Marylène Delcey1,2,3,4, Pierre Bour1,3,5, Isabelle Saniour6, Dounia El Hamrani1,3,5, Valery Ozenne1,3,5, Marie Poirier-Quinot6, and Bruno Quesson1,3,5
1IHU Liryc, Electrophysiology and Heart Modeling Institute, Fondation Bordeaux Université, Pessac, France, 2Univ. Bordeaux, Centre de recherche Cardio-Thoracique de Bordeaux, U1045, Pessac, France, 3INSERM, Centre de recherche Cardio-Thoracique de Bordeaux, U1045, Bordeaux, France, 4Siemens Healthcare SAS, Saint-Denis, France, 5Univ. Bordeaux, Centre de recherche Cardio-Thoracique de Bordeaux, U1045, Bordeaux, France, 6IR4M, UMR8081, Université Paris-Sud/CNRS, Université Paris-Saclay, Orsay, France
Synopsis
In patients presenting
cardiac electrical dysfunction, sub-millimetric resolution of
CMR would help improving
characterisation of the
arrhytmogenic substrate. Here,
a local
coil was placed
directly near the
anatomic region to image to
overpass the lack of sensitivity and selectivity
of current conventionnal coils and reduce the voxel
size. We
implemented a motion compensation method that exploits the signal of
micro-coils embodied on a catheter to
retrospectively sort radial k-space data prior to ESPIRIT image reconstruction. Using
this approach, we present high resolution (300µm), motion-resolved, image of
the heart obtained in vivo in pig.
Introduction
Precise 3D characterization of the arrhythmogenic substrate through high
resolution MRI would provide new insights for better understanding of
electrical dysfunctions in myocardium, such as Atrial
Fibrillation (AF). Despite MRI has proven to be valuable for assessing left
atrial (LA) myocardial tissue in patients suffering from AF using 3D delayed
enhancement (DE-MRI)1, the current spatial resolution (1.5x1.5x2.5
mm3) in clinical scanners remains insufficient since the atrial wall thickness
ranges 2 to 5 mm. Limited signal-to-noise ratio (SNR) inherent to large coils
surrounding the thorax, together with cardiac and respiratory motions are the
main limited factors to improve the image resolution. An intra-cardiac coil
would allow a substantial gain in both selectivity (reduced-FOV) and
sensitivity (gain in SNR) that in turn may be exploited to improve spatial
resolution. It has recently been reported2 that a local coil could
achieve high-resolution images (200 µm-in plane) on an ex vivo beating heart.
Here, we report on a proof of concept of in vivo high-resolution cardiac images
by combining a prototype MRI instrumentation with dedicated image acquisition
and reconstruction.Methods
A 2 cm
diameter receive-only loop coil was designed using a 35µm-thick copper trace and decoupled from the RF body coil during the transmission. A motion compensation
technique3 was implemented that exploits the signal of micro-coils
embodied on a MR-compatible catheter (Figure
1a). Determination of the 3D coordinates of the micro-coils with a
MR-tracking module4,5 (Figure
1b) was interleaved with acquisitions of N (typically 3 segments)
radial golden-angle k-space lines of a 2D gradient-recalled sequence (GRE). Accuracy
of the tracking was determined on phantom in static and mobile conditions. In
vivo evaluation was performed on a healthy anesthetized pig (protocol approved
by the local animal care authorities). After sedation, the animal was intubated
and mechanically ventilated (12 breaths/min). Heart rate and respiratory
signals were recorded by ECG leads and respiratory belt. An MR-compatible
steerable catheter of 9Fr (Vision-MR, Imricor, USA) was inserted through the femoral vein and navigated until the cavity of
the right ventricle. The chest of the animal was open and the loop coil was
positioned in contact with the epicardium of the left ventricle (LV)..
A localization sequence (TruFisp, TR/TE/FA =
381/1.2/80° 1.5 x 1.5x 4.5 cm3) was performed to ensure the loop
coil was correctly positioned relative to the LV before high resolution
imaging.
Acquisition
parameters were as follow: FOV =
150 x 150 mm2; Matrix = 448 x 448 px (in-plane resolution of 300 µm);
TR/TE/FA = 65/8 ms /65°; bandwidth = 255 Hz/Px; slice thickness = 3 mm; Radial
views = 12,000, no respiratory nor cardiac synchronization; 3 segments/TR.
Total acquisition time was 4 min 23 s. Tracking parameters were: FOV = 450 mm,
matrix = 512 data points, TR/TE/FA = 4.8 ms/2.8 ms/ 3°, bandwidth = 152
Hz/px. After acquisition, k-space data were sorted according to the catheter
position after converting signals recorded by the tracking coils into X, Y and
Z coordinates (Figure 1c).
The frequency spectrum of the direction presenting the dominant amplitude of
motion was plotted to visualize peaks corresponding to respiratory and cardiac
motions. A Gaussian low pass-filter (sigma = 8) was applied to retrieve the
respiratory signals. After subtraction of this filtered data to the initial
readings, a 1rst-order Butterworth low-pass filter (normalized cutoff frequency
0.2) was applied to extract displacement related to cardiac contraction. Eleven cardiac phases and 4 respiratory phases were created from these signals. Images
were reconstructed for each sub k-space using the ESPIRIT method6.Results
Tracking accuracy was found 0.3 mm on static phantom. Figure 2a displays 3D catheter
positions versus time for a 30 sec duration, together with its Fourier spectrum
in the Y direction (Figure 2b).
Figure 2c displays the
respiratory and cardiac signals extracted after processing (orange curves),
overlaid on the physiological signals recorded by external sensors (green
curves), showing good correspondence. Figure 3a is a scout view of the heart showing the coil and
catheter positions. Image in Figure
3b is obtained without data sorting, whereas image Figure 3c is the result of the ESPIRIT reconstruction of 1
cardiac phase and 1 respiratory phase after data sorting. The motion
compensation technique drastically improved the image quality when compared to
the original one (streaking artifacts are reduced and sharpness of the LV wall is
increased).Discussion
We demonstrate the feasibility of obtaining in vivo in
pig high-resolution (300 µm in-plane resolution) cardiac images using local 3D
motion descriptors (from microcoils) combined with a localized receiver coil to
increase SNR. The proposed sequence provides 3D positions of the instrument
that is closely linked to physiology, which allows for sorting k-space data
prior to iterative image reconstruction.
Such an approach opens perspectives for catheter-based imaging
(integrating both features in a single deployable instrument) of cardiac muscle
with unrivalled spatial resolution to better characterize the cardiac
substrate.Conclusion
We were able to combine a local loop coil with a motion compensation
algorithm to provide a high resolution image of the left ventricle while
overcoming motion artifacts.Acknowledgements
This work was supported by a grant from the French National Research Agency (PRC ANR) (N°ANR-19-CE19-0008-01)
This work was supported by Aquitaine Region (2016 –1R 30113 0000 7550/2016-1R 30113 0000 7553)
This project was supllied by France Life Imaging
This study received financial support from the French Government as part of the “Investments of the Future” program managed by the National Research Agency (ANR), Grant reference ANR-10-IAHU-04
References
1. Oakes RS, Badger TJ, Kholmovski EG,
et al. Detection and Quantification of Left Atrial Structural Remodeling With
Delayed-Enhancement Magnetic Resonance Imaging in Patients With Atrial
Fibrillation. Circulation. 2009;119(13):1758-1767.
2. Delcey M,
Saniour I, Vaillant F, Abell E, Bour P, Ben Hassen W, Poirier-Quinot M, Quesson
B In Proceedings of the 20th Annual Meeting of ISMRM, Sydney, Australia, 2020. Abstract
1196
3. Delcey M, Bour P, Ozenne V, Ben Hassen W, Quesson
B In Proceedings of the 20th Annual Meeting of ISMRM, Montreal, Canada, 2019.
Abstract 3847
4. Hillenbrand
CM, Elgort DR, Wong EY, et al. Active device tracking and high-resolution
intravascular MRI using a novel catheter-based, opposed-solenoid phased array
coil. Magn Reson Med. 2004;51(4):668-675. doi:10.1002/mrm.20050
5. Homagk
A-K, Umathum R, Korn M, et al. An expandable catheter loop coil for
intravascular MRI in larger blood vessels. Magn Reson Med.
2010;63(2):517-523.
6. Martin Uecker. Machine
Learning Using the BART Toolbox - Implementation of a Deep Convolutional Neural
Network for Denoising. Joint Annual Meeting ISMRM-ESMRMB, Paris
2018, In Proc. Intl. Soc. Mag. Reson. Med. 26;2802