Justin Yu1, Anshuman Panda1, and Alvin Silva1
1Department of Radiology, Mayo Clinic Arizona, Phoenix, AZ, United States
Synopsis
Quantitative MRI can precisely measure biomarkers for
disease, such as liver hemochromatosis. R2* correlates with liver iron
concentration (LIC), but data is scarce for 3T R2* mapping. The out-of-the box
vendor provided sequences for quantitative R2* measurements do not precisely
measure large R2* values in a standardized phantom. The discrepancy is greater
at higher R2* values. This can lead to underestimates in LIC quantification for
patients with severe hemochromatosis. R2* mapping from different MRI vendors
may yield different results, and patient specific QA may be necessary for clinical
utilization for quantitative R2* mapping.
Introduction
Iron overload
has primarily been diagnosed by liver biopsy in past clinical practice1.
In recent years, MRI has increasingly been utilized to quantify and map
biomarkers for disease, including FDA-approved commercial products that can
provide LIC measurements with MRI2.
One such biomarker is the relaxation rate R2*; mapping R2* in the liver
can detect high liver iron concentration (LIC) without the need for an invasive
biopsy. Many of the previous studies on LIC quantification using R2* mapping were
performed at 1.5T, with threshold values for disease being extrapolated to
higher field strengths by simply scaling the values by a constant. R2* mapping
sequences are now available clinically on 3T scanners. We seek to characterize
the accuracy of these sequences at 3T using a standardized R2* phantom. In
addition, a strategy for calibrating individual patient exams is explored by scanning the phantom and a human volunteer simultaneously. Methods
A Calimetrix (Madison,
WI, USA) R2* phantom was scanned using two 3T MRI scanners from two vendors (referred
as ‘vendor A’ and ‘vendor B’) with the whole-body transmit and multi-channel body
receive coil. Out-of-box multi-echo DIXON sequences were used for data
acquisition. The phantom manufacturer reported R2* values ranging from 16.6 to
991.4 s-1 using 12 vial inserts. The phantom was scanned three
times; once with the phantom alone in the scanner, and once with a volunteer on
the table at the same time, with two different volunteers. The volunteers were in
the right lateral recumbent position, with the phantom placed anterior to the
volunteer’s chest, at the same level as the liver. When scanning, the phantom
vials and the volunteer right liver lobe were in the same field of view. The Vendor
A mapping sequence had TR 9.23ms, TE at 0.9, 1.55, 2.2, 2.85, 3.5, 4.15, 4.8,
5.45, 6.1ms, slice thickness 3.5mm, FA of 4 deg. The Vendor B mapping sequence
had TR 9.26ms, TE at 1.26, 2.60, 3.94, 5.28, 6.62, 7.96ms, slice thickness
3.5mm, FA of 4 deg. All slices were
acquired axially, and the scans for each vendor were performed back to back. 16mm
diameter ROIs were drawn in Visage PACS and the means recorded.Results
R2* values are summarized in Figure 2. R2* mapping from vendor A and vendor B do not perform identically when measuring the phantom alone, or the phantom with a volunteer present. The data obtained from volunteer 1 are different from that of volunteer 2. In the 12 measured vials, higher discrepancies between the measured value and the phantom manufacturer reported values are found in the high R2* range vials (vials 10-12) than the low R2* vials.Discussion
The data shows that vendor A and
B had mildly variable performance for measuring R2*. Vendor A slightly overestimates R2* for vials 1-11, and Vendor B slightly underestimates
R2* for vials 9-12 as compared to scanning the phantom alone (red bar in Figure 2). This suggests that imaging centers need to
characterize the R2* mapping performance of their equipment on a scanner by
scanner basis. Furthermore, both vendors’ data shows more inaccuracy for vials in the high
range of R2* values (~400 s-1 and up)3. Thus, additional
optimization is required for the pulse sequences and/or post processing
algorithms for scanning patients with suspected higher iron concentration.
Introducing
a human body to the phantom scan seems to generate some variation in the final
measured R2* values. For vendor A, the measured values of the phantom plus the
volunteer are slightly lower than the R2* values measured with the phantom alone.
For vendor B, the results are more variable, with the measurements being
higher with volunteer 1, and the measurements being roughly the same for
volunteer 2. This suggests that in order to correlate R2* measurements with
liver iron concentration, patient specific calibration may yield better results than a
universal calibration curve for that scanner.
Scanning the
phantom alongside the patient’s liver is a potential method for patient
specific calibration, but this setup may not be well tolerated for some
patients, and it deviates from the norm of supine positioning for abdominal
scans. Additional scans with more volunteers may elucidate the effects of
scanning the phantom alongside a person.Conclusion
This preliminary
study shows that currently existing 3T R2* mapping sequences perform well for
low to moderate R2* ranges but is less accurate at high R2* ranges
corresponding to severe liver iron concentration. Further refinement to these sequences and
their post processing algorithms may improve their accuracy in this region. The
variation in measured R2* values using different scanning positions also
warrants further investigation, in order to create workflows that maximize
mapping accuracy while also preserving patient comfort and keeping exam setup
simple.Acknowledgements
No acknowledgement found.References
1.
Hernando D, Levin YS, Sirlin CB, Reeder SB.
Quantification of liver iron with MRI: State of the art and remaining
challenges. Journal of Magnetic Resonance Imaging. 2014;40(5):1003-1021.
doi:10.1002/jmri.24584
2.
Chan WC, Tejani Z, Budhani F, Massey C, Haider
MA. R2* as a surrogate measure of ferriscan iron quantification in thalassemia.
Journal of Magnetic Resonance Imaging. 2013;39(4):1007-1011.
doi:10.1002/jmri.24216
3.
Labranche R, Gilbert G, Cerny M, et al. Liver
Iron Quantification with MR Imaging: A Primer for Radiologists. RadioGraphics.
2018;38(2):392-412. doi:10.1148/rg.2018170079