Bruno Pinho-Meneses1 and Alexis Amadon1
1Université Paris-Saclay, CEA, CNRS, BAOBAB, NeuroSpin, Gif-sur-Yvette, France
Synopsis
Different B0 shim system designs are
tested on two distinct fieldmap databases for assessment of commonly employed
metrics for evaluation of B0 inhomogeneity mitigation. We verify
that those metrics might not always be the best choice for performance analysis,
as, for a same shim system, they may provide hardly comparable values across
sites. We then propose a novel, simple and robust metric for shim system
analysis, the Spherical Harmonic Rating, which is shown to provide consistent
performance across sites, providing an equivalency of the shim system to a Spherical
Harmonics basis.
Introduction
As MRI scanners
move to Ultra-High Field (UHF), B0 inhomogeneity in the human brain increases
proportionally, notably exacerbating geometric distortion and signal loss in Echo Planar
Imaging (EPI). To reduce such artifacts, a significant amount of work has been
done by several research groups to design B0 shim systems capable of pushing
the boundaries of inhomogeneity mitigation1-19 relatively to
conventional 2nd/3rd order Spherical Harmonics (SH)
shimming, integrated in commercial scanners.
The widespread
metric for evaluating B0 inhomogeneity is the standard deviation of the
magnetic field excursion ($$$\sigma_{\delta{}B_0}$$$) across voxels in the region of interest.
However, significant variability of this metric across different sites even at
the same B0 field strength has been observed7,12, making it
insufficient for assessment and comparison of performance of different shim
systems across research sites. Examples of baseline inhomogeneity reported by
different sites at different field strengths are shown in Figure 1, with cases
of strong variability even at a same site at the same field strength. Noise,
different brain mask topology and fieldmap estimation method, for instance,
have been hypothesized as potential sources of noticed variability12,
but the question remains open.
To
overcome this difficulty, we propose a simple and robust metric denominated
Spherical Harmonic Rating (SHR), which is based on a quantitative comparison of
the proposed shim system to SH-based systems.Methods
For
a given fieldmap, if the resulting inhomogeneity $$$\sigma_\text{shim}$$$ after shimming by some arbitrary system is
between that obtained by unconstrained SH shimming of orders $$$n$$$ and $$$n+1$$$ ($$$\sigma_{\text{SH}_n}\geq{}\sigma_{\text{shim}}>\sigma_{\text{SH}_{n+1}}$$$), the
SHR of such system for this particular fieldmap is given by:
$$SHR=n+\frac{\sigma_{\text{SH}_n}-\sigma_{\text{shim}}}{\sigma_{\text{SH}_n}-\sigma_{\text{SH}_{n+1}}}$$
To verify consistency of this metric across
different sites, an in-house database15 composed of 100 fieldmaps,
acquired at 3T and re-scaled to 7T, and an open-access database20
with 126 fieldmaps, at 7T, are used in shimming simulations, first with
two different matrix Multi-Coil Arrays (M-MCAs) (cf. Fig. 2a), of 24 and 48
channels. The M-MCAs are composed of regularly distributed circular loops
around a cylindrical surface of 300-mm length and 281-mm diameter. The uniform
coverage of the cylindrical surface by these generic designs should provide an
unbiased system, delivering similar performance across distinct databases.
In addition, two 2-layer 36-channel optimized
MCAs (Fig. 2b and c) designed using a SVD-based methodology10,15 are simulated and
evaluated using the SHR. A low-bias and a high-bias SVD-based MCA system
(S-MCA) with respect to the in-house database were designed based on 50 and 100
fieldmaps, respectively. Besides using only 50 subjects, relaxed field-error
tolerance was allowed in the low-bias S-MCA design.
All shim systems
are simulated over both databases considering coils with 20 wire turns and
maximum 3A per channel.
Average baseline
inhomogeneity across fieldmaps of the in-house and open-access database are 65.7-Hz and 67.2-Hz, respectively.
The SHR is
compared to the inhomogeneity metric and to relative improvement over baseline
inhomogeneity.Results and Discussion
From results in Fig. 3b and c, assessment of the
MCAs’ robustness across sites from inhomogeneity and inhomogeneity reduction
metrics is not evident. Considerable metric differences across databases for
each MCA are observed. An exact same M-MCA would appear to present over 10%
absolute difference in inhomogeneity reduction when used on different sites at
the same magnetic field and electric current constraints, despite the different
databases presenting similar values of baseline inhomogeneity.
The differences considering the inhomogeneity
metric are also large, circa 7Hz. However, although the variability in this
metric makes it unsuitable for comparison of performances, it is the standard
metric for inhomogeneity assessment and it presents valuable information regarding
the magnetic field in the Region-of-Interest.
The average SHR for the different systems is
shown in Figure 3a. From this metric, M-MCAs seem to provide practically the
same performances across different databases, which is expected for these
generic designs. This evidences the invariant characteristic of the proposed
metric. The 24 and 48-channel M-MCAs present a SHR difference between databases
of 0.01 and 0.09, respectively. For S-MCAs, we see that the low-bias system
behaves similarly to a 48-channel M-MCA in terms of robustness to inter-subject
variability, with difference in SHR across databases of 0.09. For the high-bias
S-MCA, the difference is 0.33, clearly indicating its reduced robustness, as
the difference is 3.7 times greater than for the low-bias design. Therefore,
the SHR easily highlights drop in performance of the high-bias S-MCA on the OA
database relatively to the in-house database. Conclusion
The SHR provides
numerical values consistent with the expected behavior of the different shim
systems.
Certainly,
although providing consistent values across different databases for a same shim
system, the SHR alone does not provide all the information needed for assessing
the expected image quality from some acquisition, and observing the field
excursion values across different slices and the global inhomogeneity is
necessary.
Nevertheless,
to aid clinicians and researchers to pick their shim system from the many
designs available in the literature, this metric could be helpful and we
encourage its use; or that comparisons between the inhomogeneity obtained by the
shim system design and what would be achieved with unconstrained SH shim
systems are provided. Therefore, using a common basis for analysis, consistent
across sites.Acknowledgements
No acknowledgement found.References
1. C. Juchem et al. Dynamic multi-coil shimming of the human brain at 7T. Journal of Magnetic Resonance 2011;
212(2):280–288.
2. J.W. Pan et al. Role of
Very High Order and Degree B0 Shimming for Spectroscopy Imaging of
the Human Brain at 7 Tesla. Magn. Reson. Med. 2012; 68:1007-1017.
3. H. Han et al. Integrated parallel reception, excitation, and shimming
(iPRES). Magn. Reson. Med. 2013; 70(1):241–247.
4. C. Juchem et al. Dynamic
multi-coil technique (DYNAMITE) shimming for echo-planar imaging of the human
brain at 7 Tesla. NeuroImage 2015;
105:462–472.
5. J. P. Stockmann et al. A 32-channel combined RF and
B0 shim array for 3T brain imaging. Magn.
Reson. Med. 2016; 75(1):441–451.
6. T. Kim et al. Gradient-Echo
EPI Using a High-Degree Shim Insert Coil at 7 T: Implications for BOLD fMRI.
Magn. Reson. Med. 2017; 78:1734-1745.
7. J.P. Stockmann, and L.L.
Wald. In vivo B0field shimming methods for MRI at 7 T. NeuroImage 2018.
168:71-87.
8. A. Aghaeifar et al. Dynamic B0 shimming of the
human brain at 9.4 T with a 16-channel multi-coil shim setup. Magn. Reson. Med. 2018;
80(4):1714–1725.
9. B. Pinho-Meneses and A.
Amadon. A novel few-channel coil design for human brain shimming based on
stream function Singular Value Decomposition. Proc. ISMRM 2019; p.1483.
10. B. Pinho-Meneses, M. Luong,
and A. Amadon. Optimized multi-coil array design for human brain shimming at
Ultra-High Field. Proc. ISMRM 2019; p.1477.
11. B. Pinho-Meneses and A.
Amadon. Static-magnetic-field shimming coil system for magnetic resonance
imaging. European Patent 072081EP/EPR 2019.
12. J. Zhou et al. An
orthogonal shim coil for 3T brain imaging. Magn. Reson. Med. 2019;
83:1499-1511.
13. M.C. Can et al. Designs of
shim coils with distributed currents for 3T human brain magnetic resonance
imaging. Proc. ISMRM 2019; p. 1471.
14. B. Pinho-Meneses, and A. Amadon. A
fieldmap-driven few-channel shim coil design for MRI of the human brain. Phys.
Med. Bio. 2020.
15. B. Pinho-Meneses et al. First
prototype of a Stream-Function-based Multi-Coil Array dedicated to human brain
shimming at Ultra-High-Field. Proc. ISMRM 2020; p.766.
16. A. Aghaeifar et al. A
32-channel multi-coil setup optimized for human brain shimming at 9.4T. Magn.
Reson. Med. 2020; 83:749-764.
17. H.P. Hetherington et al.
Dynamic B0 shimming for multiband imaging using high order spherical harmonic
shims. Magn. Reson. Med. 2020; 85:531-543.
18. C. Oh, E. Choi, W. Hung,
and J. Lee. Machine designing: a multi-coil local shim design using Bayesian optimization.
Proc. ISMRM 2020; p. 4227.
19. N. Arango et al.
Computational Shim Coil Design Applied to Augmenting a Shim Array with a
Non-Eye-Occluding Face Plate. Proc. ISMRM 2020; p.4224.
20. H.
Elshatlaway et al. An open dataset of 126 human brain field maps. Proc. ISMRM
2020; p. 4219.