Gunther Helms1
1Medical Radiation Physics, Clinical Sciences Lund, Lund University, Lund, Sweden
Synopsis
Controlling
the saturation of bound magnetization in variable flip angle (VFA) experiments
improves the derived parameter maps. Instead of using special CSMT RF pulses,
we kept the rmsB1 constant simply by increasing the RF pulse duration with the
square of the flip angle. At 3T, VFA T1 mapping of human brain thus showed
better compliance to the Ernst equation than at constant pulse duration or B1
amplitude, but yielded shorter T1.
Introduction
Magnetization
transfer (MT) effects in mapping T1 or T2/T1
varying the flip angle (VFA) in fast low-angle shot (FLASH) or balanced
steady-state free precession (bSSFP) sequences can be stabilized via the
average absorption of RF by the “invisible” motionally restricted macromolecular
protons1. Such “controlled saturation magnetization transfer”
(CSMT) can be achieved by complementing the root-mean-square of B1
by adding off-resonance sidebands to the RF pulse1.
As such RF pulses are not readily available on
clinical scanners, we suggest controlling rmsB1 by pulse duration (PL). This is
compared to schemes where RF duration or B1peak is hold constant in
VFA T1 mapping of human brain at 3T.Theory
For a given RF pulse shape and flip angle (FA), the
rmsB1 of an RF pulse is proportional to FA2/PL, which can be kept
constant by increasing PL with FA2. If PL is constant, rmsB1 increases
with FA2. For constant peak B1, rmsB1 increases with B1peak*FA. Absorption of RF will impose varying partial saturation to the bound pool, which -- compared to 1–cos(FA) created by the FA on free pool (Figure 1) -- will lead to MT to or from the free pool that influence the steady state signal2.Methods
Measurements were performed on a 3T MAGNETOM Prisma
using a 20-channel head-neck coil (Siemens Healthcare, Erlangen, Germany) using a prototype
sequence that allowed setting the pulse duration. A healthy male subject (29y)
and an agarose phantom (inversion recovery T1 = 1050ms) were investigated.
Sagittal non-selective volumes of 1.25 mm3 isotropic resolution (240x232x180 mm3) were acquired at
TE/TR=3.78/10ms in 1:30 minutes using partial Fourier and GRAPPA acceleration. The
flip angle of a rectangular pulse was varied through 2°, 3°, 4°, 5°, 6°, 8°,
10°, 12° and 15°, while recording the peak voltage. Three cases were
investigated:
A) constant rmsB1: A chosen minimum PL of 24us for 2° was
increased to 1800us for 15°. The peak voltage was less than 25% of the hardware
limit allowing for 4 times higher FA or 2 times shorter PL.
B) constant B1peak: PL was varied from 80us to 600us
(kept within the range of A).
C) A constant PL of 400us was chosen, to keep peak voltages within
the range of A.
The VFA signals S(FA) were plotted as Y=S(FA)/FA
over X=S(FA)*FA (FA in radian) to enhance signal bias by deviation from a straight
line3. Signal amplitude and T1 were obtained by linear
regression. The MR system’s B1+ mapping based on
multi-slice saturation recovery4 was used to account for local flip
angles deviating from the nominal value. Results and Discussion
At
constant rmsB1, the VFA signals followed the Ernst equation, that is, a
straight line in Figure 2, better than at constant PL or constant B1peak.
This is in line with results using CSMT pulses1. Specifically, the
latter showed lower signal at higher FA, where the saturation of the bound pool
increases.
After each RF pulse, more magnetization will be transferred from the
free to the invisible pool, resulting in a lower steady state. In turn, the
T1’s obtained with constant rmsB1 tended to be lower (Figure 3). Even at very long pulse lengths of 1800ms, effects of large B0 offset were not
observed orbito-frontally at 3T.
This
proof-of-principle suggests that it can be extended to higher flip angle and
use of sinc pulses. Further investigations of incidental MT-effects on T1
mapping by VFA will be compared to MP-RAGE based isotropic maps of T1 obtained from fitting the inversion recovery, which has been specifically
developed for the purpose of revealing spatial bias (abstract submitted), Conclusion
By simply increasing the RF duration with the square of
the FA, MT effects can be controlled in VFA, reducing systematic signal
deviation, but leading to lower T1 estimates.Acknowledgements
The
sequence was kindly provided by Dr A. Lutti, University of Lausanne, Switzerland.
Funding by the Swedish Research Council (NT-2014-6193) and local support by Dr. F. Testud,
Siemens Healthcare Sweden, is gratefully acknowledged. References
1) Teixeira RP AG, Malik SJ, Hajnal JV. Fast
quantitative MRI using controlled saturation magnetization transfer. Magn Reson Med 2019;81:907-920.
2) Al-Abasse Y, Helms G. Influence
of pulse length and shape on variable flip angle T1 mapping of the human brain. Proc ISMRM. 2016;24:696.
3) Helms G, Dathe H, Weiskopf N, Dechent P. Identification of
signal bias in the variable flip angle method by linear display of the
algebraic Ernst equation. Magn Reson Med 2011;66:669-677.
4) Chung S, Kim D, Breton E, Axel L.
Rapid B1+ mapping using a
preconditioning RF pulse with TurboFLASH readout. Magn Reson Med 2010;64:439-446.