Brandon Zanette1, Yonni Friedlander1,2, Marcus J Couch3,4,5, and Giles Santyr1,2
1Translational Medicine, The Hospital for Sick Children, Toronto, ON, Canada, 2Department of Medical Biophysics, University of Toronto, Toronto, ON, Canada, 3Siemens Healthcare Limited, Montreal, QC, Canada, 4McConnell Brain Imaging Centre, Montreal Neurological Institute, Montreal, QC, Canada, 5Department of Neurology and Neurosurgery, McGill University, Montreal, QC, Canada
Synopsis
Inert fluorinated (19F) gas MRI
is an emergent technology that has potential to be a
lower cost, reduced infrastructure alternative to hyperpolarized gas MRI for lung
imaging. In this work we present
an approach for non-Cartesian spiral imaging in a phantom. Compared to typical
gradient echo imaging with the same spatial resolution, the proposed approach increased
SNR by 44% in approximately a third of the acquisition time. Further simulation
and extrapolation to clinical imaging experiments indicate significant potential
for this approach to improve SNR and image quality.
Introduction
Inert
fluorinated (19F) gas MRI is an emergent technology that has potential
to be a lower cost, reduced infrastructure alternative to hyperpolarized (HP)
gas MRI for lung imaging1, providing similar and complementary
information2,3. These gases, such as sulfur
hexafluoride (SF6) or perfluoropropane (PFP; C3F8),
are biologically inert, low cost, and have favorable properties for MRI. Most
importantly, they do not require hyperpolarization, instead relying on a
relatively high number of 19F nuclei per molecule and signal
averaging; enabling imaging under thermal equilibrium conditions1.
19F gas MRI is typically performed
with gradient echo (GRE) or steady-state free precession (SSFP) imaging using
cartesian approaches, resulting in modest resolutions and SNR due to the
inherently low signal intensities and short T2*. Non-Cartesian sequences
(e.g. spiral k-space trajectories) are rapid, and more efficiently cover
k-space possibly allowing for more signal averages during a typical participant
breath-hold (≤20 sec), while also potentially permitting reduced echo times2,4.
In this preliminary work, we theoretically
and experimentally compare 2D-GRE with rapid, non-Cartesian 19F gas
imaging using a 3D stack-of-spirals (3D-SoS) pulse sequence5. Methods
Imaging
was performed on a clinical 3T scanner (MAGNETOM Prisma, Siemens, Erlangen,
Germany). A 1 L bag was filled with SF6 (Messer Canada, Mississauga,
Canada) and imaged using a 19F birdcage transmitter and 8-channel
receive array (Rapid Biomedical, Rimpar, Germany). Table 1 shows the
acquisition parameters. The spatial resolution for both sequences was 6.25×6.25×25
mm3. Transmitter voltage was adjusted to yield a 90⁰ flip angle and both
acquisitions were averaged 10 times for sufficient signal intensity. Images
were reconstructed in MATLAB (MathWorks, Natick, MA). 3D-SoS images were
reconstructed using a non-uniform FFT6. A 9×9 pixel ROI was
used to estimate SNR (Fig. 1). A simulation comparing the achievable scan
durations of both sequences for several in-plane spatial resolutions, assuming
10 slices, and keeping all other parameters in Table 1 constant, was performed.
A minimum TR 3D-SoS case for each resolution was also explored. This was
extended to estimate the number of signal averages (NSA) achievable in a 20 sec
breath-hold. Finally, the relative SNR changes (normalized to the 64x64 2D-GRE
case) accounting for differences in TE, in-plane resolutions, and number of
achievable signal averages was simulated under assumed Ernst angle imaging
conditions for PFP (T1=12 ms and T2*=2 ms)7.Results
From
Table 1 it is observed that with these parameters 3D-SoS is 2.56 times faster
per slice than 2D-GRE. Fig. 1 shows images acquired with both sequences. 3D-SoS
images are observed to have image quality comparable to 2D-GRE. However, SNR
was observed to be increased using 3D-SoS compared to 2D-GRE with values of 19.5
and 13.6, respectively. Extrapolating these results to clinical imaging (Fig.
2) with simulation demonstrates the advantage of fast, spiral imaging to reduce
scan duration and increase the NSA in a breath-hold, especially if short TR can
be maintained. Discussion
In
this work a 3D-SoS acquisition is applied to 19F gas imaging in a
phantom. 3D-SoS demonstrated a 44% improvement in SNR versus 2D-GRE for the
same spatial resolution and number of signal averages, in nearly one third of
the time. This is presumably due to the decreased minimum TE of the 3D-SoS
sequence resultant from the spiral-out trajectory. Further reductions in TE
should be achievable with shorter RF pulse durations, variable TE in kz,
or fully non-selective acquisitions. Stack-of-stars UTE has demonstrated improved
SNR with 19F gas MRI4. However, in that
study, long acquisition windows and in-plane radial undersampling in order to
comply with breath-hold constraints contributed to significant image blur, negating
the advantages of reduced TE. Here, fully-sampled data with comparatively few
acquisitions and short readouts is possible due to the efficiency of the spiral
k-space trajectory, mitigating image corruption while reducing TE.
Although
this work is very preliminary and limited to phantoms, there is significant promise
for clinical use. Fig. 2a illustrates simulated achievable scan durations for
several in-plane resolutions. These illustrate the significant reduction in
scan duration and potential for increased NSA with 3D-SoS (Fig. 2b) leading to
increased SNR (Fig. 2c). Furthermore, it is observed that SNR improvements may
be traded for improved resolution. Alternatively, reductions in scan time may
be leveraged to perform rapid, repeated imaging during extended multi-breath or
free-breathing experiments to explore gas kinetics, since 19F gases
may be freely administered in normoxic mixtures8–10. However, these simulations,
especially the minimum TR case, should be interpreted with caution, requiring experimental
corroboration and may represent an idealized scenario. Human loading conditions
may limit achievable RF power and/or such short TR may exceed SAR thresholds.
Nevertheless, there is likely a “practical middle-ground” that may be
empirically determined.
Lastly,
this work uses SF6 which may be less desirable than PFP due to the
increased RF deposition requirements to maintain Ernst angle imaging associated
with a short T1 of ~1 ms. PFP has a more favorable T1 (12
ms) reducing RF transmit requirements while remaining short enough to maintain
rapid, repeated imaging and will be used in future clinical studies. Conclusion
3D-SoS
allows for improved SNR compared to 2D-GRE under similar imaging conditions.
Extrapolating to clinical imaging indicates the potential of this acquisition
to improve 19F gas MRI in humans. Acknowledgements
Funding
was provided by the Natural Sciences and Engineering Research Council (NSERC)
of Canada. References
1. Couch
MJ, Ball IK, Li T, Fox MS, Biman B, Albert MS. 19 F MRI of the Lungs Using
Inert Fluorinated Gases: Challenges and New Developments. J Magn Reson
Imaging. 2019;49(2):343-354. doi:10.1002/jmri.26292
2. Obert AJ, Gutberlet M, Kern AL, et al.
1H-guided reconstruction of 19F gas MRI in COPD patients. Magn Reson Med.
2020;84(3):1336-1346. doi:10.1002/mrm.28209
3. McCallister A, Chung SH, Antonacci M, et
al. Comparison of single breath hyperpolarized 129 Xe MRI with dynamic 19 F MRI
in cystic fibrosis lung disease. Magn Reson Med. 2020;0(0):1-11.
doi:10.1002/mrm.28457
4. Couch MJ, Ball IK, Li T, et al. Pulmonary
ultrashort echo time 19F MR imaging with inhaled fluorinated gas mixtures in
healthy volunteers: Feasibility. Radiology. 2013;269(3):903-909. doi:10.1148/radiol.13130609
5. Zanette B, Friedlander Y, Munidasa S,
Santyr G. Comparison of 3D Stack-of-Spirals and 2D Gradient Echo for
Ventilation Mapping using Hyperpolarized 129Xe. Proc Intl Soc Mag Reson Med.
2020;28:0449.
6. Fessler JA, Member S, Sutton BP.
Nonuniform Fast Fourier Transforms Using Min-Max Interpolation. IEEE Trans
Signal Process. 2003;51(2):560-574.
7. Couch MJ, Ball IK, Li T, et al. Inert
fluorinated gas MRI: a new pulmonary imaging modality. NMR Biomed.
2014;27(12):1525-1534. doi:10.1002/nbm.3165
8. Gutberlet M, Kaireit TF, Voskrebenzev A,
et al. Free-breathing dynamic 19F gas MR imaging for mapping of regional lung
ventilation in patients with COPD. Radiology. 2018;286(3):1040-1051.
doi:10.1148/radiol.2017170591
9. Gutberlet M, Kaireit TF, Voskrebenzev A,
et al. Repeatability of Regional Lung Ventilation Quantification Using
Fluorinated (19F) Gas Magnetic Resonance Imaging. Acad Radiol.
2019;26(3):395-403. doi:10.1016/j.acra.2018.10.021
10. Goralski JL, SH C, Glass T, et al. Dynamic
perfluorinated gas MRI reveals abnormal ventilation despite normal FEV1 in
cystic fibrosis. JCI Insight. 2019:E-Pub Ahead of Print.