Natalie Schön1, Frank Seifert1, Gregory J. Metzger2, Bernd Ittermann1, and Sebastian Schmitter1,2
1Physikalisch-Technische Bundesanstalt (PTB), Braunschweig and Berlin, Germany, Berlin, Germany, 2Center for Magnetic Resonance Research, University of Minnesota, Minneapolis, MN, United States
Synopsis
Respiratory motion is a fundamental challenge for thorax MR
imaging, particular at ultra-high fields. Recent in-vivo studies have shown that
the coil's time-dependent scattering matrix (S-Matrix) can be used for respiration
gating to minimize those artefacts.
Here, we present electromagnetic simulations of the
respiration-induced changes of the transmit RF coil’s S-matrix. Analysing the impact of breathing patterns (conventional,
chest, abdominal) illustrates the combined effects of chest/diaphragm motion. The
type of respiratory motion and RF coil geometry, position and element type are
shown to impact the S-Matrix with implications on coil performance, respiratory
triggering, power monitoring and RF pulse design.
INTRODUCTION
Respiratory motion is a well-known problem for thoracic imaging. At UHF
($$$B_0\geq7T$$$) this is even more pronounced as navigators are
more difficult to use because of limited $$$B_1^+$$$[1] and self-navigation at UHF
is still under investigation [2]. A recent 7T-study
investigated the use of an alternative respiration detection method, which determines
respiration-induced variations of the scattering matrix (S-Matrix) using a
parallel transmission system [3].
Despite successful respiration detection, the measured changes varied
between elements of the 8-channel coil and it is still unclear if and how far other
factors, such as the coil's position, coil design and respiration pattern have
an impact.
In this work, we demonstrate that an
electromagnetic (EM) simulation with a respiration body model is highly
suitable to investigate such respiration-induced S-Matrix changes in detail. We
examine the impact of breathing patterns (conventional, chest and abdominal
breathing), of coil-body setups, of coil elements (loop vs. dipoles) and location
of the individual elements on respiration-induced S-Matrix variations. Finally,
such EM simulation-based S-Matrix
changes are compared to those found previously
in-vivo.METHODS
Finite-difference
time-domain (FDTD) EM simulations were performed with Sim4Life 6.0 (ZMT) using a 7T 16-TX/RX-channel body array coil
(8 modules consisting of 1 dipole + 1 loop element each; cf.Fig.1.5)[4,5]. The coil was loaded by different
respiration-resolved body models that were generated for different respiratory
states using XCAT[6]. Conductivity and relative permittivity
were assigned according to [7] (Fig.1.4).
Three breathing
patterns varing by diaphragm and chest displacement were modelled: abdominal,
chest, and conventional (i.e. chest+abdominal) breathing (Tab. 1). Models in
five respiratory states were generated for each breathing pattern. The fifth
state refers to maximum inhalation which can exceed displacement by~50%
compared to deep breathing, according to own measurements. Two RF coil setups were
investigated for deep, conventional breathing: a spatially fixed coil setup
[8,9], generating varying minimum distances of
10-30mm between chest and modules and a moving coil setup where the
anterior modules move simultaneously with respiration ensuring a 10mm fixed
minimum coil-body distance. For abdominal breathing the chest does not move,
which motivates using only the fixed coil setup. For chest breathing we
chose the moving coil setup which better reflects the experimental situation.
For each of the 20
simulations (5 respiratory states; 4 combinations of breathing pattern and coil
setup) a complex S-Matrix was extracted (Fig.1.6). An elementwise co-simulation
performed in Python (3.7) compensates staircasing artefacts in
rotated elements, introduced by the meshing process (Fig.1.7).
The 16x16 S-Matrix elements
are defined by: $$S_{ij}(t)=\frac{V_{fwd,i}}{V_{ret,j}}$$ with $$$V_{fwd,i}$$$forward voltage at channel i, $$$V_{ret,j}$$$: returned voltage at channel j[3]. (Fig.1.8) is decomposed into a time-invariant
and a time-varying respiration-induced component[3]:
$$S_{ij}(t)=S_{ij}^{(0)}+\Delta S_{ij}^{(motion)}(t)$$RESULTS
The highest $$$|\Delta S_{ij}^{(motion)}|$$$ changes occur by Loop 8 for the fixed coil setup and conventional
breathing of $$$|\Delta S_{L8L8}^{(motion)}|=8.89\cdot10^{-2}$$$ (cf. Fig.2+Tab.1). An order of magnitude
smaller values were observed for the other three setups (Tab.1). Depending on the location of the module, different sensitivities to
variations in surface and internal geometries were observed. Generally, the
central posterior dipoles appear to be more sensitive than the posterior loops
($$$4^{th}$$$ and $$$5^{th}$$$ module, Fig.2). The opposite is observed for
the front elements for conventional breathing/fix setup (Figs.2&4). This
might be caused by higher spatial coverage of loops and a larger penetration
depth of the dipoles. Fig.3 exemplarily shows the SD of $$$\Delta S_{ij}^{(motion)}$$$ over all respiratory states for the moving coil with conventional breathing. Strongest variations are observed on the
diagonal matrix elements (i=j, Fig.3),. This is followed by $$$S_{i\neq j}$$$ when j is either the neighbouring
module of i or the loop/dipole element of the same TX-module. Similar
observations were made for the other patterns/setups.
The largest $$$\Delta S^{(motion)}$$$-variation is clearly seen for front
modules (Fig.4b) during conventional breathing/fixed coil (gray
curve). Posterior elements (Fig.4c) show no relevant difference in $$$\Delta S^{(motion)}$$$ between moving (black curve) and fix setup for conventional breathing. Both findings reflect,
as expected, the coil-body distance plays an essential role. For posterior
elements the diaphragm motion (blue curve) appears to be less relevant compared
to the chest motion (violet curve), this is not observed for front elements. Fig.4b
shows similar variations of $$$\Delta S^{(motion)}$$$ for conventional breathing/moving
coil, abdominal and chest breathing.DISCUSSION AND CONCLUSION
This work investigates
respiration-induced changes of the S-Matrix by using a dedicated EM simulation
with a respiration-resolved body model. Overall, the results confirm experimental
results by Hess et al.[3], where variations of up
to $$$10^{-3}$$$ were found. Here, however, values up to $$$9 \cdot 10^{-2}$$$ were obtained, which may be related to different coil-element sizes and
larger peak diaphragm/chest displacements ($$$d_{diaphragm}$$$=48mm/$$$d_{chest}$$$=14mm)
compared to experiments. Fixed coil setups seem preferable over moving setups for
motion detection based on S-matrix changes, however, recent work showed higher respiration-induced
variations of SAR for this setup[4].
In conclusion, the separate
investigation of diaphragm and chest motion indicates $$$\Delta S_{ii}^{(motion)}$$$ of posterior elements appears less
sensitive for diaphragm motion, but none
of the motions dominates
the impact on $$$\Delta S_{ii}^{(motion)}$$$ of front elements. In this case both motion types have rather similar relevance for $$$\Delta S_{ii}^{(motion)}$$$ and both affect coil performance,
pulse design and power monitoring. Acknowledgements
This work was supported by the German Research Foundation
(DFG), grant number SCHM 2677/2-1.References
[1] S. Schmitter, S. Schnell, K. Uğurbil, M.
Markl, and P. F. Van de Moortele, “Towards high-resolution 4D flow MRI in the
human aorta using kt-GRAPPA and B1+ shimming at 7T,” J. Magn. Reson. Imaging,
vol. 44, no. 2, pp. 486–499, 2016.
[2] S. Dietrich et al., “3D Free‐breathing
multichannel absolute Mapping in the human body at 7T,” Magn. Reson. Med.,
no. October, p. mrm.28602, Dec. 2020.
[3] A. T. Hess, E. M. Tunnicliffe, C. T.
Rodgers, and M. D. Robson, “Diaphragm position can be accurately estimated from
the scattering of a parallel transmit RF coil at 7 T,” Magn. Reson. Med.,
vol. 79, no. 4, pp. 2164–2169, Apr. 2018.
[4] N. Schön et al., “Impact of
respiration on B1+ field and SAR distribution at 7 T using a novel EM
simulation setup,” Proc. Intl. Soc. Mag. Reson. Med., no. 1120, 2020.
[5] M. A. Ertürk, A. J. E. Raaijmakers, G.
Adriany, K. Uğurbil, and G. J. Metzger, “A 16-channel combined loop-dipole
transceiver array for 7 Tesla body MRI,” Magn. Reson. Med., vol. 77, no.
2, pp. 884–894, 2017.
[6] W. P. Segars, G. Sturgeon, S. Mendonca, J.
Grimes, and B. M. W. Tsui, “4D XCAT phantom for multimodality imaging research,”
Med. Phys., vol. 37, pp. 4902–4915, 2010.
[7] Itis Foundation, “Dielectric Properties.”
[Online]. Available:
https://itis.swiss/virtual-population/tissue-properties/database/dielectric-properties/%0A.
[8] J. T. Vaughan et al., “Whole-body
imaging at 7T: Preliminary results,” Magn. Reson. Med., vol. 61, no. 1,
pp. 244–248, 2009.
[9] S. Orzada et al., “A 32-channel
parallel transmit system add-on for 7T MRI,” PLoS One, vol. 14, no. 9,
pp. 1–20, 2019.