Ryan McNaughton1, Hernan Jara1,2, Ning Hua2,3, Andy Ellison2,3, Lee Goldstein1,2,3, and Stephan Anderson2,3
1Boston University, Boston, MA, United States, 2Boston University Medical Center, Boston, MA, United States, 3Center for Translational Neuroimaging, Boston, MA, United States
Synopsis
Purpose: To test the PD-T1-T2 qMRI accuracy
of ultra-high spatial resolution compressed sense Tri-TSE. Methods: A healthy male volunteer (35yo) was scanned with an ultra-high
spatial resolution compressed sensing Tri-TSE pulse sequence. Maps of PD, T1,
and T2 were generated with voxel size of 0.4 x 0.4 x 1.2 mm3. Results: PD, T1, and T2 accuracy are
not affected by threefold compressed sensing acceleration. Conclusion: Compressed sensing does not appear to negatively impact
MS-qMRI accuracy and opens the door to ultrafast brain MS-qMRI at current
clinical spatial resolution or to a new high spatial resolution standard in the
clinical context.
Introduction
Increasing the speed of multispectral qMRI (MS-qMRI)
acquisitions is of paramount importance for developing clinically practical quantitative
protocols, particularly for high spatial resolution applications. Compressed
sensing (CS) is a recently developed technique based on the premise that if “the
underlying image exhibits transform sparsity, and if k-space undersampling
results in incoherent artifacts in that transform domain, then the image can be
recovered from randomly undersampled frequency domain data, provided an
appropriate nonlinear recovery scheme is used”1. In
this work, we tested a commercially available CS implementation in combination
with an ultra-high spatial resolution variant of the Tri-TSE pulse sequence to
test for MS-qMRI accuracy relative to its non-CS counterpart.Methods
A
healthy male volunteer (35yo)
was consented with an IRB approved research protocol. We tested CS-Tri-TSE at
3T (Philips, Ingenia Elition X) with a CS factor of 3 to generate self-coregistered
maps of the relaxation times (T1 and T2) and proton density (PD) at a native voxel size of 0.4 x 0.4 x 1.2 mm3. Other
selected parameters: TRlong=11,142ms, TRshort=550ms, TE1,2=13.75 and 110ms,
Ns=130 slices, Scan time=17min. The scanner CS-reconstructed directly acquired
dicom images were further processed with MS-qMRI algorithms programmed in Python
3.7 with the Enthought Deployment Manager. Two preparation steps are required
and performed with Fiji: 1) manual editing of the segmented intracranial matter
(ICM) segment and 2) manual delineation of the cerebellum’s superior aspect. MS-qMRI
algorithms were derived as functions of pulse sequence parameters according to
the Bloch equation model of the Tri-TSE, which is applicable across all MRI
platforms (Eq. 1-3).
Eq. 1: $$$T_2=cf_3\frac{TE1_{eff}-TE2_{eff}}{ln\left(\frac{DA_2}{DA_1}\right)}$$$
Eq. 2: $$$PD=\frac{cf_1}{C_{coil}}\cdot\frac{DA_1\exp\left(\frac{TE1_{eff}}{T_2}\right)+DA_2\exp\left(\frac{TE2_{eff}}{T_2}\right)}{1-\exp\left({\frac{-\left(TR_{long}-TSEshot_{DE}\right)}{T_1}}\right)}$$$
Eq. 3: $$$T_1=Root_{hybr}\left\{T_1+\frac{TR_{short}}{ln\left(\frac{1-\left(\frac{DA_3}{DA_1}\right)\left[\left(1-\exp\left(\frac{-TR_{long}}{T_1}\right)\right)-cf_2\exp\left(\frac{-\left(TR_{long}-TSEshot_{DE}\right)}{T_1}\right)\right]}{1+cf_2\exp\left(\frac{TSEshot_{SE}}{T_1}\right)}\right)}\right\}$$$Results
As shown in Fig. 1, the image quality
of the CS-Tri-TSE images is nearly indistinguishable from the non-CS
counterparts in terms of contrast and SNR. Remarkably, flow ghosting artifacts
(not shown) were less severe in the CS-Tri-TSE images, specifically in the
posterior fossa. Selected qMRI maps are shown in Fig. 2 with the corresponding
pixel value scale bars showing good quantitative agreement with values reported
in the literature2 as
further confirmed in the whole-brain histograms (Fig. 3). These histograms
further show that there is less parameter spread in the CS-maps, possibly
indicating less vulnerability to pulsation induced variability.Discussion
We tested a three-fold acceleration factor and
found that MS-qMRI accuracy and map quality are not affected. It is likely that
further improvements in pulse sequence development and compressed sensing
technologies could lead to routine ultra-high spatial resolution in the
clinical context.Conclusions
Compressed
sensing does not appear to negatively impact MS-qMRI accuracy and opens the door
to either, ultrafast brain MS-qMRI at current spatial resolution or to a new
high spatial resolution standard in the clinical context. This work could have implications for protocol unification and
standardization, and for ultra-high spatial resolution Synthetic-MRI and white
matter fibrography.Acknowledgements
No acknowledgement found.References
1. Lustig M, Donoho D, Pauly JM. Sparse
MRI: The application of compressed sensing for rapid MR imaging. Magnetic
Resonance in Medicine 2007;58(6):1182-1195.
2. Rooney
W, Johnson G, Li X, Cohen E, Kim S, Ugurbil K, Springer C. Magnetic field and
tissue dependencies of human brain longitudinal 1H2O relaxation in vivo.
Magnetic Resonance in Medicine 2007;57(2):308-318.