Bobby Runderkamp1, Thomas Roos2, Wietske van der Zwaag2, Matthan Caan3, Gustav Strijkers3, and Aart Nederveen1
1Radiology and Nuclear Medicine, Amsterdam UMC, Amsterdam, Netherlands, 2Spinoza Center for Neuroimaging, Amsterdam, Netherlands, 3Department of Biomedical Engineering and Physics, Amsterdam UMC, Amsterdam, Netherlands
Synopsis
Abdominal
MRI at ultra-high field benefits from increased SNR but is challenged by high
B1+-inhomogeneity. We use eight-channel parallel transmission to homogenize B1+
signal in 7T liver MRI and compare CP-transmission against phase shimming and a
kt-points pulse. Increased signal homogeneity was obtained with the kt-points
pulse and phase shimming compared to CP-transmission for small volunteers. Combining
the kt-points pulse with a compressed sensing reconstruction facilitated single
breath-hold high-resolution liver MRI. With improved B1-mapping and kt-point
optimization, kt-points is expected to also improve homogeneity in larger
volunteers, and to outperform phase shimming in achieving homogeneous signal in
7T liver MRI.
Introduction
The signal
gain facilitated by 7T MRI may be exploited to improve abdominal image quality.
However at this high field strength, the B1+ wavelength becomes smaller than
typical abdominal dimensions, which introduces B1+ inhomogeneities that cannot
be corrected for with traditional circularly polarized (CP) transmission. The aim of this work was to use
multi-channel parallel transmission to address B1+ inhomogeneities in the liver
in 3D, comparing phase shimming with a kt-points pulse1. This was
combined with SENSE and Compressed Sensing2 (CS) to enable
high-resolution 7T liver MRI in a single breath-hold.Methods
Five
healthy volunteers with a wide range in abdominal size were scanned with a 7T
Philips Achieva scanner (Philips, Best, The Netherlands) using an 8-channel
Tx/Rx fractionated dipole antenna body array3 (MRCoils, Netherlands,
Figure 1b). For each volunteer, a 3D sub-Ernst angle GRE per transmit channel
(single-channel-on) and a B1-map (DREAM4, all-channels-on, CP-transmission)
were combined to calculate single-channel B1-maps. These B1-maps were
noise-clipped to remove erroneous B1 values in areas with too low signal. Together
with a 2nd order B0-shimmed 3D-GRE, these maps served as input for
phase shimming and kt-points pulse calculation. Shimming volumes-of-interest
(VOI) in the whole liver (volunteers 1,3,4 and 5) or part of the liver
(volunteer 2) were manually drawn.
For phase shimming, the 8 Tx-channel phase offsets were individually adapted to
minimize the RF standard-deviation/(mean)2 in the VOI, using the MRCodeTool
toolbox (MRCode, Netherlands). Kt-points pulses were calculated with five
subpulses, max. blip strength = 33 mT/m, max. blip slew rate = 166 mT/m/ms and a
maximum excitation k-space extent of 1/6 cm-1. Channel RF amplitudes
and phases and the 3D excitation k-space trajectory were calculated using an
interleaved greedy-local optimization algorithm5 in a home-built
MRCodeTool extension6 (Figure 1d).
To evaluate B1+-homogenization techniques, 3D-GRE scans were acquired using either
CP-transmission, phase shimming, or kt-points. The 3D kt-points pulses were
non-selective, requiring large FOVs to cover the entire transmission volume of
the antennae. All scans were acquired in a single breath-hold. Details on the
scan parameters are given in Figure 1a.
To increase in-plane resolution while
preserving scan-time, the kt-points acquisition was combined with SENSE and CS-acceleration
(x3.65, volunteer 1). For CS, a random undersampling pattern was employed (Figure
1c) using the in-house developed PROUD software patch7,8. Reception
coil sensitivity estimation from a fully sampled k-space center and image
reconstruction were performed using BART9 and MRecon (Gyrotools, Zurich,
Switzerland) in MATLAB. Regularization parameter optimization was performed
heuristically. To assess image quality, signal intensity distributions in the
3D VOI for the CP-transmission, phase-shimmed and kt-points scans were
evaluated.Results
Figures
2&3 respectively show two slices of the CP-transmission, phase-shimmed and
kt-points scans, and figure 4 shows the signal intensity distributions in the
VOIs for all volunteers. In the smallest volunteers, 1&2, phase shimming
and kt-points show considerable increase in signal and homogeneity compared to CP-transmission.
Kt-points shows slightly improved homogeneity over phase shimming in areas
without noise-clipping, although small areas with signal dropout persist in
noise-clipped B1 regions. A modest increase in signal is seen for volunteer
4&5 for phase shimming and volunteer 4 for kt-points. In volunteer 3,
kt-points improved homogeneity; however, in the largest volunteer, 5, kt-points
reduced image quality.
Figure 5 shows two slices comparing the original kt-points scan with
acceleration by SENSE and CS. The increased resolution improves depiction of
fine structures at the cost of aliasing artifacts with SENSE. No such artifacts
are present in CS-reconstructed images, at the expense of slightly decreased
image quality in the lateral liver segment.Discussion
The
results show that in volunteers with a small waist, kt-points and phase
shimming can homogenize a considerable part of the liver. In areas where the
B1-maps are measured correctly, kt-points performs superior over phase
shimming. We think that the remaining small areas with signal dropouts might be
improved by increasing the max. excitation k-space extent of the kt-points
pulse and increasing the number of sub-pulses.
Less improvement is seen deeper in the abdomen, where the B1-map has insufficient
SNR and large areas were removed from the data using noise-clipping. As a
result, for volunteers with a larger waist, the calculated excitation k-space
trajectory was one in which all sub-pulses were emitted in the vicinity of a
single k-space location. This results in a suboptimal pulse producing a
gradient-like signal drop-off in the left-right axis. Possible improvements for
this inadequate B1-map are increasing SNR by reducing resolution and increasing
NSA, optimizing the noise clip factor or using Fourier-PE DREAM10. We also intend to make the kt-points sub-pulses slab-selective, allowing a
smaller FOV to be scanned and thus a smaller voxel size. Finally,
measuring homogeneity using flip angle distributions might be preferable over
signal distributions.Conclusion
We
were able to implement kt-points and phase shimming in 7T liver MRI using
multi-channel parallel transmission for 3D B1+-homogenization. Increased signal
and homogeneity were obtained with kt-points and phase shimming compared to CP-transmission
for small volunteers. Combining kt-points with a compressed sensing reconstruction
facilitated single breath-hold high-resolution liver MRI. With improved B1-mapping
and algorithm optimization, kt-points is expected to outperform phase shimming
in all body sizes and achieve a highly homogeneous excitation of the liver at
7T.Acknowledgements
No acknowledgement found.References
- Cloos MA,
Boulant N, Luong M, Ferrand G, Giacomini E, Le Bihan D, Amadon A. kT-points:
short three-dimensional tailored RF pulses for flip-angle homogenization over
an extended volume. Magn Reson Med.
2012;67(1):72-80. doi: 10.1002/mrm.22978.
- Lustig M,
Donoho D, Pauly JM. Sparse MRI: The Application of Compressed Sensing for Rapid
MR Imaging. Magn Reson Med.
2007;58:1182-1195. doi: 10.1002/mrm.21391.
- Raaijmakers AJE, Italiaander M, Voogt IJ,
Luijten PR, Hoogduin JM, Klomp DWJ, van den Berg CAT. The Fractionated Dipole
Antenna: A New Antenna for Body Imaging at 7 Tesla. Magn Reson Med. 2016;75:1366-1374. doi: 10.1002/mrm.25596.
-
Nehrke K,
Versluis MJ, Webb A, Börnert P. Volumetric B1(+) mapping of the brain at 7T
using DREAM. Magn Reson Med.
2014;71(1):246-56. doi: 10.1002/mrm.24667.
-
Grissom WA,
Khalighi M, Sacolick LI, Rutt BK, Vogel MW. Small-tip-angle spokes pulse design
using interleaved greedy and local optimization methods. Magn Reson Med. 2012;68(5):1553-62. doi: 10.1002/mrm.24165.
- Roos T,
Knapen T, van der Zwaag W. Reducing B1+ inhomogeneities in the brain at 7T:
Pads or Parallel transmit? Proceedings of
the 11th Annual Meeting ISMRM Benelux Chapter, 2019; Leiden,
Netherlands
- Peper ES, Gottwald LM, Zhang Q,
Coolen BF, van Ooij P, Strijkers GJ, Nederveen AJ. 30 times accelerated 4D flow
MRI in the carotids using a Pseudo Spiral Cartesian acquisition and a Total
Variation constrained Compressed Sensing reconstruction. Proceedings of the 27rd Annual Meeting ISMRM, 2018;Paris, France.
-
Gottwald LM, Peper ES, Zhang Q,
Coolen BF, Strijkers GJ, Planken RN, Nederveen AJ, van Ooij P. Pseudo Spiral
Compressed Sensing for Aortic 4D flow MRI: a Comparison with k-t Principal
Component Analysis. Proceedings of the
27rd Annual Meeting ISMRM, 2018;Paris, France.
-
Uecker M. Berkeley Advanced Reconstruction Toolbox
Target Audience: 2015;163(1991):25176. doi:10.1002/mrm.25176.
-
Tse DHY,
Poole MS, Magill AW, Felder J, Brenner D, Jon Shah N. Encoding methods for
B1(+) mapping in parallel transmit systems at ultra high field. J Magn Reson. 2014;245:125-32. doi:
10.1016/j.jmr.2014.06.006.