Efe Ilicak1, Jascha Zapp1, Safa Ozdemir1, Lothar R. Schad1, and Frank G. Zöllner1,2
1Computer Assisted Clinical Medicine, Medical Faculty Mannheim, Heidelberg University, Mannheim, Germany, 2Mannheim Institute for Intelligent Systems in Medicine, Medical Faculty Mannheim, Heidelberg University, Mannheim, Germany
Synopsis
Functional
lung imaging is of great importance for diagnosis of pulmonary diseases.
Previously, a non-contrast-enhanced method called Fourier Decomposition was
proposed for assessing pulmonary functions based on balanced steady-stated free
precession pulse sequence. However, this pulse sequence is known to be
sensitive to magnetic field inhomogeneities. Here, we propose a phase-cycled
acquisition for improved robustness against field inhomogeneities. In vivo
results from 1.5 T and 3 T scanners are provided to demonstrate the performance
of phase-cycled acquisitions for functional lung imaging. Preliminary results
indicate that phase cycling can be a viable option for reducing limitations
arising from magnetic field inhomogeneities.
Introduction
Functional
imaging is of great importance for diagnosis and monitoring of prevalent lung
diseases. To this end, a non-contrast-enhanced method called Fourier
Decomposition (FD) was previously demonstrated to achieve local ventilation and
perfusion information during free breathing.1 In FD MRI, a series of images are
acquired during free breathing using balanced steady-state free precession (bSSFP)
sequence to capture signal variations stemming from respiratory and cardiac
signal modulations.
Although FD
MRI has shown clinical promise, the bSSFP sequence is sensitive to magnetic
field inhomogeneities, resulting in irrecoverable signal loss known as banding
artefacts.2 Despite significant improvements, occurrence
of banding artefacts reduces the robustness of the method, especially hindering
its usefulness at a commonly used clinical field strength of 3 T.3,4 In this work, we incorporate RF
phase cycling into bSSFP acquisitions to improve FD MRI’s robustness against field
inhomogeneities. We present in vivo results to demonstrate the performance of
the proposed method. Methods
In FD MRI, a
2D bSSFP sequence with a constant RF phase commonly acquires K images, usually with 3-4 frames/sec.1 Afterwards, this time-series is
registered,5 and analysed using temporal Fourier
transform to obtain regional density changes corresponding to respiratory and
cardiac frequencies. Consequently, ventilation- and perfusion-weighted maps are
generated.6
In
multi-acquisition bSSFP imaging, K
images are acquired with N different
RF phase increments $$$\Delta\phi_n$$$ with $$$n \in [1 \ N]$$$, to change the spatial location of the banding artefacts.7,8 Here, we have used N=4 different phase cycles with a block-wise phase-cycling scheme,9 meaning that K/4 images were acquired for each phase cycle.
For evaluation, in
vivo constant phase and phase-cycled acquisitions were obtained for two
volunteers using a 1.5 T scanner (Magnetom Aera, Siemens Healthineers, Erlangen, Germany) with TR/TE = 1.88/0.80 ms, TA = 174
sec; and using a 3 T scanner (Magnetom Skyra, Siemens Healthineers, Erlangen, Germany) with TR/TE =
2.31/1.02 ms, TA = 188 sec. The rest of the parameters were kept identical
between the scanners and were as following: FOV = 450 mm x 450 mm, slice
thickness = 15 mm, GRAPPA factor = 3, flip angle = 50$$$^\circ$$$, bandwidth = 1302 Hz/Px.
For each acquisition, 280 images were acquired using standard shim with a 0.2 s
pause between measurements. For the FD analyses, these acquisitions were divided into 4 subgroups (SG) and
initial 10 images were excluded to eliminate transient state behavior. Each of
these subgroups was then registered individually and the FD was performed to obtain ventilation and
perfusion maps of each subgroup. Combinations of these subgroups were also
generated.10
To assess the
image quality, parenchymal signal-to-noise (SNR) was calculated on registered
image series with respect to voxels containing air; and the contrast of
ventilation (CNRV) and perfusion (CNRQ) were calculated with respect to their surrounding
tissue. Results
Tables 1 and
2 show the SNR and CNR averaged across the subjects for 1.5 T and 3 T
acquisitions. At 1.5 T, $$$\Delta\phi = 0$$$ generates the
lowest SNR and CNR scores, meanwhile, $$$\Delta\phi = \pi/2$$$ and $$$\Delta\phi = 3\pi/2$$$ perform
similarly in terms of perfusion contrast. At 3 T, the difference of SNR and CNR
between the phase cycles diminishes, due to the increased field inhomogeneities.
Figure 1 shows
the functional maps of each subgroup at 1.5 T for constant phase (a) and
phase-cycled (b) acquisitions. As expected, the constant phase acquisition
provides similar contrast throughout the experiment whereas phase-cycled
acquisition is able to generate functional maps
different spatial features.
Here, $$$\Delta\phi = \pi$$$ provides more
comprehensive perfusion maps in terms of CNR, compared to other phase cycles. Figure 2
shows the functional maps of each subgroup at 3 T. Here, conventional $$$\Delta\phi = \pi$$$ acquisition suffers
from increased field inhomogeneities, especially visible in the perfusion
maps. Whereas, with RF phase cycling, the proposed method improves robustness
against field inhomogeneities and is able to obtain more comprehensive perfusion
maps.
Figure 3 shows
the combined maps overlaid on top of a single cross-section for 1.5 T (a) and 3
T (b) acquisitions. At 1.5 T, phase-cycled acquisition suffers from lower
signal values in the perfusion maps due to averaging effects. Nonetheless, phase-cycled maps provide similar
contrast and prominent structures compared to their constant phase version. At 3 T, banding artefacts are more
prominent due to increased field inhomogeneity as expected, and the constant
phase acquisition is
less able to reproduce structures and a comprehensive perfusion map. On the other hand, the phase-cycled
acquisitions are able to improve the robustness against field inhomogeneity and
reproduce prominent structures and a comprehensive perfusion map. Discussion & Conclusion
We have
developed and demonstrated phase-cycled bSSFP acquisitions for functional lung
imaging. Individual and combined ventilation- and perfusion-weighted maps were
successfully displayed at 1.5 T and 3 T field strengths. While further studies
with more subjects are warranted, our preliminary results indicate that phase
cycling can be incorporated to
FD MRI to improve robustness against magnetic field inhomogeneities, at the expense of reduced scan
efficiency. Overall, phase cycling can be a viable option for reducing limitations
arising from magnetic field inhomogeneities and therefore, increase the
clinical value of FD MRI. Acknowledgements
This work
was supported by Deutsche Forschungsgemeinschaft (grant number: DFG 397806429).References
- Bauman G, Puderbach M, Deimling M, et
al. Non-contrast-enhanced perfusion and ventilation
assessment of the human lung by means of Fourier decomposition in proton MRI. Magn
Reson Med. 2009;62(3):656-664. doi:10.1002/mrm.22031
- Bangerter
NK, Hargreaves BA, Vasanawala SS, et al. Analysis of multiple-acquisition SSFP.
Magn Reson Med. 2004;51(5):1038-1047.
-
Bauman
G, Pusterla O, Bieri O. Ultra-fast Steady-State Free Precession Pulse Sequence
for Fourier Decomposition Pulmonary MRI. Magn Reson Med.
2016;75(4):1647-1653. doi:10.1002/mrm.25697
-
Bauman
G, Pusterla O, Bieri O. Functional lung imaging with transient spoiled gradient
echo. Magn Reson Med. October 2018. doi:10.1002/mrm.27535
- Chefd’hotel
C, Hermosillo G, Faugeras O. Flows of diffeomorphisms for multimodal image
registration. In: Proceedings IEEE International Symposium on Biomedical
Imaging. IEEE; :753-756. doi:10.1109/ISBI.2002.1029367
- Ilicak
E, Chacon-Caldera J, Zapp J, Schad LR, Zoellner FG. Compressed Sensing
reconstruction in non-contrast-enhanced functional lung MRI using Fourier
Decomposition: An initial study. Proc Intl Soc Mag Reson Med.
2019;27:1897.
-
Çukur
T. Accelerated Phase-Cycled SSFP Imaging With Compressed Sensing. IEEE Trans
Med Imaging. 2015;34(1):107-115. doi:10.1109/TMI.2014.2346814
-
Lauzon
ML, Frayne R. Analytical characterization of RF phase-cycled balanced
steady-state free precession. Concepts Magn Reson. 2009;34A(3):133-143.
-
Benkert
T, Ehses P, Blaimer M, Jakob PM, Breuer FA. Dynamically phase-cycled radial
balanced SSFP imaging for efficient banding removal. Magn Reson Med.
2015;73(1):182-194.
-
Brennan
DG. Linear Diversity Combining Techniques. Proc IRE.
1959;47(6):1075-1102. doi:10.1109/JRPROC.1959.287136