Matthew A. McCready1, William B. Handler1, Francisco Martinez2,3, Timothy J. Scholl2,3, and Blaine A. Chronik1,2
1Physics and Astronomy, Western University, London, ON, Canada, 2Medical Biophysics, Western University, London, ON, Canada, 3Robarts Research Institute, Western University, London, ON, Canada
Synopsis
Delta relaxation enhanced magnetic resonance (dreMR) is a field-shifting quantitative molecular imaging method. The dreMR method may be used to produce images with signal proportional to concentration of contrast agents with longitudinal relaxivity dispersion. Here we find that field inhomogeneities and ramping periods, which were previously ignored, cause significant errors in dreMR images. These errors include apparent non-zero signal from dispersion-free tissue, and apparent change in signal due to contrast agents. We show that these effects can be mitigated by use of an improved homogeneity design method, and increased slew rates.
INTRODUCTION
Current medical imaging trends show a growing need for quantitative
molecular imaging in preclinical studies. Thus far, positron emission
tomography (PET) has been the prevailing method but has an associated radiation
dose resulting in undesirable effects in longitudinal studies.1 Delta
relaxation enhanced magnetic resonance (dreMR or “dreamer”) is a
contrast-enhanced MRI method for quantitative molecular imaging. The dreMR technique
uses a B0 insert coil to shift the strength of the static
field in a pulse preparation phase of a pulse sequence. Using contrast agents
with longitudinal relaxivity (r1) dispersion, images taken at
different field strengths can be subtracted, resulting in signal proportional
to the concentration of these agents.2 This subtraction is weighted
by the different field strengths to cancel out dispersion-free signal. In
development of the dreMR method and design of previous dreMR coils,2,3
it was assumed that field homogeneity is unimportant for the dreMR system, and
effects of ramping periods were also ignored. Here, we find that field
inhomogeneities, and these ramping periods have significant effects on dreMR
images and that the improved homogeneity design method we recently developed4
can reduce these effects.METHODS
To see inhomogeneity effects analytically, we use the original dreMR
subtraction where the images now differ in field with their weightings by the
given inhomogeneity. We apply this method in MATLAB to various concentration
distributions of the contrast agent VivoTrax using its r1 dispersion data and simulated field maps given by dreMR
coil designs. Artificial noise consistent with real dreMR images is added to
these simulations. Ramping effects require a new set of Bloch equations, where
longitudinal field strength varies in time, and therefore so does r1 in the presence of
contrast agent. We consider only the longitudinal magnetization as dreMR is
unaffected by behavior of the transverse. This
is solved numerically in MATLAB using ode45 for the ramping periods, and solved
analytically using the assumptions that ramping field strength is linear in
time, r1 is linear with Bz, and overall
ramping time is small, so we ignore terms of O(Δt2) or higher.RESULTS
Dispersion-free signal in the presence of inhomogeneities is given
analytically through (1), where, ΔB* is the dreMR field with some inhomogeneity. A simulated dreMR subtraction with
an inhomogeneous coil, and one of our new coil designs can be seen in Figure 1.
Here, cylinders of various VivoTrax concentrations are included in the domain.
The same simulation for each coil is shown in Figure 2 where concentration of
VivoTrax is homogenous across the domain. The field contours for each coil are
given in Figure 3. Dispersion-free
signal accounting for a ramping period is given analytically through (2), where
ξ is the dreMR
coil slew rate, and Δt2 the flat top duration of the dreMR pulse. A
comparison of this model with ode45 calculations is given for various slew
rates in Figure 4.
$$(1) I_0~=2B_0 (\Delta B-\Delta B^*)/(B_0^2-\Delta B^2)$$
$$(2) I_0~=(2M_0B_0/T_1)(\Delta B^2/(B_0^2-\Delta B^2))(1/\xi)exp(-\Delta t_2/T_1)$$DISCUSSION
Inhomogeneities in the dreMR field shift were found to result in improper
cancelling of signal from dispersion-free pixels, and artefacts in signal from
contrast agents. These effects result from a difference between expected field
in the subtraction weighting, and the actual field at a given point. The
shading effects seen in Figure 1, are clearly caused by field inhomogeneities
as seen by comparison of Figures 2 and 3. The effect was seen to produce a
large difference in signal, reaching a percent difference of 14% in the example
of Figure 2. Using our new design method,4 we can improve
homogeneity and minimize these effects as seen in Figures 1 and 2. Ramping
effects were also found to result in improper cancelling of signal from
dispersion-free tissue, and a change in signal from contrast agents. These
effects can be minimized by increasing slew rate. While decreasing ΔB
can also minimize this effect, doing so decreases signal from contrast agents
and causes a loss of contrast.CONCLUSION
Field inhomogeneities can lead to shading artefacts in dreMR images, and
signal from dispersion-free regions. These effects can be minimized by using a
layer of specifically placed, field-correcting windings within a dreMR coil.
The ramping periods of dreMR pulses result in signal from dispersion-free
regions and a change in signal from contrast agents, but these effects can be
mitigated by using a high slew rate. By using our improved homogeneity design
method and a high slew rate, we can noticeably improve quality of images from
this quantitative molecular imaging method.Acknowledgements
The authors would like to acknowledge financial support from NSERC, the Ontario Research Fund, and the NSERC Canada Graduate Scholarship - Masters.References
1.
Hildebrandt I, et al. Anesthesia and Other Considerations for in Vivo Imaging
of Small Animals. ILAR Journal. 2008;49(1):17-26.
2.
Alford J, Rutt B, Scholl T, et al. Delta relaxation enhanced MR: Improving
activation-specificity of molecular probes through R1 dispersion imaging.
Magnetic Resonance in Medicine. 2009;61(4):796–802.
3.
Harris C, Handler W, Araya Y, et al. Development and optimization of hardware
for delta relaxation enhanced MRI. Magnetic Resonance in Medicine. 2013;72(4):1182–1190.
4. McCready M,
Handler W, Chronik B. Improving homogeneity in delta relaxation enhanced magnetic
resonance using boundary element method. ISMRM AGM. 2020;3312.