted non-degradable resolution
phantom presenting periodic structures in all of the three dimensions, not-only
for the in-plane phase or frequency encoding direction5, but also
the third slice-selective, phase or "super-resolution" encoding axis.
Such phantoms do not exist yet for the spatial microscopy range with voxel
sizes VS < 100 x 100 x 100 µm3.Subjects and methods
The design of the proposed reference resolution standard phantom for
isotropic 3D- MR-micro-imaging and
microscopy comprised a set of 3D- cavity containing cubes at identical bar-to-cavity
ratio (duty cycle dc = 1) with min. 4 periods (periodic length: a, see fig. 1) for each spatial frequency
ranging from a1 = 512 µm, down to a9 = 2 µm. Such
3D-structures would allow for checking spatial resolution from 2 to 500 line
pairs (lp) per mm. The demands on miniaturization and spatial accuracy for the
establishment of these reference objects at different characteristic structural
widths of more than 6 orders of magnitude (106) were very
challenging but could be (partly) fulfilled using 3D-Direct Laser writing
(3D-DLW), based on a 2-photon absorption for localised polymerisation in a
negative resist6. After 2 years of optimization we achieved a 3D-DLW
manufacturing protocol for a cube set comprising 9 cubes (ai = 128,
98, 64, 48, 32, 24, 16, 8, and 4 µm), each offering 4.5 periods for the
coverage of several modulations with excellent duty cycle (fig.1). This set of
cubes was positioned in a small tube containing silicon oil for
MR-visualization and checked for MR visibility. For demonstration of
suitability for quality control on spatial resolution in MR-µ-imaging and
MR-microscopy a prototype insert on a human High-field 7T scanner7
was used.Results
The 3D cube set allowed to quickly check the impact of several hardware,
pulse sequence and MR-protocol parameters, e.g. bandwidth/pixel, frequency-/phase encoding and nominal slice thickness, on the spatial resolution in all of
the 3 dimensions based on one MR-measurement data set. In most of the cases
there is a significant difference between voxel size and actual spatial
resolution. As an example the MR-microscopic image (sagittal view) resulting
from the 3D-data set of a Cartesian frequency and phase encoding high-resolution
3D-gradient echo sequence is shown in fig. 3. Whilst the cube-grid with a/2 =
48 µm is resolved, the inner structure of the cube with ai = 32 µm is not
detectable, though the nominal voxel size is less (31x31x30 µm3).
Moreover the 3D- MR-microscopic data set on the 3D- resolution phantom
also allows for the quantitative evaluation of the Modulation Transfer Function
(MTF) in all of the 3 spatial dimensions for different spatial encodings. A
quantitative criterion on the relative intensity modulation depth (Ir) might be
used (e.g. Irres = 50% Irmax)8. In the case of
the 3D-gradient echo (GRE) sequence this criterion, using a MTF fitting
function, resulted in an effective spatial resolution of ares/2 = 45
µm (11,1 line-pairs/mm).Discussion/Conclusion
Quantitative
quality control on acutal achieved spatial resolution with difference to voxel
size becomes increasingly important for the upcoming perspectives for MR-micro-imaging
and microscopic resolution even on human scanners. With difference to previous
investigations3,5 we propose a 3D-cube set concept covering the
spatial range from ai = 512 µm to 1 µm. We could realize partly this design
concept with a cube set ranging from ai = 128 µm down to ai = 4 µm using
3D-Direct Laser Writing (3D-DLW) based on the 2-photon initiated polymerization
process. As an example the spatial resolution for a 3D-GRE sequence with
nominal voxel VS ≈ (30 µm)3 is
investigated, the difference between voxel size and spatial resolution is
shown.
The proposed investigational 3D phantom set might
serve in future as basic concept for a standard on quantitative quality control
referring to one of the most important image quality criterions in MRI: the
spatial resolution.Acknowledgements
This project was supported by the
KNMF-project nr. 2015-013-006488; A. Berg, 2-3-Dimensional
μ-Structures for Magnetic Resonance Microscopy. References
1 Laistler, E., Dymerska, B., Sieg, J., Goluch,
S., Frass‐Kriegl, R., Kuehne, A. and Moser, E. In vivo
MRI of the human finger at 7 T. Magn. Reson. Med. 2018;79:588-592.
2 European Norm EN 61223-3-7 (IEC
62B/575/CDV) 2005.08.01. Evaluation and routine testing in medical imaging
Departments; Part 3-7: Acceptance and constancy tests – Determination of
essential image characteristics of magnetic resonance equipment.
3 American College of Radiology; MR
accreditation program; Testing Instructions, small phantom guidance. 2018. https://www.acraccreditation.org/-/media/ACRAccreditation/Documents/MRI/SmallPhantomGuidance.pdf.
Accessed December 11, 2020.
4 Fellner C, Muller W, Georgia J,
Taubenreuther U, Fellner FA, Kalender WA; A high-resolution phantom for
MRI; Magnetic Resonance Imaging 2001;19:899–904.
5 Berg A, Boerner M. Quantification of microscopic spatial resolution on MR-scanners using
micro-phantoms manufactured with Deep-X-ray Lithography. Proc.
Int. Soc. Mag. Reson. Med. 2020;28:4267.
6 Hengsbach
S, Lantada AD. Rapid prototyping of multi-scale biomedical microdevices by
combining additive manufacturing technologies. Biomed Microdevices. 2014;16(4):617-27.
7 Berg A, Potthast A, Starewicz P.
MR-MICROSCOPY ON A HUMAN 7T-SCANNER; Proc. ISMRM 2010: progr nr. 1048.
8 Lerski RA, de Certaines JD. II
Performance assessment and quality control in MRI by Eurospin test objects and
protocols. Magnetic Resonance Imaging 1993;11(6):817-833.