Difei Wang1, Rüdiger Stirnberg1, Eberhard Pracht1, and Tony Stöcker1,2
1German Centre for Neurodegenerative Diseases (DZNE), Bonn, Germany, 2Department of Physics and Astronomy, University of Bonn, Bonn, Germany
Synopsis
We propose a fast MPM protocol at 7T using skipped-CAIPI 3D-EPI with
simple PTx water-excitation based on 3 kT-points. By
comparison to corresponding CP mode scans, the 3 kT-points
excitation mainly improves the B1+ field homogeneity in
the Cerebellum. Using MPM B1+ field correction, this
simple improvement is sufficient to achieve good and homogeneous T1,
PD and T2* estimates throughout the brain. However, the
lack of MT homogenization still results in the inadequate MTsat
CNR. By combining MPM with EPI and PTx, we obtained quantitative
whole-brain parameter maps of high quality, except for Cerebellar
MTsat within 3 minutes scan time.
Introduction
Recently, following a fast multiparameter mapping (MPM)
implementation at 3T1, a corresponding protocol at 7T was proposed2 using 3D-EPI with segmented CAIPIRINHA (skipped-CAIPI) sampling3. In this work we propose to overcome a critical part of the
remaining challenges of the previous work
by utilizing water-selective kT-points pulses4.Methods
All data were acquired on a Siemens MAGNETOM 7T Plus scanner using
32-channel head receive coil and 8-channel transmit array. One set of
3D-EPI T1w, PDw and MTw images was acquired at 4
equidistant TEs between 4.64 and 19.64 ms at 1 mm isotropic
resolution. A 20.2x2z1 skipped-CAIPI sampling was
employed, resulting in an EPI factor of 4. Repetition time/flip angle
were chosen as follows: 27 ms/20° for T1w,
27 ms/3° for PDw,and 45 ms/4° for MTw. A single rectangular RF pulse with long
duration (1.0 ms) was used for simple fat suppression using
traditional circularly polarized excitation5. The corresponding
PTx scans were acquired using 3 kT-points RF
excitation pulses with the same nominal flip angles. Each sub-pulse
also had 1.0 ms duration for fat suppression as previously used for
fMRI4. Both CP and PTx MTw scans used the same Gaussian-shaped CP
mode RF pulse applied prior to each excitation.
In
order to keep the specific absorption rate (SAR) within the safety
limits, the nominal MT flip angle had to be reduced from 320° in the
CP mode MTw scan to 260° in the PTx-MTw scan. A second PTx-MTw scan
was acquired with prolonged TR (74 ms) to achieve the same 320°
nominal MT flip angle within SAR limits. The total acquisition time
of the CP mode MPM scans and the PTx MPM scans (incl. both MTw
scans) were 2:59 and 5:05, respectively. All imaging parameters are
summarized in Tab.1.
Two B1
maps were acquired using the same CP mode and PTx pulses by modifying
the 3D-EPI sequence to include two subsequent TRs according to the
Actual Flip angle Imaging (AFI) method6 with TR2 = 100
ms/TR1 = 20 ms, nominal flip angle = 50°, 5 mm isotropic
resolution, 4.2x2z1 skipped-CAIPI sampling (EPI factor 5),
TA = 0:10 for both.
All
parameter maps are calculated with the hMRI toolbox7.Results
Fig.1 shows representative sagittal, coronal and axial slices of all
scans at TE = 4.64 ms. In the PDw images, it can be observed that PTx
excitation is more homogeneous in the Cerebellum (red arrow). In the
T1w images, this shows as an improved WM/GM contrast
compared to the CP scan. However the PTx scans have slightly reduced
signals above the sphenoid sinus (blue arrow). The different nominal
saturation flip angles and TRs resulted in nearly 100% SAR estimation
(in “1st level safety mode“).
Fig.2 demonstrates the example sagittal slices of distortion
corrected MTw image along with MTw scan with uncorrected AP and PA
phase encoding. The difference is neglectable, therefore the
remaining results are presented for AP phase encoding only.
Fig.3 shows the quantitative parameter maps acquired using CP mode
and PTx pulses together with the corresponding B1+ scale
maps.
Fig.4 shows the (a) MTsat histogram and bar plots of (b) T2*,
(c) PD*, (d) T1 estimates in different regions
of interest (ROI) from CP and PTx scans. The smaller
error bars of T1 in WM and GM in PTx scans show that the transmit
field inhomogeneities could be compensated for in the parameter maps.
However, low MT saturation flip angles of all three scans are not
adequate to provide enough MT contrast: the low CNR is not sufficient
to create clearly separated modes in histogram8. In some ROIs, the
T1 estimates of PTx scans are slightly higher than that of
CP scans and the literature values.
Discussion
Our study shows that MPM protocol using skipped-CAIPI 3D-EPI with PTx
pulses is feasible.
The simplest form of
water-selective kT-points
PTx pulses is utilized in order to reduce the B1+ field
inhomogeneity with reasonable pulse prolongation and minimal SAR
increase. The PTx pulses achieve improved excitation in specific regions
like Cerebellum, which CP mode pulses could not excite sufficiently.
With an additionally acquired B1+ map, the remaining
transmit field variation was successfully counterbalanced, which was
not possible using CP mode pulses. Thus more homogeneous quantitative
parameter maps of high-quality can be obtained.
Although
PTx pulses provide the expected benefits, the acquisition currently
suffers from SAR limitations. The MT saturation flip angle had to be
reduced at the price of lower CNR. Currently, if MTsat is not
required, PTx scans can provide quantitative T1, T2*
and PD* maps at 1mm isotropic resolution of high quality
within 2 minutes scan time.
In a
future study, we will investigate the use of parallel transmit MT
pulses, or different CP mode MT saturation, like TIAMO13.Conclusion
We have presented a fast MPM protocol at 7T using skipped-CAIPI
3D-EPI with simple PTx pulses. This sufficiently improved the CNR of
T1w and PDw scans in the Cerebellum, while the MTw scans
still suffered from CP mode saturation in the current implementation.
With an additional B1+ map acquired, more homogeneous T1,
PD* and T2* maps of high-resolution
and high-quality can be obtained throughout the brain.Acknowledgements
No acknowledgement found.References
1. Wang D, Stöcker T, Stirnberg R. Fast Quantitative Multiparametric
Mapping using 3D-EPI with Segmented CAIPIRINHA Sampling at 3T. In:
Proceedings of the international society of magnetic resonance in
medicine.; 2020. pp. 4–7.
2. Wang D,
Stirnberg R, Stöcker T. Fast high-quality and -resolution multi
parameter mapping (MPM) at 7 Tesla using skippedCAIPI 3D-EPI. ESMRMB 2020. pp. 2–4.
3. Stirnberg R, Stöcker T. Segmented K‐space blipped‐controlled
aliasing in parallel imaging for high spatiotemporal resolution EPI.
Magnetic Resonance in Medicine 2021;85:1540–155.
4. Le Ster C, Moreno A, Mauconduit F, et al. Comparison of SMS-EPI
and 3D-EPI at 7T in an fMRI localizer study with matched
spatiotemporal resolution and homogenized excitation profiles. PLOS ONE 2019;14:e0225286.
5. Stirnberg R,
Brenner D, Stöcker T, Shah NJ. Rapid fat suppression for
three-dimensional echo planar imaging with minimized specific
absorption rate. Magn. Reson. Med. 2016;76:1517–1523.
6. Yarnykh VL.
Actual flip-angle imaging in the pulsed steady state: A method for
rapid three-dimensional mapping of the transmitted radiofrequency
field. Magnetic Resonance in Medicine 2007;57:192–200.
7. Tabelow K, Balteau E, Ashburner J, et al. hMRI – A toolbox for
quantitative MRI in neuroscience and clinical research. Neuroimage
2019;194:191–210.
8. Helms G, Dathe
H, Kallenberg K, Dechent P. High-resolution maps of magnetization
transfer with inherent correction for RF inhomogeneity and T1
relaxation obtained from 3D FLASH MRI. Magn. Reson. Med.
2008;60:1396–1407.
9. Andersson JLR, Skare S, Ashburner J. How to correct susceptibility
distortions in spin-echo echo-planar images: application to diffusion
tensor imaging. NeuroImage 2003;20:870–888.
10. Smith SM,
Jenkinson M, Woolrich MW, et al. Advances in functional and
structural MR image analysis and implementation as FSL. NeuroImage
2004;23:208–219.
11. Peters AM, Brookes MJ, Hoogenraad FG, et al. T2* measurements in
human brain at 1.5, 3 and 7 t. Magnetic Resonance Imaging
2007;25:748–753.
12. Rooney WD, Johnson G, Li X, et al. Magnetic field and tissue
dependencies of human brain longitudinal 1H2O relaxation in vivo.
Magnetic Resonance in Medicine 2007;57:308–318.
13. Orzada S,
Maderwald S, Poser BA, Bitz AK, Quick HH, Ladd ME. RF excitation
using time interleaved acquisition of modes (TIAMO) to address B1
inhomogeneity in high‐field MRI. Magnetic Resonance in Medicine
2010;64:327–333.