Claudiu Schirda1
1Radiology, University of Pittsburgh School of Medicine, Pittsburgh, PA, United States
Synopsis
We propose using a gradient readout, fast MRSI
acquisition and a support or stand-alone 3D multi-echo, 3-point Dixon GRE scan for
MRI scanner and coil quality assurance (QA). The fast MRSI acquisition with the
support 3D GRE scan uses the BRAINO phantom, takes less than 10 minutes and could
be used for 1 to 3 months QA. A modified (with an increased FOV), stand-alone
3D GRE scan, takes 3-minutes, could be used either with a spherical or
cylindrical phantom and is proposed for daily QA, in addition to the manufacturer
EPI-based stability scan. Fully automated reconstruction programs provided at
github.com/FastMRSI.
Introduction
Quality Assurance (QA) is
an important part of the MRI studies, ensuring data reliability. Daily phantom
scans are part of the manufacturer recommended QA (typically a 5-10 mins EPI
stability scan) and weekly, monthly or trimestral phantom scans are usually used
by single- or multi-center studies, for control and calibration. Given the sensitivity
of spectroscopic imaging acquisitions to B0 and B1 field inhomogeneities,
gradient readout imperfections, we propose using a gradient-readout, fast MRSI
and a 3D multi-echo GRE acquisition for QA. The metabolite concentrations in a
GE manufactured BRAINO phantom, similar to what is typically found in human
brain, are 3 to 4 order of magnitude smaller compared to the water signal.
While the typical MRSI voxel size is 2 to 3 orders of magnitude bigger, the SNR
in a spectroscopy acquisition is still substantially lower than for a regular water
imaging acquisition. Thus, stable
measurements would indicate normal scanner and coil functioning, while deviations
could be detected sooner.Methods
Gradient readout, fast MRSI
acquisition:
Using a 3T Prisma scanner (Siemens, Erlagen,
Germany) BRAINO phantom in the centre of either the 32-channel head, 64-channel
head or 20-channel head and neck coil, a 30x30x12, 7-mm isotropic rosette
spectroscopic imaging (RSI) scan1 with WET water suppression, TR/TE=1000/19.2ms,
FOV=210x210x84 mm3, a 1:48 min:sec water reference (RSI no WET,
TR=370ms, FA=15deg), and a 1:28 min:sec 3D GRE scan with same FOV as RSI, 128x128x48 matrix,
FA=8, TE1,2,3=2.46/4.92/8.61ms were acquired.
Data is reconstructed using rsi_recon (Linux
or Windows) and spectroscopic images and spectra are saved to NIFTI and .mat files.
When LCModel2 is available, siarraysDATA file is processed with lcm_rsi
Linux program. Both programs are parallelized, to take advantage of multi-core
CPUs. For lcm_rsi, installing Linux “parallels” is recommended, to take
advantage of CPU’s hyperthreading, when available.
LCModel
output for all voxels is read and used to generate baseline-subtracted, 0 and 1st
order phased spectra on which SNR for NAA, Cre and Cho is calculated (peak
divided to noise standard deviation, calculated in metabolite-free region). Linewidth (LW), metabolite concentrations, CRLBs, LCM SNR, etc
are extracted and metabolite concentration and CRLB maps are generated for
later visual inspection, if needed. Comma
separated (CSV) and Excel files with metabolite concentrations for all voxels,
ratios to Cr and CRLBs are generated.
Nifti files with spectra and metabolite maps are written out for easy
inspection with freeview (part of Freesurfer) or fsleyes (part of FSL). Averages
for LW, SNR, etc calculated in central spheroids, around the centre of the acquisition
ROI are written to the summary text file. The 3D GRE scan, if acquired, is used
to calculate SNR in each coil, as described below for the stand-alone scan and results
reported in same summary text file.
Stand-alone multi-echo 3D GRE scan:
A 3-echo 3D GRE
scan as above, but with 128x128x128 matrix and FOV=256mm can be acquired for
daily QA. Raw/twix data is reconstructed with greSNR program (github.com/FastMRSI/greQA).
SNR is calculated in each channel for the central slice (signal in a square ROI
around maximum intensity voxel, noise as standard deviation in square ROIs in
the 4 image corners) and an SNR3d is also calculated (signal in a cube ROI
around maximum intensity voxel, noise as standard deviation in cube ROIs in the
8 corners). For sum-of-squares combined image signal for SNR is calculated in central
square, signal for SNR3d in central cube.
The choice of echo times allows for calculation of “water” and “fat”
images (even though no fat in phantom we used) as well as B0 maps. While we’ll
implement the 3-point Dixon3 for calculations in the future, the
water and fat images are currently calculated simply as I2+I3 and I2-I3 with I2 and I3 images corresponding to TE2
and TE3.Results
Data for one exam and
full reconstruction results here:
https://pitt.box.com/s/61v6829svyt4lnxapow0st69rtkg6go1
Examples for channel SNR and SNR3d calculations
in Figures 1 and 2.Discussion
We propose using a
gradient readout, fast MRSI acquisition for 1-3 months QA and a 3D multi-echo GRE
for daily QA. Programs to automatically reconstruct the data and generate simple
to read summary text files (in addition to individual coil channels evaluation:
SNR, image, etc) are provided (github.com/FastMRSI). In conjunction with a framework
such as Yarra (http://yarraframework.com),
highly sensitive scanner and coil QA is possible with a high degree of automation.Acknowledgements
No acknowledgement found.References
[1] Schirda et
al. (2018). Magnetic Resonance in Medicine, 79(5), 2470-2480.
[2] Provencher,
S.W. (1993). Magnetic Resonance in Medicine 30,
672-679.
[3] Jingfei Ma. (2008). Journal of Magnetic
Resonance Imaging, 28, 543-558.