Hyungseok Jang1, Yajun Ma1, Michael Carl2, Saeed Jerban1, Eric Y Chang1,3, and Jiang Du1
1University of California, San Diego, San Diego, CA, United States, 2GE Healthcare, San Diego, CA, United States, 3Veterans Affairs San Diego Healthcare System, San Diego, CA, United States
Synopsis
In
this study, we explored the feasibility of applying single-point Dixon (spDixon)-based
fat suppression to a novel ultrashort echo time-based double echo steady state
(UTE-DESS) sequence for highly efficient morphological musculoskeletal (MSK)
imaging. Compared to the two-point Dixon-based approach used with conventional UTE-DESS
MSK imaging which requires two echoes for fat-water separation, we demonstrated
a new technique which requires only a single image in, removing the need for additional
data acquisitions. The feasibility and efficacy of our approach were
demonstrated in human knee joints.
Introduction
In double echo steady state (DESS) imaging,
fat signal commonly appears with high intensity due to a high T2/T1
ratio. To suppress this signal in DESS, the water excitation technique using a
spectral-spatial radiofrequency (RF) pulse has been widely used1–6. Recently, an ultrashort echo time-based
DESS (UTE-DESS) sequence was deemed feasible for imaging both short and long T2
tissues in human knee joint7. UTE-DESS imaging is disadvantaged by
strong fat signal (just as in DESS), but unfortunately the conventional water
excitation technique using composite, long-duration (~10ms) RF pulses8,9 cannot be applied in UTE-DESS targeting TEs
shorter than 0.1ms. Alternatively, fat suppression can be achieved
with two-point Dixon7, though this approach requires a longer
scan time to acquire an additional set of images with a different time delay. In
this study, we showed the feasibility of efficient single-point Dixon (spDixon)-based fat
suppression in UTE-DESS for morphological musculoskeletal
imaging. Methods
Figures 1A and 1B show the pulse sequence diagram and
k-space trajectory of 3D Cones-based UTE-DESS (UTE-Cones-DESS) sequence used in this study to acquire two images from free induction decay (FID)-like S+ and echo-like S- signal. The
spDixon method decomposes complex MR signal into individual fat and water
signals with a known chemical shift-induced phase difference, θ (Figure 1C)10,11. A
prerequisite of spDixon is the removal of any phase errors which
have been added to the complex MR signal, including initial phase offset, $$$\phi_{0}$$$, and
phase error due to B0 field inhomogeneity, $$$\phi_{B0}$$$. In
general, $$$\phi_{B0}$$$ is
determined by the field inhomogeneity, ΔB0, and by a
time delay during FID,
$$$\tau$$$, such that $$\phi_{B0} = \gamma ΔB_0 \tau$$, where
$$$\gamma$$$ is the gyromagnetic ratio of proton. $$$\phi_{0}$$$ is more complicated to model as it involves many factors including electric property of tissues, RF
coil, data readout, and reconstruction. Here,
we proposed to use the intrinsic signal properties of S+ and S-
in DESS to measure $$$\phi_{0}$$$ without any additional acquisitions. S-
is acquired at an echo time (TE) of (2xTR–$$$\tau$$$) in
the signal passage toward refocusing in the opposite direction to S+
on the transverse plane. Therefore, the phase of S+ and S-
in DESS can be written as $$\phi_{S_+} = \phi_{0}+\phi_{B0}+\phi_{c},$$ $$\phi_{S_-} = \phi_{0}-\phi_{B0}-\phi_{c}+\pi,$$ where
$$$\phi_{c}$$$ is the phase of the complex signal resulting
from a combination of fat and water signals. Then, $$$\phi_{0}$$$ can be
solved for by using both equations such that $$\phi_{0} = (\phi_{S_+}+\phi_{S_-}-\pi)/2.$$
Figure 1D shows a block diagram for the
proposed workflow to perform spDixon in UTE-DESS imaging, where it is expected
that $$$\phi_{B0}$$$ will be much smaller than $$$\phi_{0}$$$ because of the short $$$\tau$$$. Therefore,
correction of $$$\phi_{B0}$$$ is merely optional for rapid morphological
UTE-DESS imaging.
To evaluate the feasibility, six healthy
volunteers (two females, aged 35.3±8.8 years) and five patients with
osteoarthritis (OA) (five males, aged 51.0±8.1 years) were recruited and
underwent knee imaging in a 3T MRI scanner (GE-MR750) using a transmit/receive
8-channel knee coil. The protocol included UTE-Cones-DESS, field map acquisition, T1-weighted
fast spin echo (T1-FSE), and T2-weighted fast spin echo
(T2-FSE), as shown in Figure 1E. Results
Figure 2
shows the results of spDixon in UTE-DESS (37-year-old male). Without correction of $$$\phi_{0}$$$, spDixon yielded largely misestimated water and
fat signals (Figures 2G,H,L,M), while with correction of
$$$\phi_{0}$$$ spDixon
provided more reasonable fat and water images (Figures 2I,J,N,O). The estimated
$$$\phi_{B0}$$$ for the
prescribed $$$\tau$$$ was
negligibly small (Figure 3C) compared to $$$\phi_{c}$$$ (Figure 3A) and $$$\phi_{0}$$$ (Figure 2C).
Figures 3D-I show the comparison between spDixon without and with correction of
$$$\phi_{B0}$$$: the difference was small and did not
impair morphological information (red arrows).
Figure 4 shows the results of UTE-DESS imaging
with spDixon on the same volunteer, in which fat-suppressed water images were
obtained for both S+ and S- by utilizing the proposed
spDixon approach. Both weighted echo-subtraction (WES) with and without spDixon
achieved high contrast specific to the short T2 tissues for the osteochondral
junction (yellow arrows), tendons (red arrows), menisci (blue arrows), and
ligaments (green arrows).
Figure 5A-H shows results from a
representative OA patient (52-year-old male). Long T2 water signal from fluid is well-detected
in the water image of S-, which corresponds well with the T2-FSE
image (red arrows in Figures 5C,F). Sclerosis of subchondral bone is also well-depicted
as bright contrast in the S+ water image and the WES water image (yellow
arrows in Figures 5D,E,G). Note that the lesion is not well-depicted in the WES
image without spDixon due to the unsuppressed fat signal (Figure 5H). Figures 5I-P
show results from another OA patient (56-year-old male). Meniscal tear is detected in the S-
water image and the WES water image with or without spDixon (yellow arrows in
Figure 5M-P), which corresponds well with the T2-FSE and T1-FSE
images. Discussion and Conclusion
We
demonstrated that fat suppression overall improved lesion detection of long T2
water and subchondral bone sclerosis (Figure 5). spDixon-based fat suppression
strategy has the advantage of shorter scan time over the two-point Dixon-based
approach since intrinsic information derived from the UTE-DESS signal is
utilized. The proposed UTE-DESS with spDixon may provide a new imaging tool to
assess short T2 tissues such as the osteochondral junction, tendons,
ligaments, and menisci.Acknowledgements
The authors acknowledge grant support
from the NIH (R01AR062581, R01AR068987, R01AR075825, and R21AR075851), Veterans
Affairs (I01RX002604 and I01CX001388), and GE Healthcare.References
1. Kohl
S, Meier S, Ahmad SS, et al. Accuracy of cartilage-specific 3-Tesla 3D-DESS
magnetic resonance imaging in the diagnosis of chondral lesions: comparison
with knee arthroscopy. J. Orthop. Surg. Res. 2015;10:191 doi:
10.1186/s13018-015-0326-1.
2. Friedrich KM, Reiter G, Kaiser B, et al. High-resolution cartilage
imaging of the knee at 3T: Basic evaluation of modern isotropic 3D
MR-sequences. Eur. J. Radiol. 2011;78:398–405 doi: 10.1016/j.ejrad.2010.01.008.
3. Moriya S, Miki Y, Yokobayashi T, Ishikawa M. Three-Dimensional
Double-Echo Steady-State (3D-DESS) magnetic resonance imaging of the knee:
Contrast optimization by adjusting flip angle. Acta radiol. 2009;50:507–511
doi: 10.1080/02841850902849444.
4. Moriya S, Miki Y, Matsuno Y, Okada M. Three-dimensional double-echo
steady-state (3D-DESS) magnetic resonance imaging of the knee: establishment of
flip angles for evaluation of cartilage at 1.5 T and 3.0 T. Acta radiol.
2012;53:790–794 doi: 10.1258/ar.2012.110532.
5. Fujii H, Fujita A, Yang A, et al. Visualization of the peripheral
branches of the Mandibular division of the trigeminal nerve on 3D double-echo
steady-state with water excitation sequence. Am. J. Neuroradiol.
2015;36:1333–1337 doi: 10.3174/ajnr.A4288.
6. Qin Y, Zhang J, Li P, Wang Y. 3D Double-Echo Steady-State with Water
Excitation MR Imaging of the Intraparotid Facial Nerve at 1.5T: A Pilot Study.
Am. J. Neuroradiol. 2011;32:1167–1172 doi: 10.3174/ajnr.A2480.
7. Chaudhari AS, Sveinsson B, Moran CJ, et al. Imaging and T2 relaxometry
of short-T2 connective tissues in the knee using ultrashort echo-time
double-echo steady-state (UTEDESS). Magn. Reson. Med. 2017;78:2136–2148 doi:
10.1002/mrm.26577.
8. Hardy PA, Recht MP, Piraino DW. Fat suppressed MRI of articular
cartilage with a spatial-spectral excitation pulse. J. Magn. Reson. Imaging
1998;8:1279–1287 doi: 10.1002/jmri.1880080615.
9. Schick F, Forster J, Machann J, Huppert P, Claussen CD. Highly
selective water and fat imaging applying multislice sequences without
sensitivity toB1 field inhomogeneities. Magn. Reson. Med. 1997;38:269–274 doi:
10.1002/mrm.1910380216.
10. Ma J. A single-point dixon technique for fat-suppressed fast 3D
gradient-echo imaging with a flexible echo time. J. Magn. Reson. Imaging
2008;27:881–890 doi: 10.1002/jmri.21281.
11. Jang H, Carl M, Ma Y, et al. Fat suppression for ultrashort echo time
imaging using a single-point Dixon method. NMR Biomed. 2019:e4069 doi:
10.1002/nbm.4069.