Kazuo Kodaira1, Michinobu Nagao2, Masami Yoneyama3, Yasutomo Katsumata3, Takumi Ogawa1, Yutaka Hamatani1, Isao Shiina1, Yasuhiro Goto1, Mamoru Takeyama1, Isao Tanaka1, and Shuji Sakai2
1Department of Radiological Services, Tokyo Women's Medical University Hospital, tokyo, Japan, 2Department of Diagnostic imaging & Nuclear Medicine, Tokyo Women's Medical University Hospital, tokyo, Japan, 3Philips Japan, tokyo, Japan
Synopsis
Diaphragmatic one-dimensional navigation (NAV) is the
conventional approach for CMRA respiratory motion compensation. This method has limitations such as the cumbersomeness of
setting NAV and giving stresses to patients by wrapping the compression band
around the chest/abdomen for minimizing respiratory artifacts. Image-based
2D navigator (iNAV) can solve this limitation
because it directly corrects the translational movement of the heart in two directions (R-L
and F-H) by using 2D real-time imaging. We
investigate the feasibility of iNAV with loosely wrapped abdominal compression
band compared to the conventional NAV approach.
Introduction
The sequences of various methods have been proposed for whole heart coronary magnetic resonance angiography (WHC-MRA)1-4. In conventional
WHC-MRA, it is typically performed during free-breathing, and respiratory
motion is prospectively compensated by using a 1D-right-diaphragmatic-navigator
(NAV), measuring motion in the FH-direction5. However, the limitations
of the NAV-method are the cumbersomeness of setting the NAV on the right-diaphragm
and giving stresses to patients by wrapping the compression-band around the
chest/abdomen for minimizing respiratory artifacts.
Image-based-2D-navigator (iNAV) has the possibility to solve these problems. The iNAV allows prospective
correction of translational motion of the heart in two-directions (R-L, F-H) by
using 2D-real-time imaging to track the region surrounding the coronary
arteries throughout the cardiac cycle5 (Fig.
1-2). Furthermore, this method allows direct motion estimation and correction
of respiratory induced motion of the heart, obviating the need for a
diaphragmatic motion model6. Consequently,
the iNAV enables more robust suppression of respiratory motion artifacts for
WHC-MRA compared with conventional NAV, and images can be acquired in shorter
time and with improved image quality5-7. Therefore,
by using iNAV, we hypothesized that the image quality
of WHC-MRA can be maintained while shortening the scan time even if the
abdominal compression-band is loosened. In this study, we investigated the
feasibility of iNAV with a loosely wrapped compression-band for WHC-MRA by
comparing it with conventional NAV.Methods
A total of six volunteers (6 males; age range: 23~44) were
examined on a 1.5T MRI (Ingenia CX, Philips Healthcare). The study was approved
by the local IRB, and written informed consent was obtained from all subjects.
We used 3D-balanced-TFE (bTFE) for WHC-MRA images. Using bTFE as a base
sequence, we compared NAV-approach and iNAV-approach with loosely or tightly
wrapped abdominal compression-band
(NAV-loose/NAV-tight/iNAV-loose/iNAV-tight). For iNav scans, motion corrections for both
head-foot and left-right directions were enabled. For respiratory gating in
iNav, we applied Constant Respiratory efficiency UsIng Single End-expiratory
threshold (CRUISE) algorithm8. CRUISE assumes a fixed scan
efficiency of 50%. Therefore, the total scan time will be exactly double of the
displayed time.
The imaging-parameters of
bTFE: FOV=300×300mm, voxel-size=1.6×1.6×1.6mm, SENSE-factor=2.4, TR/TE/FA=2.2/1.1/80.
Scan time was recorded for each sequence. Image quality was
evaluated by visual score at the 10-points (RCA: #1/2/3/4 LAD: #5/6/7/8 CX:
#11/13) based on American-Heart-Association classification. We evaluated them
as 4-point grades (grade “4” was excellent, “1” was severe) by two blinded
readers.
For quantitative comparison, signal-to-noise ratio (SNR) and contrast-to-noise ratio
(CNR) were measured. The SNR was assessed in the blood, epicardial fat
and myocardium. To allow quantitative SNR measurements, we used a noise-measurement-method proposed by Zwanenburg et al9. Each sequence was
repeated with exactly the same receiver gain, but without any RF and gradient-pulses.
The reconstructed images showed only noise, including
the noise added due to the SENSE-reconstruction. The standard-deviation of a region of interest of the corresponding area
in the noise image was used as metric for the noise. SNRblood, SNRepicardial fat and SNRmyocardium were then calculated as follows:
SNRA = SI(A) / SDnoise(A)
Where SI are the mean average signal intensity of the blood, epicardial fat and myocardium respectively, and the corresponding SDnoise is the standard-deviation at the same location on
the noise images.
Subsequently, we measured the
CNR for comparing image contrast quantitatively. The CNR was estimated for blood and epicardial fat (CNRblood-epicardial fat). The CNRblood-epicardial fat was calculated by the following
equations:
CNRA-B = [SI(A) - SI(B)] / 0.5 [SDnoise(A) + SDnoise(B)]
The SNR and CNR were assessed by one-way repeated
measures analysis of variance (ANOVA) and the post- hoc Tukey-Kramer test.Results
Figure 3
shows the representative images using NAV-loose, NAV-tight, iNAV-loose and iNAV-tight. The results of visual score and average scan time are shown in Figure 4.
The iNAV-approach showed significantly higher
than NAV-approach regarding
RCA, LAD, except for the comparison between NAV-tight and iNAV-loose. Regarding CX, there was no significant
difference in the comparison of all sequences.
Average
scan time was significantly shorter iNAV-approach compared
with NAV-approach, except for the
comparison between NAV-tight and iNAV-loose.
Figure 5 shows SNR and CNR comparison
among four sequences. For the SNRblood, there was no significant
difference in the comparison of all sequences. For the SNRepicardial fat, iNAV-approach showed significantly lower value compared to NAV-approach. For the SNRmyocardium, there was no significant
difference in the comparison of all sequences. For the CNRblood-epicardial fat, iNAV-approach showed significantly lower value compared to NAV-approach.Discussion
From the
results of SNR and CNR, it is considered that iNAV-approach
has a better contrast between arteries and
background fat tissues than NAV-approach. We considered that iNAV-approach has higher fat
suppression effect than NAV-approach
because the additional SPIR-pulse in the iNAV might affect the results of fat suppression. Therefore, it is considered that iNAV-approach
showed a higher value in the visual score than NAV-approach.Conclusion
By using
iNAV for WHC-MRA, image quality could be maintained without tightly wrapping
the abdominal compression-band, and the scan time could be shortened compared
to the conventional NAV-approach. Therefore, iNAV can reduce patients’ stress. In
addition, iNAV eliminates the cumbersomeness of setting the NAV on the right
diaphragm, reducing the burden on the scanner. Therefore, iNAV could improve the workflow of
WHC-MRA.Acknowledgements
No acknowledgements
found.References
(1) Kazuo Kodaira, et al. Whole heart coronary MRA
with 3D non-selective bSSFP-DIXON: comparison with conventional methods. Proceedings of the Virtual
Conference & Exhibbition of ISMRM (2020).
(2) Oliver
M. Weber, PhD et al. Free-Breathing, Three-Dimensional Coronary Artery Magnetic
Resonance Angiography: Comparison of Sequences. Journal of Magenetic Resonance
Imaging. 2004; 20: 395-402.
(3) Yuji
Iyama et al. Single-Breath-Hold Whole-heart Unenhanced Coronary MRA Using
Multi-shot Gradient Echo EPI at 3T: Comparison with Free-breathing
Turbo-field-echo Coronary MRA on Healthy Volunteers. Magn
Reson Med Sci. 2018; 17: 161-167.
(4) Teresa
Correia et al. Technical note: Accelerated nonrigid motion-compensated isotropic 3D coronary MR angiography. Med Phys. 2018 Jan; 45(1): 214-222.
(5) Markus
Henningsson et al. Prospaective Respiratory Motion Correction for Coronary MR
Angiography Using a 2D Image Navigator. Magn
Reson Med. 2013; 69: 486-494.
(6) Markus
Henningsson et al. Whole-Heart Coronary MR Anigiography Using Image-Based
Navigation for the Detection of Coronary Anomalies in Adult Patients With
Congenital Heart Disease. Journal of Magenetic Resonance
Imaging. 2016; 43(4): 947-955.
(7) Markus Henningsson
et al. Whole-Heart Coronary MR Angiography With 2D Self-Navigated Image
Reconstruction. Magn Reson Med. 2012; 67: 437-445.
(8) Markus Henningsson et al. Coronary MR Angiography
Using Image-Based Respiratory Motion Compensation With Inline Correction and
Fixed Gating Efficiency. Magn Reson Med. 2018; 79:
416-422.
(9) Zwanenburg
et al. MR Angiography of the
cerebral perforating arteries with magnetization prepared anatomical reference
at 7T: comparison with time-of-flight. J Magn Reson Imaging. 2008; 28:
1519–1526.