Pan Liu1, Sidy Fall2, and Olivier Baledent2,3
1University of Picardie Jules Verne/CHIMERE EA 7516, Amiens, France, 2Facing Faces Institute/CHIMERE EA 7516, Amiens, France, 3Medical image processing department, University hospital Amiens, Amiens, France
Synopsis
Compared with CINE phase
contrast MRI (CINE-PC), echo-planar imaging phase contrast (EPI-PC) can achieve
real-time quantification of blood flow, with lower SNR. In this study, the
pulsating model of the simulated cerebral vasculature was used to verify
the accuracy of EPI-PC. The imaging time of EPI-PC was 62ms/image at 1.2*1.2mm2 spatial resolution and FOV=100*60mm2. The reconstructed EPI-PC flow curve was extracted by
homemade post-processing software. After comparison with CINE-PC flow
curve, it was concluded that EPI-PC can provide an average flow with less than
3% error, and its flow curve will be similar to the CINE-PC flow curve in shape.
INTRODUCTION
Today, CINE phase-contrast
MRI (CINE-PC) [1] is the primary noninvasive method for quantifying blood flow
and cerebrospinal fluid (CSF) in the cerebrovascular system by using cardiac
synchronization [2]. However, it cannot provide the effect of respiration on
blood/CSF flow [3] as it can only provide the average cardiac cycle
reconstructed using all cardiac cycles obtained during the acquisition.
Echo-planar imaging phase contrast (EPI-PC) [4] is a recent technique can achieve
real-time quantification of blood/CSF flow and does not require
synchronization. Meanwhile, EPI-PC is sensitive to noise and has a low
signal-to-noise ratio (SNR). The purpose of this study was to validate the
accuracy of EPI-PC in vitro and to quantify the effect of pixel size on both
sequences.Material
The study was performed on a clinical 3T
scanner (Philips Achieva, maximum gradient 80mT/m, gradient increase rate
120mT/m/ms) with a 32-channel head coil.
A flow phantom consisting of 4 tubes in
series with different diameters (Φ:9.5/6.4/4.4/2 mm) was attached to the pump
to mimicking the vessels in craniospinal system (Figure.1), and a static water-filled
tube was placed next to the tubes for calibration. The water pump provides
pulsatile flow at 99 bpm with an average flow rate of 1150 mm3/s. This
experiment will image the first tube (Φ9.5 mm) with CINE-PC and EPI-PC.
CINE-PC
provides 32 phase images to show an average pulse cycle. And EPI-PC contains
150 phase images to show the flow rate in real time (Figure.2).
Post-processing is performed by home-made
software [5] to extract the flow curves. Calibration is performed by measuring
the average velocity of the static tube. In addition, for comparison with the
CINE-PC, The EPI-PC flow signal is reconstructed as flow curve with 32 sampling
points in the same format as the CINE-PC flow curve.METHODS
Both sequences were imaged 10 times using
the default parameters (Figure.3) for tube-1. The average flow rate of the pump
(1150 mm3/s) and the cross-sectional area of the tube (70.8 mm2)
were used as gold standards, and the confidence intervals for the average flow
rate and the segmentation area were set to 100 ± 10% (average flow: [1035 mm3/s:
1265 mm3/s]; segmentation area: [63.72 mm2: 77.8 mm2]).
The Bland & Altman test was then used to verify the similarity of the
shapes of these two curves.
Secondly, based on the default parameters, pixel size
was used as a variable for 10 group images (from 0.8 mm to 4.4 mm with 0.4 mm
interval), and each group was repeated four times to observe its effect on the
two sequences.RESULTS
After 10 measurements, reconstructed EPI-PC
flow curves were synchronized with CINE-PC flow curves at the peak flow, as
shown in Figure.4-A. The average flow of EPI-PC was 1116 ± 24.5 mm3/s with a
coefficient of variation of 2.3%. The average flow of CINE-PC was 1239 ± 26.3
mm3/s with a coefficient of variation of 2.3%. Both the average flow and the
segmentation area (EPI-PC is 70.1 mm2 and CINE-PC is 70 mm2) are within the
confidence interval. The Bland-Altman plot (Figure.4-B) reflects that the
differences of the 32 points corresponding to the two curves are within the
limits of agreement (LoA), which can indicate that the two curves have similar
shapes.
Figure.5 shows that the average flow of
EPI-PC is within the confidence interval when the pixel size is between 1.2 mm
and 2.8 mm, while the pixel size of CINE-PC needs to be set between 1.2 mm and
2.4 mm to provide an accurate average flow.DISCUSSION
Under the present experimental conditions,
the reconstructed EPI-PC flow curve can replace the CINE-PC flow curve.
However, in vivo, cerebral arterial blood flow waves usually have higher
frequency components (4Hz-5Hz), and for the current sampling frequency, EPI-PC
may have difficulty in correctly representing these high frequency waveforms.
CINE-PC completes multiple phase encodings
in each pulse cycle and fills them in K-space of the corresponding phase
images, so that even if the acquisition time increases, it does not affect its
pseudo-sampling interval ∆t. However, for EPI-PC with real-time imaging, ∆t is
directly related to the acquisition time. Therefore, increasing the sampling
frequency can improve the accuracy of the reconstructed curve shape without
affecting the accuracy of the results. In the case of larger pixel size, not
only the SNR can be improved, but also the sampling frequency
can be increased. However, if the pixel size is too large, a partial volume effect
will occur [6], which distorts the results. The findings of this study indicate
that setting the pixel size of EPI-PC between 25%-30% of the target vessel/CSF diameter can obtain a higher SNR and a higher temporal
resolution.Conclusion
This experiment demonstrates the
feasibility of using EPI-PC for in vivo real-time imaging. After
post-processing with specific software, EPI-PC can provide the same flow curve
as CINE-PC and obtain time-dimensional data. Increasing the sampling frequency
allows for better display of the details of the reconstructed flow curve from
EPI-PC. With the application of more acceleration techniques such as compressed
sensing, EPI sequences will have greater potential for clinical diagnosis and
for studying the effect of respiration on blood/CSF flow.Acknowledgements
Thanks to David Chechin
from Phillips for his scientific support. Thanks to the
staff members at the Facing Faces Institute (Amiens, France) for technical
assistance.References
[1] Nj, Pelc, Herfkens Rj, Shimakawa A, and Enzmann Dr. Phase
Contrast Cine Magnetic Resonance Imaging. Magnetic Resonance
Quarterly 7, no 4 (1 octobre 1991): 229‑54.
[2] Balédent, O., Henry-Feugeas, M. C. and Idy-Peretti,
I. Cerebrospinal fluid dynamics and relation with blood flow: a magnetic
resonance study with semiautomated cerebrospinal fluid segmentation. Invest
Radiol 36, (2001): 368–377.
[3] Chen, Liyong, Alexander Beckett, Ajay Verma, and
David A. Feinberg. Dynamics of Respiratory and Cardiac CSF Motion Revealed
with Real-Time Simultaneous Multi-Slice EPI Velocity Phase Contrast Imaging.
NeuroImage 122 (15 novembre 2015): 281‑87.
[4] Pan LIU, Armelle LOKOSSOU, Sidy FALL, Malek MAKKI, and Olivier
BALEDENT. Post Processing Software for Echo Planar Imaging Phase Contrast
Sequence. ISMRM 27th, (2019): N. 4823.
[5] Tang Chao, Duane D. Blatter,
et Dennis L. Parker. Accuracy of Phase-Contrast Flow Measurements in the
Presence of Partial-Volume Effects. Journal of Magnetic Resonance Imaging
3, (1993): no 2.