Amir Masoud Afsahi1, Zhao Wei1, Michael Carl2, Saeed Jerban1, Hyungseok Jang1, Nicole Le1, Jiang Du1, Eric Y. Chang1,3, and Ya-Jun Ma1
1Department of Radiology, University of California, San Diego, San Diego, CA, United States, 2GE HealthCare, San Diego, CA, United States, 3Radiology Service, Veterans Affairs, San Diego Healthcare System, San Diego, CA, United States
Synopsis
To explore the
application of a 3D slab-selective UTE sequence as a superior technique to 3D
ZTE at 3T field strength and to compare with 3D CT as a gold standard in Lumbar
spinal bone assessment.
Introduction
Computed
tomography (CT) offers both high-resolution and high-contrast imaging of bone, and
is recognized as the gold standard imaging modality for spinal bone evaluation.
However, due to the ionizing radiation exposure associated with CT, this
imaging modality is not recommended for children or for patients requiring
frequent examinations. In comparison, magnetic resonance imaging (MRI) is a superior
method given that it does not involve radiation exposure and that it is able to
provide high-contrast visualization of soft tissue. However, bone signals
cannot be effectively detected by conventional MRI sequences due to their low
proton density and short transverse relaxation times of bone (T2~390
μs at 3T). Fortunately, ultrashort echo time (UTE) and
zero echo time (ZTE) sequences with echo times (TEs) less than 100 μs are
able to capture these bone signals1,2. Most recently, ZTE sequences combined with data
post-processing have been applied to skull, shoulder, cervical spine, and hip imaging,
and have demonstrated similar contrast to CT3-5. However, ZTE application in the lumbar spine may
encounter some technical challenges, including artifacts caused by respiratory
motion resulting from the use of non-selective rectangular radiofrequency (RF)
pulses, and spatial signal inhomogeneity caused by the potential coil ring-down
factor. In contrast, UTE takes advantage of slab-selectivity using a soft pulse
for signal excitation that can reduce artifacts related to respiratory motion. Further,
because UTE allows for adjustment of echo time, the tissue signal inhomogeneity
induced by the coil ring-down factor can be reduced. In this study, we
investigated the performance of 3D UTE and ZTE sequences in lumbar spinal bone
imaging and compared UTE imaging with the gold standard CT imaging.Methods and Materials
Figure
1 shows the features of 3D ZTE and UTE sequences. The ZTE sequence utilizes a non-selective
rectangular RF pulse with short duration for excitation (8 μs),
followed by 3D center-out radial sampling (Fig. 1A). To minimize echo time, readout
gradients are turned on prior to signal excitation so that the gradient
encoding begins simultaneously with signal excitation. 3D UTE sequence enables
slab selection by using a soft-half pulse for excitation (~1 ms) together with
a slice-selective gradient (Fig. 1B). After excitation, the spatial-encoding
gradient is turned on and simultaneous data acquisition begins. In data
post-processing, the N4ITK bias correction algorithm was performed for both UTE
and ZTE images. Then, the corrected images were logarithmically transformed and
inverted to generate high bone contrast.
In
this study, we first to compare 3D UTE and ZTE sequences in lumbar spinal bone
imaging, a 29-year-old healthy female volunteer was scanned. All scans were
performed in coronal plane to exclude soft tissue in the abdomen (i.e., the major
tissues to cause respiratory motion artifacts). The detailed sequence
parameters of UTE and ZTE are found in Table 1. Second, to investigate the best
resolution for UTE spinal bone imaging, five scans with different isotropic resolutions
(i.e., 0.9, 1.0, 1.2, 1.6, and 2 mm3) were performed on the
above-described healthy volunteer. A 72-year-old male patient
with spinal bone fractures was also recruited and underwent UTE bone imaging for
comparison against his most recent CT images.Results and Discussion
Figure
2 presents the comparative ZTE and UTE lumbar spinal bone images of the healthy
volunteer. As expected, UTE imaging showed much better bone structural
information compared with ZTE. Figure 3 presents the five UTE scans with
different resolutions for this same volunteer. The signal-to-noise ratio (SNR) was
lower when the resolution was higher. However, when the resolution was too low,
the bone structures were not very well-demonstrated. The resolution with a voxel
size of 1.2 mm3 showed a good balance between image SNR and bone structures.
Figure 4 compares UTE and CT spine bone images of the patient with spinal bone
fractures. Vertebral fracture was clearly visible in both CT and UTE images. There
were also high levels of agreement between UTE MRI and gold standard CT scan. Conclusion
The
3D UTE sequence was superior to the 3D ZTE sequence in lumbar spinal bone
imaging because of its reduced sensitivity to respiratory motion. Furthermore,
the UTE imaging of spine offered comparable results to gold standard CT. Therefore,
3D UTE MRI is suitable as a clinically-compatible technique in spinal bone
imaging and has strong potential for effective incorporation into the clinical
workflow.Acknowledgements
The authors
acknowledge grant support from the NIH (R01NS092650,
and R21AR075851),
Veterans Affairs (I01RX002604 and I01CX001388), and GE Healthcare.References
1. Du J, Carl M, Bydder M, Takahashi A,
Chung CB, Bydder GM. Qualitative and quantitative ultrashort echo time (UTE)
imaging of cortical bone. J Magn Reson. 2010 Dec;207(2):304–11.
2. Delso G, Wiesinger F, Sacolick LI,
Kaushik SS, Shanbhag DD, Martin H. Clinical Evaluation of Zero-Echo-Time MR
Imaging for the Segmentation of the Skull. 2014;417–23.
3. Breighner RE, Bogner EA, Lee SC, Koff
MF, Potter HG. Evaluation of Osseous Morphology of the Hip Using Zero Echo Time
Magnetic Resonance Imaging. Am J Sports Med [Internet]. 2019 Oct 21;47(14):3460–8.
Available from: https://doi.org/10.1177/0363546519878170
4. Wiesinger F, Sacolick LI, Menini A,
Kaushik SS, Ahn S, Veit-Haibach P, et al. Zero TE MR bone imaging in the head.
Magn Reson Med. 2016 Jan;75(1):107–14.
5. Weiger M, Brunner DO,
Dietrich BE, Pruessmann KP, Colin FM. ZTE Imaging in Humans. 2013;332:328–32.