Anne Slawig1, Andreas Max Weng1, Bernhard Petritsch1, Simon Veldhoen1, and Herbert Köstler1
1Department of Diagnostic and Interventional Radiology, University Hospital Würzburg, Würzburg, Germany
Synopsis
MR imaging of the lung is a challenge due to
the low proton density in lung parenchyma and the very short relaxation times.
To overcome low signal and SNR problems UTE-sequences have successfully been
applied to lung imaging. Here, the benefit of a 2D UTE-acquisition for the
determination of ventilation and perfusion, as determined by SENCEFUL-MRI, is
evaluated. It is shown that SENCEFUL-MRI clearly benefits from the use of a
UTE-sequence.
Introduction
MR imaging of the lung is very challenging but of
high interest for functional analysis, especially in pediatric patients or to
monitor chronic lung diseases.
Recently,
MR-sequences using ultra-short echo times (UTE-MRI) have been proposed for functional
lung imaging (1–3). One approach to determine functional lung
parameters is the SENCEFUL algorithm (4), which has already proven itself in multiple clinical
applications (1,5,6) and has been combined with 3D UTE to evaluate
ventilation values.
Here,
the benefit of a 2D-UTE-acquisition for the determination of both, ventilation
and perfusion, by SENCEFUL-MRI is evaluated.Materials and Methods
All experiments were performed at a 3T-MRI
(Siemens MAGNETOM Prisma). Coronal 2D slices in the lung were acquired in 3
healthy volunteers. One Cartesian scan using a quasi-random line ordering was
acquired at minimum possible echo time (TE=0.69ms) for comparison. Radial
acquisitions with a center-out read-out scheme were acquired at TE= 0.69 ms,
0.40 ms and 0.34 ms. Other imaging parameters were kept constant: TR=2.0ms,
flip angle=8°, resolution=3.0x3.0mm2, slice thickness=10mm, measurement
time=2min each. In the Cartesian measurement an additional short acquisition
window is needed after each read-out to acquire the DC-signal, whereas in the
radial scan the first points of each read-out serve as DC-signal.
All
data was retrospectively gated into 30 respiratory phases and 21 cardiac
phases, using the DC-signal. A segmentation of the lung was performed by a
region-growing algorithm and functional parameters were determined using the
standard SENCEFUL algorithm (4). Ventilation values were
normalized to end-expiration signal and overall lung volume change during one
breathing cycle. Perfusion values were normalized to a pure blood ROI in the
descending aorta. For quality control, the distribution of the ventilation
values and the perfusion phase values inside the lung ROI was analyzed.Results
Figure 1, 2 and 3 show results of an exemplary
volunteer. Anatomical images in Figure 1 show the expected increase in signal
intensity in the lung parenchyma with shorter TE. Also, the overall impression
of the images changes as signal intensities approach a pure proton-density
weighted contrast and artefacts due to chemical shift disappear. Anatomical images
at the same TE, but different read-outs, look similar.
The
ventilation maps (Figure 2a) appear smoother with decreasing TE. The
distribution of ventilation values in the radial acquisitions becomes narrower with
shorter TE as shown by the decreasing interquartile range (IQR) (Figure2 b).
IQR in a radial acquisition is slightly higher than in the Cartesian
acquisition at the same TE.
Perfusion
maps are comparable through all measurements and echo times (Figure 3a). Large
vessels in distal areas are clearly visible, while perfusion values decrease in
the proximal regions.
Perfusion
phase, as deviation from the phase in the pure-blood reference ROI (Figure 3b),
appears smooth, large deviations only occur at the edges and close to the
heart.
The
distribution of phase values is small (Figure 3c) and slightly decreases with
lower echo times in the radial acquisition.
All
findings were reproducible in all 3 volunteers, the IQR of ventilation and
perfusion phase for all measurements are shown in Table 1.Discussion
Anatomical images show the expected effects of
a decreasing echo time. The quantification of ventilation benefits from shorter
echo times. In a healthy volunteer, the ventilation is expected to be uniform
in a coronal slice, which translates into a small IQR. As the determination is
based on a change in signal intensity in the lung parenchyma, the increased
signal intensity at short TE render the determination more robust, as supported
by the obtained smaller IQR of the ventilation value distribution.
Similar
results, but weaker effect size, were found for the perfusion maps. This could
be caused by the comparatively long T2* relaxation times of blood.
For
the Cartesian acquisition a minimum echo time of 0.69ms was given by the RF-pulse,
slice selection and read-out dephasing times. To achieve shorter TE a radial
acquisition was implemented. Unfortunately, these suffer from a non-perfect
distribution of sampling points in k-space, leading to lower SNR than in a
comparable Cartesian acquisition. Therefore, an initial reduction in the
quality of functional parameter maps at similar TE occurs. Nevertheless, the
loss can be more than compensated by the shorter TE possible in radial imaging.
Additional
advantages of radial acquisitions are general robustness towards motion and the
implicit acquisition of the DC-signal in each readout. In this study, TR and
acquisition time was kept constant for better comparability. However, the
included DC signal and the shorter TE of radial acquisitions also would allow
for shorter overall measurement times.
In
contrast to the more common 3D approach for UTE, the 2D acquisition allows the determination
of perfusion related parameters. But the necessity of slice selection limits
minimum TE, such that a 2D sequence cannot achieve the same shortness of TE as
a 3D approach. Additionally, through-plane motion (especially of the heart and
large vessels) can cause difficulties in image registration and the
determination of functional values.Conclusion
It is shown here that SENCEFUL-MRI clearly benefits from the use of a
UTE-sequence. As signal in the lung decays rapidly, shorter echo times can
provide increased signal intensity. The improved signal-to-noise consequently
allows for a more robust determination of functional parameters.Acknowledgements
No acknowledgement found.References
1. Pereira LM, Wech T, Weng AM, et al. UTE-SENCEFUL: first results for 3D high-resolution lung ventilation
imaging. Magnetic Resonance in Medicine 2019;81:2464–2473
2.
Balasch A, Metze P, Stumpf K, et al. 2D Ultrashort Echo-Time Functional Lung
Imaging. Journal of Magnetic Resonance Imaging 2020;52:1637–1644 doi:
https://doi.org/10.1002/jmri.27269.
3. Heidenreich JF, Weng
AM, Metz C, et al. Three-dimensional
Ultrashort Echo Time MRI for Functional Lung Imaging in Cystic Fibrosis.
Radiology 2020;296:191–199
4.
Fischer A, Weick S, Ritter CO, et al. SElf-gated Non-Contrast-Enhanced
FUnctional Lung imaging (SENCEFUL) using a quasi-random fast low-angle shot
(FLASH) sequence and proton MRI. NMR in Biomedicine 2014;27:907–917
5.
Kunz A, Weng A, Wirth C, et al. Funktionelle native Lungen-MRT
(SENCEFUL) zur Ermittlung pulmonaler Perfusionsdefizite bei Cystischer Fibrose.
In: RöFo - Fortschritte auf dem Gebiet der Röntgenstrahlen und der bildgebenden
Verfahren. Vol. 189. Georg Thieme
Verlag KG; 2017. p. WISS 314.8.
6.
Veldhoen S, Weng AM, Knapp J, et al. Self-gated Non–Contrast-enhanced
Functional Lung MR Imaging for Quantitative Ventilation Assessment in Patients
with Cystic Fibrosis. Radiology 2016;283:242–251