Thomas K.F. Foo1, Louis Frigo2, Myung-Ho In3, Nastaren Abad1, Vincent B Ho4, and Matt A Bernstein3
1GE Research, Niskayuna, NY, United States, 2GE Healthcare, Waukesha, WI, United States, 3Mayo Clinic, Rochester, MN, United States, 4Uniformed Services University of the Health Sciences, Bethesda, MD, United States
Synopsis
A phase contrast
pulse sequence was modified to use a single-sided bipolar encoding to calibrate
the transverse gradient offsets for asymmetric gradient coils. The phase
difference approach cancels the concomitant gradient field effects from all
gradient waveforms except for the bipolar gradient waveforms. By fitting the
measured phase offset in the phase difference images to the applied bipolar
gradient amplitudes, the gradient offset value can be calculated. This was used
for pre-emphasis compensation for the zeroth and first order concomitant
gradient fields.
Introduction
Head-only
asymmetric gradients [1-3] are an approach to build high efficiency, small diameter
gradient coils for high-field imaging of the brain, extremities, and infants, but
still allow sufficient clinical access to accommodate imaging of human subjects.
However, asymmetric gradient coils also introduce additional zeroth-, first-order,
and higher-odd-order concomitant fields that need to be compensated for to
achieve reliable image quality [4-6]. Salient in the compensation
schemes is the knowledge and calibration of the spatial offsets of the asymmetric
transverse gradient coils in the z-direction. The Compact 3T (C3T) [2] and MAGNUS [3] head gradients have asymmetric
transverse gradient coils, while the z gradient coil is symmetric. Let z0x
and z0y denote the x and y gradient offsets along the
z-direction, for such asymmetric transverse coils, respectively.
To
adequately correct for zeroth and first order concomitant gradient terms
using pre-emphasis, one must accurately determine the value of these offsets.
The offsets can be estimated from the EM analysis of
the gradient coil. However, manufacturing tolerances and assembly can introduce
errors. The pre-emphasis can be calibrated using trial-and-error methods but a
more systematic approach is to use a calibration procedure to determine the z0x
and z0y offsets, which is described herein.Methods
The zeroth and first order concomitant fields are given by:
$$B_{error, 0th} \approx \frac{1}{2B_0} \left [ G_x^2 z_{0x}^2 + G_y^2 z_{0y}^2 \right ] \hspace{0.5in} (1) $$
and
$$B_{error,1st} \approx - \frac{G_x G_z z_{0x}}{2 B_0} x - \frac{G_y G_z z_{0y}}{2 B_0} y + \frac{ (G_x^2 z_{0x} + G_y^2 z_{0y})}{B_0}z \hspace{0.5in} (2) $$
assuming that $$$\alpha$$$ =
0.5 as the z-gradient coil is symmetric. To determine the values of z0x and z0y
for the Compact 3T (C3T) and MAGNUS gradient coils, a modified phase contrast
acquisition sequence with single-sided bipolar gradient waveform was used (Figure 1). The phase difference between the images acquired with the (+) and (-) encoding
waveforms yields a non-zero phase in a stationary phantom due to the
concomitant fields from the bipolar waveform. As the phase from the (+) or (-) bipolar waveforms are
$$ \phi = \gamma \int_{\tau} B_{error, 0th}(t) dt \hspace{0.5in} (3) $$
the phase difference is
$$ \Delta \phi = \phi^{(+)} - \phi^{(-)} $$
$$ \hspace{3.0in} = \gamma \left[ \int_{\tau} B_{error, 0th}^{(+)} (t) dt - \int_{\tau} B_{error, 0th}^{(-)} (t) dt \right] \hspace{0.5in} (4) $$
By varying the amplitude of the (-) encoding waveform through a range
of values, the phase difference measurements is fitted as a function of Gx2
or Gy2. As g3
is varied,
the gradient offset is determined from the fit to Eq.(4), where a1 is the coefficient of the quadratic term,
$$z_{0x} = \sqrt{ a_1(\frac{2B_0}{\gamma}) \left( t_{pw,3} + t_{pw,4}\left( \frac{t_{p,3} + t_{r,3}}{t_{p,4} + t_{r,4}} \right)^2 \right)^{-1} } \hspace{0.5in} (5) $$
where tp,3, tp,4, tr,3, tr,4 are the pulse widths of the bipolar waveforms (Figure 1), and tpw,i = [tp,i
+ (2/3) tr,i], with i=1,2,3,4
indicating the first and second gradient lobes of the (+) and (-) encoding
waveforms. For single-sided encoding, g1 = 0, and g2
is simply the gradient lobe acting as the dephasing gradient for the readout
gradient.
All experiments were
conducted on a C3T (80 mT/m and 700 T/m/s) and two MAGNUS (200
mT/m and 500 T/m/s) systems with 32-channel (NOVA Medical,
Wilmington, MA) receiver coils. An oil-filled 14-cm diameter phantom was used
for the experiments. To calibrate the offset for the x gradient, an axial 24-cm
acquisition with R/L set as the frequency encoding direction was used with a 256
x 256 acquisition matrix with 10-mm slice thickness. “Flow” encoding was along
the R/L direction. To calibrate the y gradient offset, “Flow” encoding was
along the A/P direction. As a bipolar encoding waveform is used in a stationary
phantom, a zero phase is expected. After the phantom settles, any residual phase
would be due to eddy currents or concomitant fields. The amplitude of the
bipolar encoding (g3) waveform was varied from $$$\pm$$$40% of the
maximum gradient amplitude.Results
From the EM design of the C3T and MAGNUS gradients, the phase
difference or measured phase error as a function of g3 is
shown in Figure 2 for the C3T gradient as well MAGNUS (unit #2). Using the
nominal EM value of z0x of 12 cm in Eq. (4) was slightly off from
the measured phase variation in both the C3T and MAGNUS gradients. Using the
measured value of z0x, the zeroth order concomitant frequency
correction provided adequate compensation of the concomitant field effects
(Figure 3). The
measured z0x offsets were for C3T, 10.7$$$\pm$$$ 0.2 cm, MAGNUS, 12.7$$$\pm$$$0.2 cm and 12.9$$$\pm$$$0.2 cm.
The departure of the measured values from the nominal EM values by $$$\pm$$$1
cm was expected due to manufacturing and assembly tolerances, as was the 0.2 cm difference between the two MAGNUS gradients coils.Discussion
An efficient and simple method to calibrate the concomitant gradient
field compensation using a phase difference approach has been described. The
gradient offset parameters are used for pre-emphasis compensation for the
zeroth and first order concomitant field effects (Figures 4 and 5). This proposed method eliminates a
trial-and-error approach to determine the correct scaling parameters for the
pre-emphasis correction and accounts for variations in the gradient offsets for
asymmetric gradient coils due to manufacturing tolerances.Acknowledgements
Grant funding: NIH U01EB026976,
NIH U01EB024450, and
CDMRP W81XWH-16-2-0054References
1. Weiger M, et al. Magn Reson Med 2018; 79:
3256–66.
2. Foo TKF, et al. Magn Reson Med 2018; 80:
2232–45.
3. Foo TKF, et al. Magn Reson Med 2020; 83:
2356–69.
4. Tao S, et al. Magn Reson Med 2017; 77:
2250–62
5. Weavers PT, et al. Magn Reson Med 2018;
79: 1538–44.
6. Meier C, et al. Magn Reson Med 2008; 60:
128–34.