Yaohui Wang1,2, Qiuliang Wang1,2, Yang Liu1, Jigang Zhao1, Hui Wang1, Junsheng Cheng1, and Feng Liu3
1Institute of Electrical Engineering, Chinese Academy of Sciences, Beijing, China, 2University of Chinese Academy of Sciences, Beijing, China, 3School of Information Technology and Electrical Engineering, The University of Queensland, Brisbane, Australia
Synopsis
A new route for
ultrahigh field MRI shimming was proposed which utilized hybrid superconducting
matrix coil and multi-order room-temperature shim coils. The proposed scheme
was aimed to create a stable and accurate shimming solution for high
homogeneous magnetic field magnet. A real 9.4T whole-body MRI magnetic field
profile was used to validate the shimming strategy and a remarkable effect was
achieved. Therefore, the new shimming method deserves to generalize to
ultrahigh field MRI system.
Introduction
The current
magnet shimming strategies in MRI can be mainly divided into two categories,
namely field-based method or harmonics-based method. Passive shimming is a
commonly-used field-based method, which utilizes several shim trays with
independent iron pockets [1, 2]. The quantity
of the iron piece in an individual pocket is an optimizable parameter to
elevate the field homogeneity. There are also some reports using coils or loop
matrix inserted in the MRI magnet bore, which increase the field homogeneity by
adjusting the current of each coil element [3, 4]. On the other
hand, room-temperature shim coils [5, 6] are typically
employed to implement the harmonics-based shimming. A single shim coil
generates a magnetic field profile corresponding to the harmonic decomposition,
and it is controlled independently to offset certain amount of harmonic
components of the overall magnetic field. Compared with the coil shimming
technique, the passive shimming can be economical. However, its thermal
instability issue caused by gradient coils is unpredictable, which is possible
to deteriorate the field quality when the scanner is under operation. A replacement
of shim tray by superconducting matrix shim coil residing in the magnet system
can seek a constant environment for the shim device. Meanwhile, the spare space
originally for shim tray can be used to add more shim coils and cooling pipes.
A more powerful shimming framework by combing a superconducting matrix coil and
multi-order room-temperature shim coils is proposed for ultrahigh field MRI in
this work.Method
The superconducting matrix coil is composed of several saddle coils with
a matrix arrangement 8-row × 8-column. The target function for the shimming
optimization is to minimize the magnetic field deviation over the imaging
region. For the safety of the superconducting matrix coil and the thermal load
control of current lead, the maximum current in a single matrix coil element is
constrained. Eq. (1) and (2) express the optimization function,$$\min:\sum_{s=1}^S(Bz_s+{\bf{AI}}-x)^2\qquad\qquad(1)$$$$-I_{max}<I_k<I_{max}\qquad\qquad(2)$$where Bzs
is the bare magnetic field of the magnet system, A is the transfer matrix of the superconducting matrix coil, I is the current vector, x is the mean value of the shimmed magnetic
field, Imax is the maximum
allowable current, which is constrained within 5 A and k is the port number of the matrix coil.
Since the superconducting matrix coil is located in
the magnet system, the gap of the primary and shielding coils in the gradient
assembly can be used to arrange more shim coils. Fig. 1 shows the coil pattern
of the superconducting matrix coil and a set of third and fourth order room-temperature
shim coils. The fifth and sixth order zonal shim coils are also included for a
full offset of the zonal harmonic components.
For
validation, the proposed novel hybrid strategy was tested for the shimming of magnetic field of a 9.4T whole-body
superconducting magnet fabricated by IEE CAS is used for the validation of [7].
The illustration
of the locations of the superconducting matrix coil and room-temperature shim
coils were shown in Fig. 2.Results
Fig. 3
compares the magnetic field distribution among bare field, shimming with
superconducting matrix coil and additional shimming with room-temperature shim
coils. Table I lists the field homogeneity evaluation over the imaging volume.
The
initial magnetic field homogeneity over the 450-mm imaging volume (sphere) is
97.77 ppm with peak-peak value and 29.48 ppm (RMS). With optimized currents in
the matrix coil element, the peak-peak homogeneity increased to 17.96 ppm and
RMS homogeneity increased to 5.86 ppm. Subsequently, room-temperature shim
coils is responsible to cancel the remaining harmonic components, reaching a
final homogeneity with peak-peak value only 2.03 ppm and RMS value only 0.36
ppm.Discussions
To achieve an adequate field-adjustment capability of the
superconducting matrix coil, the coil element was selected with a large arc
side and the matrix coil elements were divided into four layers (as shown in
Fig. 1) to avoid overlapping. In our next work, the layout of the matrix
elements will be optimized to achieve a better solution.Conclusion
In
this work, we proposed a new shimming technique that combines superconducting
matrix shim coil and multiple-order room-temperature shim coils. The proposed
technique was tested in the case of shimming a 9.4T whole-body MRI magnet, and it
showed that the hybrid shimming
technique was capable of handling larger inhomogeneous magnetic field profile
with higher-order spherical harmonics.Acknowledgements
This work is funded by National Key
Research and Development Project (No. 2020YFF01014702, No. 2019YFC0117604), Beijing
Science and Technology Plan (No. Z181100003818020) and International Partnership
Program of Chinese Academy of Sciences (No. 182111KYB20170067, No. 182111KYSB20170039).References
[1] A. Belov, V. Bushuev, M. Emelianov,
V. Eregin, Y. Severgin, S. Sytchevski, et
al., "Passive shimming of the superconducting magnet for MRI," IEEE Transactions on Applied
Superconductivity, vol. 5, pp. 679-681, 1995.
[2] X. Kong, M. Zhu, L. Xia, Q. Wang, Y.
Li, X. Zhu, et al., "Passive
shimming of a superconducting magnet using the L1-norm regularized least square
algorithm," Journal of Magnetic
Resonance, vol. 263, pp. 122-125, 2016.
[3] C. Juchem, T. W. Nixon, S. McIntyre,
V. O. Boer, D. L. Rothman, and R. A. de Graaf, "Dynamic multi-coil
shimming of the human brain at 7T," Journal
of Magnetic Resonance, vol. 212, pp. 280-288, 2011.
[4] S. Littin, F. Jia, K. J. Layton, S.
Kroboth, H. Yu, J. Hennig, et al.,
"Development and implementation of an 84-channel matrix gradient
coil," Magnetic Resonance in
Medicine, vol. 79, pp. 1181-1191, 2018.
[5] P. Hudson, S. D. Hudson, W. B.
Handler, and B. A. Chronik, "Finite-length shim coil design using a
fourier series minimum inductance and minimum power algorithm," Concepts in Magnetic Resonance Part B:
Magnetic Resonance Engineering, vol. 37B, pp. 245-253, 2010/10/01 2010.
[6] S. Clare, J. Evans, and P. Jezzard,
"Requirements for room temperature shimming of the human brain," Magnetic Resonance in Medicine, vol. 55,
pp. 210-214, 2006.
[7] Q. Wang, Y. Dai, B. Zhao, S. Song, C.
Wang, L. Li, et al., "A
Superconducting Magnet System for Whole-Body Metabolism Imaging," IEEE Transactions on Applied
Superconductivity, vol. 22, pp. 4400905-4400905, 2012.