Matt A Bernstein1 and Seung-Kyun Lee2
1Radiology, Mayo Clinic, Rochester, MN, United States, 2Biomedical Engineering, Sungkyunkwan University, Suwon, Korea, Republic of
Synopsis
Applying systematic dimensional analysis based linear
algebraic methods, we derive the important scaling relation for MR gradient
design: gradient coil inductance L scales as the square gradient efficiency (measured
in T/m/A) and the fifth power of radius.
The inclusion of "turns" as a unit in the dimensional matrix produces the result uniquely, greatly
simplifying the analysis.
Introduction
An important result in gradient design is the inductance L
of gradient coil scales as the fifth power of the coil’s radius a. The proportionality relationship L α a5 holds provided the primary and shield layers are
sufficiently well-separated [1-2]. Coil inductance plays a central role in
gradient performance because the stored energy is given by Ws = ½LI2 where I
is the coil current, and gradient risetime τ=LI (V-IR) -1, where V is the applied voltage and R
is the coil resistance.
The scaling relationship L α a5
has
had a major impact on MR gradient performance as well as power and
chilled-water requirements as the standard patient aperture has increased from
60 to 70 cm, which approximately doubles inductance. Also, along with reduced
peripheral nerve stimulation [3,4], the scaling of L strongly motivated for advent of smaller bore
size scanners and gradient coil inserts, e.g. [5-7]. Here we examine the basis
of this important scaling relationship using systematic, dimensional analysis methods
that are based on linear algebra [8-9]. Methods
To apply systematic dimensional analysis we seek a set of
exponents (e1,e2
… en
) such that the product of a set
of n physical parameters (q1e1
× q2e2 ×...qnen) equals to
1, i.e., is dimensionless. We are free to choose the set of q based on experience
or modeling, as detailed in [8-9].
Consider the five physical quantities L (inductance in H), G (gradient amplitude in
T/m), a (radius
in m), μ0
(permeability
of free space in H/m), and I
(current in A). L is the quantity being investigated, and the
inclusion of the other four is motivated by their appearance as dimensioned quantities
in the expression for G in simple gradients for which an analytical result
can be written, e.g., a Maxwell coil pair (e.g., see Eq. 6 in [10]).
The next step in systematic dimensional analysis is to decompose
the units for each physical quantity into fundamental units [8], e.g., 1T/m = m-1
kg
s-2
A-1. The exponents e satisfy the
homogeneous linear equation De=0, where D is called the
dimensional matrix. Its n columns are labeled by the physical
parameters, and its rows are labeled by the units (Fig. 1). In addition to the
relevant basic SI units (m,kg,s and A), the fifth row in D represents
the unit turns. This inclusion is motivated by the product N×I in units of amp-turns
appearing commonly in transformer and coil design. Recall L scales as N2 and G scales linearly with N, allowing us to complete the bottom row of D (Fig. 1). For illustrative purposes, we also
analyze the problem omitting turns as a unit, i.e., constructing the 4x5
dimensional matrix D’ (Fig.2).
In general, the number of linearly independent, non-zero
solutions for e is given by n-rank(D),
and is equal to the number of dimensionless products that can be formed. The solutions for e were determined by
Gaussian elimination.Results
The matrix D has rank=4,
so there is only 5-4=1 distinct, non-zero solutions for e shown in Fig.
1. According to the Buckingham’s theorem [8], the resulting dimensionless quantity
equals some constant, which dimensional analysis does not specify, yielding
L½
I μ0½ (G a5/2) -1 = const...….[1]
Squaring the expression in
Eq.1, absorbing μ0
into the constant, and defining the gradient
efficiency η≡GI -1
yields
L α (G2 I -2) a5 = η2a5….....……[2]
The quadratic dependence of
inductance on gradient efficiency illustrates a fundamental tradeoff between
maximal gradient amplitude and slew rate [1-2].
If instead “turns” is excluded as a row in
the dimensional matrix D' (Fig. 2), then there are two distinct, non-zero
solution to D'e=0 because rank(D')= 3. When there are two dimensionless parameters, Buckingham’s theorem
implies they are related by an (unspecified) function [8]. This yields
a μ0/L = f[ LI (Ga3)-1 ]..............[3]
Typically the unspecified
function f needs to be determined with further measurements or simulations. The
scaling relation L α η2a5
can still be obtained from Eq. 3, but only with the
choice f(x) = x-2
, which is equivalent to choosing e4
= ½ e5 in Fig 2. Without the physical insight of the
dependence of L and G on turns, there is no a priori reason to
make these choices.Discussion and Conclusion
The analysis that includes the number of turns as unit is
considerably more useful because the scaling relationship L α η2a5
emerges as
the unique result. Unlike the analysis omitting turns and resulting in
Eq. 3, no further measurements or simulations are required to obtain the
unknown functional dependence in Eq.3.
While turns is a dimensionless unit (i.e., not a length,
mass, charge, etc.), this is not a concern. There are many other dimensionless
units in commonly use, e.g., moles of a substance, or radians to measure angle,
which is dimensionless since it is the ratio of two lengths.
In conclusion, systematically applying dimensional analysis using
linear algebraic tools and physically reasonable assumptions, we have derived the
important scaling relation for gradient design: L α η2a5
. Use of turns in the dimensional matrix
greatly simplifies the analysis.Acknowledgements
This work was partially supported by research grant NIH U01
EB024450.References
1.
Turner R. Gradient coil design: a review of
methods. Magn Reson Imaging. 1993;11(7):903-20.
2.
Hidalgo-Tobon SS. Theory of gradient coil design
methods for magnetic resonance imaging. Concepts in Magnetic Resonance Part A, 2010;
36A(4) 223–242.
3.
Lee SK, Mathieu JB, Graziani D, Piel J,
Budesheim E, Fiveland E, Hardy CJ, Tan ET, Amm B, Foo TK, Bernstein MA, Huston
J 3rd, Shu Y, Schenck JF. Peripheral nerve stimulation characteristics of an
asymmetric head-only gradient coil compatible with a high-channel-count
receiver array. Magn Reson Med. 2016 Dec;76(6):1939-1950.
4.
In MH, Shu Y, Trzasko JD, Yarach U, Kang D, Gray
EM, Huston J, Bernstein MA. Reducing PNS with minimal performance penalties via
simple pulse sequence modifications on a high-performance compact 3T scanner.
Phys Med Biol. 2020 Jul 31;65(15):15NT02.
5.
Endt vom A, Kimmlingen R, Riegler J, Eberlein E,
Schmitt F. A high-performance head gradient coil for 7T systems. Proceedings of
the 14th Annual Meeting of the International Society of Magnetic Resonance in
Medicine; 2006; p. 1370.
6.
Foo TKF, Laskaris E, Vermilyea M, Xu M, Thompson
P, Conte G, Van Epps C, Immer C, Lee SK, Tan ET, Graziani D, Mathieu JB, Hardy
CJ, Schenck JF, Fiveland E, Stautner W, Ricci J, Piel J, Park K, Hua Y, Bai Y,
Kagan A, Stanley D, Weavers PT, Gray E, Shu Y, Frick MA, Campeau NG, Trzasko J,
Huston J 3rd, Bernstein MA. Lightweight, compact, and high-performance 3T MR
system for imaging the brain and extremities. Magn Reson Med. 2018
Nov;80(5):2232-2245.
7.
Foo TKF, Tan ET, Vermilyea ME, Hua Y, Fiveland
EW, Piel JE, Park K, Ricci J, Thompson PS, Graziani D, Conte G, Kagan A, Bai Y,
Vasil C, Tarasek M, Yeo DTB, Snell F, Lee D, Dean A, DeMarco JK, Shih RY, Hood
MN, Chae H, Ho VB. Highly efficient head-only magnetic field insert gradient
coil for achieving simultaneous high gradient amplitude and slew rate at 3.0T (MAGNUS)
for brain microstructure imaging. Magn Reson Med. 2020 Jun;83(6):2356-2369.
8.
Szirtes T. Applied dimensional analysis and
modeling. McGraw-Hill 1997.
9.
Bernstein MA, Friedman WA. Thinking about
equations. Chapter 6, Introduction to dimensional analysis and scaling. Wiley 2009.
10.
Pascone R, Vullo T, Cahill PT. Theoretical and
experimental analysis of magnetic field gradients for MRI. 1993, available on
IEEE Explore.