Zhenhua Shen1, Xuchen Zhu1, Shihong Han1, Fuyi Fang1, Wei Luo1, Shao Che1, Zidong Wei1, Jinguang Zong1, Yongquan Ye2, Bo Li1, Shuheng Zhang1, Anthony Vu2, Weiguo Zhang2, and Guobin Li1
1United Imaging Healthcare, Shanghai, China, 2UIH America, Inc., Houston, TX, United States
Synopsis
For
the first time, routine clinical body imaging with large FOV on a whole-body 5T
MRI system is demonstrated. With multi-channel RF parallel transmission
hardware architecture and static RF shimming techniques, the uniformity of the RF
transmission field is shown to be well controlled for imaging quality guarantee.
Preliminary results show great promise for body imaging at 5T.
Introduction
Whole
body MR imaging at ultra-high field remains challenging due to transmit B1(B1+)
field inhomogeneity, as radiofrequency (RF) wavelength in human tissues is much
shorter than the dimensions of human torso1 and causing constructive/destructive
interference patterns in resultant images. In order to mitigate the issue, RF
shimming methods including static parallel transmission (static pTx, or static
RF shimming) and dynamic parallel transmission (dynamic pTx, or dynamic RF
shimming) have been proposed on the basis of multi-channel transmission hardware
architecture.2 Here we present the initial
results of abdomen and pelvis imaging at 5T with static RF shimming, and
demonstrate the capability of the 5T MR system for routine body imaging.Methods
Experiments
were performed on a whole body 5T MRI scanner (United Imaging
Healthcare, Shanghai, China) as shown in Figure 1. The parallel RF transmission
system consists of eight independent RF power amplifiers (RFPAs), each with a
peak power of 8kW. The volume transmission coil used was an eight-channel loop
array coil. For the initial testing, only the amplitude and phase of each
channel were changed to optimize the target B1+ fields which is commonly used
in 3T system as static RF shimming method. The static RF shimming procedure starts
with a prescan calibration step with a B1+ mapping sequence. The total acquisition time for all eight
channels in prescan takes less than 1 minute. Desired weights of each transmit
channel are then calculated using magnitude least squares optimization and used
for the later scan.3 The T2-weighted (T2W) fast spin echo (FSE)
sequence with fat saturation (FOV: 380mm*300mm, Matrix: 336*225, TR: 4000ms,
TE: 87.2ms, Slice Thickness: 6mm, Slice Number: 24) was selected as the demonstrating
protocol for abdominal imaging (Figure 2). Another routine T2W FSE sequence (FOV:
350mm*350mm, Matrix: 416*312, TR: 4800ms, TE: 75.04ms, Slice Thickness: 3mm,
Slice Number: 26) was chosen for image quality assessment of pelvic imaging
(Figure 3).Results
Figure 2 and Figure 3 demonstrate
the B1+ profiles of both circular polarization (CP) mode and optimized transmitting mode used after
static RF shimming process in abdomen and pelvis. The coefficient of variance
(CV) reflects the degree of B1 homogeneity, and greater CV values correspond to
greater inhomogeneity. Before static RF shimming, severe shading artifacts can
be seen in T2W FSE images. After static RF shimming, B1+ uniformity within the
abdomen and pelvis was improved to yield smaller CV values, and the resultant
image became more uniform. The CV value decreased by nearly 16% compared with
that of CP mode in abdomen, and by 33% in pelvis.Conclusion
Preliminary
results show great promise for routine whole body MR clinical imaging at 5T, an
area where 7T has been struggling due to more severe B1+ inhomogeneity and SAR
constraints. The static RF shimming procedure takes less than a minute and can
be implemented as part of the prescan procedures. Further improvements, such as
by dynamic RF shimming, will be investigated in future study.Acknowledgements
This work is supported by National Key R&D Program of China N0.2017YFC0108800.References
- Erturk, M. A.; Li, X.;
Van de Moortele, P. F.; Ugurbil,
K.; Metzger, G. J., Evolution of UHF Body Imaging in the Human Torso at 7T:
Technology, Applications, and Future Directions. Top Magn Reson Imaging 2019,
28 (3), 101-124.
- Padormo, F.; Beqiri, A.; Hajnal, J. V.; Malik, S. J., Parallel
transmission for ultrahigh-field imaging. NMR
Biomed 2016, 29 (9), 1145-61.
- Setsompop, K.; Wald, L.
L.; Alagappan, V.; Gagoski, B. A.; Adalsteinsson, E., Magnitude
least squares optimization for parallel radio frequency excitation design
demonstrated at 7 Tesla with eight channels. Magn Reson Med 2008, 59 (4), 908-15.