Nastaren Abad1, Afis Ajala1, Yihe Hua1, and Tom K.F Foo1
1General Electric Global Research, Niskayuna, NY, United States
Synopsis
Impulse response function (IRF) harvesting is a
powerful technique for gradient characterization. In this study a field
monitoring system was utilized to characterize a head-only, high-performance
gradient coil (MAGNUS), enabling expansion of the spatial response pattern into
higher order spherical harmonics. The predictive power of IRF-estimated waveforms
and field evolution is demonstrated by comparing to nominal and active field
monitored approaches for arbitrary gradient waveforms. The calculated modulation
function indicated good correspondence for arbitrary waveforms and k-space trajectories
varying with slew rate, amplitudes and number of interleaves.
Introduction
Field perturbations caused
by eddy currents induced in gradient coils and conducting structures,
mechanical oscillations and fluctuations due to fast gradient switching are well known sources of image artifacts, compromise gradient performance, and impact MR image interpretations.
In previous studies, field monitoring with a dynamic field camera, has demonstrated improved image quality1.
However, field monitoring is limited by fast decay times of the probes (T2=35ms,
signal lifetime~50ms), with simultaneous measurements requiring the probes to either be integrated with the primary phased-array receiver coil, complicating
patient set-up or to be performed in a separate scan session causing system
interruptions.
A
practical alternative is to characterize gradient performance by using an
impulse response function (IRF), allowing deviations of the target gradient
waveforms to be corrected. Utilizing linear systems theory, the gradient system
can be approximated as linear time-invariant (LTI), providing a method to estimate
system performance and to predict arbitrary waveforms and trajectories. This
approach has previously been proposed in several studies2-3, where
short field measurements with known inputs were used. In this study a field
monitoring system1 (Skope, Zurich, Switzerland)
was used to characterize the gradient IRF, for a high-performance head-only gradient4.
The use of a field monitoring system, with multiple probes allows the decomposition
of the spatial response into higher order spherical harmonics, providing higher
order and cross-term information. Preliminary experimental results highlighting
calibration and performance of the gradient IRF are presented. Methods
Field
measurements were performed on a whole-body 3T magnet (GE SIGNA MR750),
equipped with a high-performance head-only gradient (MAGNUS)4
achieving 200mT/m and 500T/m/s with a standard 1MVA gradient driver. Triangular
waveforms (Figure 1), providing a close surrogate to delta functions were used
to probe the system response. A combination of ~60 waveforms, individually
applied to each axis, with rise times ranging between 31-320ms were used,
while maintaining the field cameras dynamic range. Field outputs were collated2
to yield the IRF using, $$
IRF(ω)= (∑Input_j^* (ω) Output_j (ω))/(∑|Input_j (ω)|^2 )$$
Data
was acquired without pre-emphasis compensation. As the MAGNUS gradient has
asymmetric transverse gradient axes (x-y), associated zeroth and first order
concomitant gradient are present. The IRF measurement included these, as well
as second order concomitant fields (also present in fully symmetric gradient
coils).
For
validation, IRF predictions were used to estimate arbitrary waveforms by
comparing predictions to nominal and field-monitored measurements. For
this, sinc and triangle waveforms were chosen. To estimate trajectories, a 2D single-slice spiral with SR =156T/m/s, 3072 points, and TE/TR=2.4/25ms
was acquired on a small grid-phantom and in a healthy volunteer (under an IRB
approved protocol) with a total readout time of 3ms per spiral interleave arm. k-space
trajectory prediction from the IRF, obtained by time domain convolution with
the nominal gradient input, was used for regridding non-cartesian k-space data.
The reconstructed images were compared to reconstruction with nominal trajectories as
well as using the measured trajectories (using field monitoring).Results and Discussion
Channel
specific mechanical oscillations in the IRF with distinct peaks at <1% of
the full response, were noted (Figure 2). The time domain representation
highlights the system delay without pre-emphasis correction. In addition to
delay times, the IRF is also able to capture subtle mechanically induced field
oscillations. Error plots (Figure 3) highlight correspondence between the IRF
prediction and field monitored approach for two arbitrary input waveforms. The
correspondence between IRF predicted and monitored trajectory correction is highlighted
with phantom (Figure 4 A/B) in vivo trajectory-based reconstructions (Figure
5).
Interestingly,
strong aliasing is not visible even with the nominal reconstruction, for both
phantom and in vivo implementations potentially attributable the systems
eddy current profile (design specs <0.05%). The benefit of the IRF prediction
is realized for acquisitions with longer readouts, where monitoring is limited
by probe dephasing (Figure 4 C&D), but adequate compensation can be
achieved with the predicted trajectories.
Residual
artifacts are potentially due to the absence of active compensation for linear (and
higher order) components of the concomitant field as well as static off
resonance, which can be compensated with B0 correction. Furthermore, the clip-on
sensors cover the edge of the imaging volume, where for a small head gradient
with a design 26 cm FOV for linearity and uniformity, measurements are
dependent on the accurate positional determination of the sensors. Errors in position calibration, or even displacement of the clip-ons can contribute to
errors. Conclusions
Unlike
concurrent field monitoring, the IRF needs to be determined once for a system,
providing a practical method to not only tune pre-emphasis, but account for
mechanically induced oscillations and correct for gradient imperfections. From
an image reconstruction perspective, the predicted trajectories can be actively
incorporated for real time correction, without scanner downtime. Furthermore,
the predicted waveforms and trajectories demonstrate equivalent correction to
the field monitoring system. Though an improvement to image quality is
apparent, this approach needs to be further tested for a wider range of
implementations, such as incorporating higher order field terms to refine or thermal
effects to test departure from the LTI approximation. Acknowledgements
Grant funding from
NIH U01EB28976, CDMRP W81XWH-16-2-0054References
1.
Dietrich BE, et al., Magn Reson Med.
2016
2.
Addy,
N. O., et al., Magn. Reson. Med. 2012.
3.
Vannesjo,
S. J. et al., Magn. Reson. Med. 2013.
4.
Foo TK, et al., Magn
Reson Med. 2019