Patrick Metze1, Tobias Speidel1, Fabian Straubmüller1, and Volker Rasche1,2
1Internal Medicine II, Ulm University Medical Center, Ulm, Germany, 2Core Facility Small Animal Imaging (CF-SANI), Ulm University, Ulm, Germany
Synopsis
In
this work we evaluate the proposed nonuniform self-gating (nuSG) method for
lung imaging in highly irregular respiratory patterns and compare its
performance to image-based and k-space-based gating methods and to a breath-hold
reference. Functional parameters (proton fraction, fractional ventilation) and
image sharpness are evaluated, with the nuSG algorithm showing a superior
performance in case of fractional ventilation and image sharpness. In more
uniform motion patterns its performance is similar to image-based SG and both
techniques outperform k-space based gating.
Introduction/Purpose:
MRI
lung imaging has emerged as a promising alternative to computed tomography due
to its capabilities in the visualization and assessment of functional and
morphological information, without the use of ionizing radiation1,2. However,
short T2* relaxation times and low proton densities in the lung lead to low SNR,
with scan durations being limited by maximum breath-hold durations. Dedicated
sequences such as ultrashort echo time (UTE3) and zero echo time (ZTE4) better
exploit the available signal and have furthermore been optimized for lung
imaging5-9. While breath-hold imaging is possible for healthy and
cooperative subjects10,11,12, patients with lung diseases often exhibit
limited breath-hold capabilities, leading to the need of free-breathing
acquisitions in combination with motion compensation strategies. Traditional
self-gating approaches rely on the identification of a characteristic frequency
and thus perform best for uniform motion, but often irregular breathing
patterns due to anxiety, shortness of breath or coughing are observed.
Therefore, we adapted the nonuniform self-gating (nuSG) framework13,14,15 for
lung imaging and evaluated the method in six healthy volunteers, comparing the
results to established k-space16 and image-based9,17 gating techniques.Methods:
All imaging experiments were performed on a
clinical 1.5 T whole-body MRI system (Achieva 1.5 T, Philips Healthcare, Best,
The Netherlands) using a dedicated 32-channel cardiac coil. A 2D UTE sequence (parameters
are given in Table 1) was combined with a tiny golden angle (tyGA) acquisition
scheme18,19 to minimize motion sensitivity, eddy current related artefacts
and to provide a better angular coverage for sliding-window and self-gating
reconstructions. At two coronal slice locations two breath-hold acquisitions
(inspiration and expiration) and two free-breathing acquisitions with uniform
and irregular respiratory patterns were performed. The free-breathing acquisitions were
reconstructed using nuSG13,14,15, image-based self-gating9 and a k-space center based
self-gating16 (Figure 1). To obtain quantitative information on lung
density, proton
fraction values (PF) in inspiration and expiration were
pixelwise calculated according to $$PF = \frac{SI_{lung}}{SI_{muscle}}\textrm{exp}\left(\frac{TE}{T^*_2}\right),$$ with
SI being the signal intensity of the lung parenchyma and intercostal muscle, TE
the echo time and $$$T^*_2$$$ = 2.11 ms20. After nonrigid image registration of
exhaled and inhaled motion states using the Medical Image Registration Toolbox
for Matlab (MIRT)21, fractional ventilation maps (FV) were calculated
according to$$FV = \frac{SI_{EX}-SI_{IN}}{SI_{EX}}$$
To evaluate the residual motion blur after gating, image sharpness was
analysed based on a standard metric (e.g. 13, 22) that identifies the position
of 25% and 75% of the maximum signal intensity on an interpolated intensity
profile placed over the lung liver interface. For image sharpness analyses, the
lung-liver interface is divided into three areas that are automatically
detected by slope and variance analyses. The areas are described by linear
equations and the x-distance of the intersections is chosen as a measure for
sharpness.Results:
Figure 2 shows reconstructed coronal slices of
all three gating approaches. For inspiration, image sharpness of the nuSG approach
is clearly superior to img-based and k-space-based self-gating. Further nuSG
allows the reconstruction of the breathing motion over the entire acquisition
period (Figure 3).
All quantitative measurements are summarized in
Figure 4. Both sharpness measures show the superiority of nuSG and img-based
self-gating. Sharpness for breath-hold images is higher in expiration compared
to inspiration. nuSG sharpness is on a similar level across both motion states
and both respiratory patterns, with proton fractions being overestimated.
Differences between inspiration and expiration are noticeable for breath-holds,
image-based gating and to a lesser extent in nuSG and k-space based SG.
Concerning fractional ventilation, nuSG comes closest to the BH reference, with
a slight underestimation.Discussion:
The presented results are in accordance with
previous studies (9, 17) and indicate that k-space based self-gating is clearly
outperformed by image-based and non-uniform self-gating. Especially for irregular
respiration with varying ex- and inspiratory positions, the lung liver
interface appears blurred in k-space based-SG. The reduced inspiratory image sharpness
for the breathhold approach is most likely caused by respiratory drifts. For
uniform breathing patterns the image-based approach performed excellent but partly
fails for irregular breathing. Non-uniform self-gating works equally well for
data acquired during uniform and irregular breathing, thus exhibiting better sharpness
measures for inspiration even compared to the BH reference.
The higher proton fraction values in the nuSG
reconstructions could be due to the reconstruction with less data. The
resulting lower SNR increases the influence of noise and could lead to an
increased lung signal in magnitude images. However, there seems to be no
negative effect on FV.
As the difference between inspiration and
expiration is typically higher in BH imaging compared to free breathing, it is obvious
that the calculated FV is higher compared to gated imaging. While nuSG and
image-based SG yield similar FV, ksp-based SG yields reduced values, most likely
due to residual image blur.Conclusion:
The nuSG
approach in combination with 2D-tyUTE achieves functional values in line with
the image-based SG approach as well as with the reference BH technique, while
offering the advantage of intrinsically reconstructing a movie of the
underlying time course of the motion, yielding a possible time-resolved evaluation
of lung parenchyma changes even for highly non-uniform motion.Acknowledgements
The authors thank the Ulm University Centre for Translational Imaging MoMAN for its support.References
1.
Hirsch
et al., Pediatr Radiol 2020
2. Serra et al., Chest 2011, 140 (6)
3. Holmes et al., Radiography 2005, 11
(3)
4.
Weiger et al., Magn Reson Med
2013, 70 (2)
5.
Bae
et al., Eur Radiol 2019, 29 (5)
6.
Bae
et al., Eur Radiol 2020
7.
Burris et
al., Radiol 2016, 278 (1)
8.
Zucker et al., J Magn Reson Imaging
2018, 47 (1)
9.
Tibiletti
et al., Magn Reson Med 2016, 75 (3)
10.
Wild et
al., Insights into Imaging 2012, 3 (4)
11.
Biederer
et al., Insights into Imaging 2012, 3 (4)
12.
Lederlin
et al., J Magn Reson Imaging 2014, 40 (4)
13.
Wundrak
et al., Magn Reson Med 2016, 76 (3)
14.
Metze
et al., Proceed ISMRM 2020 #0438
15.
Speidel
et al., Proceed ISMRM 2020 #1087
16.
Weick et al., J Magn Reson Imaging
2013, 37 (3)
17.
Balasch et al., J Magn Reson Imaging
2020
18.
Wundrak
et al., IEEE T Med Imaging, 2014, 34 (6)
19.
Wundrak
et al., Magn Reson Med, 2016, 75 (6)
20.
Yu et al.,
Magn Reson Med, 2011, 66 (1)
21.
Myronenko, Mirt, 2010
22.
Larson et
al. Magn Reson Med, 2005, 53 (1)